The ABX Pentra 80 is a compact bench-top hematology analyzer that offers fast and accurate blood analysis. With a touch screen interface, 80 samples per hour processing capacity, and a wide range of parameters, it provides convenience and efficiency in small laboratories. The system includes reagents, controls, and allows for both manual and automatic sampling. It features a rack loader for easy sample handling and methods and technologies for accurate analysis. With customer maintenance features and environmental protection measures, it is a reliable and user-friendly option for hematology testing.
E N D
Product specification • Software specification • Performance • Overview • Reagents • Controls • Parameters • Sampling • Rack loader • Methods & technologies • Flags and alarms • Customer maintenance • Environmental protection
Overview Touch screen Reagent Compartment Processed Tubes Rack Loader 10 racks of 10 tubes Stat Sample Bay
Overview Compact bench top system No external computer is requested 12 inches Size: 34.1 x 23.3 x 22 in Weight: 55 kg / 122 lb 54 cm 57 cm 82 cm
Reagents • 5 reagents • ABX Cleaner • ABX Biolyse • ABX Basolyse II • ABX Eosinofix • + • ABX Diluent Onboard reagents
Reagents • ABX Pentra 80 provides : • Cyanide free lyse • Small cubitainer volumes • Bar- codes, batch number, level count down, logs • Waste level sensor
Control bloods • CBC & 5DIFF control bloods : • Bar-codes, batch number, expire date, logs • Target values uploaded by floppy disk • Multi-language package insert • CE marking
Parameters • CBC Mode (12 Parameters) • WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PCT, PDW • 5 DIFF Mode ( 26 Parameters) • WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, PCT, PDW • LYM, MON, NEU, EOS, BASO (% and #) • ALY (Atypical Lymphocytes) (% and #) • LIC (Large Immature Cells) (% and #) • 80 samples per hour • Manual or automatic • CBC or DIF
Sampling • ABX patented the MDSS concept: • Whole blood micro-sampling • 30 µl (CBC) • 53 µl (CBC+DIFF) • No sampling valve • No maintenance • No clogging • Customised Dilution Ratio • Extended linearity without manual intervention
sampling Automatic mode Random access 30 µl (CBC) 53 µl (CBC+DIFF) Internal and external bar-code reader Manual (open or close) mode
Rack loader • Correction of tube height • Position check • Automatic correction of a wrongly positioned tube • Automatic checking of tube presence • Accurate rack identification to avoid errors on sample ID Rack identification
Rack loader Rotation swirling End-over-end mixing (reference method Tube identification Accurate tube identification to avoid errors on sample ID
Methods and technologies • RBC/PLT - Impedance • The impedance principle is based on the reference method • HGB - Spectro-photometry • HCT - Numeric integration • WBC - Impedance & optical • BASO - Impedance & specific lyse
Methods and technologies DHSS • 1. Cytochemistry • Lyses red blood cells • Stains nuclei and cytoplasmic granules of leukocytes • Fixes the leukocytes in their original state (48 hours post-draw stability) • 2. Double hydrodynamic focusing • Hydrodynamic focusing before and after the calibrated micro-aperture • 3. Focused Flow Impedance • Cell volume by impedance method • 4. Light absorbance • Analysis of the internal structure by measuring the light absorbance of cells
Methods and technologies • RBC/PLT - Impedance • The impedance principle is based on the reference method • HGB - Spectro-photometry • HCT - Numeric integration • WBC - Impedance & optical • BASO - Impedance & specific lyse
Methods and technologies Product specification Methods & technologies • The determination of the 5 per differential uses : • X -> Cellular volume • Y ->Intensity of staining Eosinophils Large Immature Cells Neutrophils Monocytes Lymphocytes Atypical Lymphocytes
Methods and technologies • RBC/PLT - Impedance • The impedance principle is based on the reference method • HGB - Spectro-photometry • HCT - Numeric integration • WBC - Impedance & optical • BASO - Impedance & specific lyse
Flags and alarms • General flags • “l” and “h” (normal limit) • “L” and “H” (panic limit) • “x” (reject) • “!” (suspicion) • Pathologic messages RBC and PLT curves
Flags and alarms ALY: (Atypical Lymphocytes) Atypical Lymphocytes Plasmocytes, hairy Cells LIC: (Large Immature Cells) Blasts, myelocytes Promyelocytes, Metamyelocytes 5 per differential matrix 1- Priority alarms
Flags and alarms Nl: (Neutro-lympho) Hypo-segmented neutrophils Sezary cells, hairy cells Ne: (Neutro-eosino) Hyper-segmented neutrophils Hypo-granular eosinophils Mn: (Mono-neutro) Immature granulocytes, band cells 5 per differential matrix 2- Separation alarms
Flags and alarms No: (Noise) Platelet aggregates, NRBCs Lysis resistant RBCs Rm: (Right-mono) Large monocytes, blasts Ln: (Left-neutro) Cellular debris Rn: (Right-neutro) Immature granulocytes Ll, Ll1: (Left-lympho) Platelet aggregates, NRBCs 5 per differential matrix 3- Zone alarms
Customer maintenance Only one automatic shut-down per day Stepper motors replace the old compressor concept -> No noise, no maintenance A double needle replace the old shear valve concept -> Small sampling volume w/o maintenance Liquid valve replace the old pinch valve concept -> No clogging, no maintenance
Environmental protection Liquid waste 32 ml / sample
Environmental protection Product specification Environmental protection Solid waste 1,83 g / sample
Environmental protection Product specification Environmental protection Power Requirements • Power supply From 100 Vac to 240 Vac (50 HZ to 60 Hz) • Power consumption Maximum 230 VA
Product specification • Software specification • Performance • System specification • Worklist • Results • Data processing • Validation station
System Colour LCD touch screen: 12 inches Windows XP Floppy disk CD ROM drive 1xRS232C, 2xUSB port, 1x Ethernet port Random access (DIFF or CBC mode) Onboard validation-station Onboard user manuel User defined flagging Several operator Ids (4 access levels) Real-time management of patient’s report Automatic re-run with user-defined dilution factor (CDR)
System Actions toolbar Contextual toolbar
System • Mono or bi-directional connection: • ASTM or ABX format • Transfer of • Demographic data • Test data • Blood type(man, woman, child) • Tests requested (CBC, CBC/DIFF) • Query Mode
System • Qualify assurance management • Bar-codes • Lot number • Expiration date • Logs • 24 QC files • XB • Levey-Jenning graphs
System • Logs for : • Maintenance • Calibration • Controls • Errors • Data Handling • Patient record • Settings • Blank cycles • Connection Logs access by date
Worklist Real time visualisation Easy to set up Maximum security Manual entry or host connection Demographic data, department... Comment
Result 26 parameters 3 histograms & 1 matrix Zoom capability User-defined printout header Pathological messages Comments
Data processing Status screen Result statistics Cycles counters Rack processing Reagent management Remaining level Lot number Bar-code Expired date
Data processing Virtual view from rack performed Rack number Tube position Color code for type test Access to patient data by tube touch
Validation station Old # D+ % D- New • Automatic validation with user-defined : • Flags • Alarms • Limits • Delta check
Validation station On the same screen Patient history Run Re-run (w or w/o dilution) Patient reportbuilt from: Run Re-run Manual entry
Validation station Manual entry 5 Diff +Immature cells NRBC’s 100% diff control Automatic calculations Pathology level entry Comments entry
Validation station • Archives management • 10.000 patient reports • With curves and matrix • Flags, comment • Search capability by: • Date or patient ID
Product specification • Software specification • Performance • Linearity • Precision • Accuracy • Standards and Directives
Linearity Specifications - Performance Linearity Linearity limits were obtained and validated using the following dilution ratio: - WBC/LMNE 1/3 - RBC/HGB/PLT 1/2
Precision Parameters %CV Range WBC <1.5 % 4.0- 11.0 103 /µl RBC <1.5 % 4.0- 6.0 106 /µl HGB <1.0 % 11.0- 18.0 g/dl HCT <1.5 % 35- 55% RDW <2.0 % 80- 100 PLT <5.0 % 150- 400 103 /µ MPV <3.0 % 7.6- 10.9 NE% <3.0 % 50- 80% LY% <4.0 % 25- 50% MO% <8.0 % 2- 10% EOS% <15.0 % 0- 5% BAS% <20.0 % 0- 2%
Accuracy Parameters Mean % Mean difference difference WBC < 3 0.2 RBC < 3 0.10 HGB < 3 0.3 HCT < 4 1.5 PLT < 5 10
Standards and Directives • Directives • 98/79/EC - IVD medical devices • 89/336/EEC - EMC • 73/23/EEC - Low voltage • Standards • IEC 61010-1: 1990 + A1:1992 + A2:1995 • EN 61326 : 1998 • EN 61000-3-2 : 1995 • EN 61000-3-3 : 1998 • CSA - C22.2 No. 1010-1 : 1992 • UL 3101-1 : 1993
Product specification • Software specification • Performance • Features • Main features • Exclusive features
Main features Qualification & quantification of ALY and LIC Automatic and immediate rerun Real open tube and close tube for manual mode Real capillary mode (30 µl for CBC, 53µl for CBC+Diff) Integrated validation station Compact, simple, reliable (no shear valve, no compressor, only one PCB) Small reagent consumption Automatic shutdown cycle Low power consumption