1 / 30

SSD Status Before and After Dessica Installation

SSD Status Before and After Dessica Installation. Massimo Venaruzzo 04/10/2012 SSD – EN/CV Meeting. Outline. SSD General Status Before Dessica Humidity Trends Before Dessica Half ladder leackage current before Dessica SSD General Status with Dessica Humidity Trends with Dessica

enye
Download Presentation

SSD Status Before and After Dessica Installation

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. SSD Status Before and AfterDessicaInstallation Massimo Venaruzzo 04/10/2012 SSD – EN/CV Meeting

  2. Outline • SSD General Status BeforeDessica • HumidityTrendsBeforeDessica • HalfladderleackagecurrentbeforeDessica • SSD General Status withDessica • HumidityTrendswithDessica • HalfladderleackagecurrentwithDessica • MeasurementsperformedbySSD people • Conclusions

  3. SSD General Status – BeforeDessica Off topreventtrips 11/144 (~8%) Fluctuatingcurrent High temp High current Noise in the DQM

  4. HumidityTrendsBeforeDessica CR5 SSD-C (detector entrance)

  5. HumidityTrendsBeforeDessica SSD-A (detector exit)

  6. HalfladdersleakagecurrentbeforeDessica - 1 A607 (off) C609 (off)

  7. HalfladdersleakagecurrentbeforeDessica - 2 A624 (kept on and usedasreference) C601 (kept on and usedasreference)

  8. SSD General Status – WithDessica Currentstilltoo high High temp High current

  9. HumidityTrendsWithDessica CR5 SSD T Stable ~19.9 – 20.1 deg SSD-C DessicaStartup P2 Power Cut Dessicaalarm test

  10. HumidityTrendsWithDessica DessicaStartup SSD-A Techincal Stop 3

  11. HalfladdersleakagecurrentwithDessica - 1 A607 C609

  12. HalfladdersleakagecurrentwithDessica - 2 A624 C601

  13. Off CriticalLadderCurrents: Overall situation

  14. Measurementsperformedby SSD - 1 04/09, 16:30 - Dessica started with this setpoints: - RH: 15% (--> 2.03 g/kg) - T: 19 deg (measured near the machine ~18.3 deg, decreased to ~16.6 in the following days) - Flow measured ~314 m^3/h that increased to ~330 m^3/h in the following hours - P measured ~1.5 kPa - SSD C2-sensor RH (at the detector entrance): ~26% - SSD A -sensor RH (at the detector exit) : ~26% - A624: 195 uA - C601: 400 uA 07/09, 09.30 - Flow reduced to ~ 300 m^3/h, other parameters fixed - P measured: 1.26 kPa - SSD C2-sensor RH: ~20.3% - SSD A -sensor RH: ~21.5% - A624: 180 uA - C601: 375 uA

  15. Measurementsperformedby SSD - 2 10/09, 09.30 - RH setpoint reduced to 10% (-->1.35 g/kg), other parameters fixed - SSD C2-sensor RH: ~16% - SSD A -sensor RH: ~18% - A624: 160 uA - C601: 332 uA 11/09, 09.30 - Flow reduced to ~ 250 m^3/h (the nominal one), other parameters fixed - P measured: 0.864 kPa - SSD C2-sensor RH: ~16.8% - SSD A -sensor RH: ~17.9% - A624: 157 uA - C601: 327 uA Cleanessmeasurements

  16. Measurementsperformedby SSD - 3 Dessica flow 150 m³/h Flow reduction ~ 91% Dessica flow 250 m³/h Flow reduction ~ 80% Dessica flow 300 m³/h Flow reduction ~78.5% Dessica flow 350 m³/h Flow reduction ~ 78.3%

  17. SSD – C side (detector entrance) during SSD measurements DessicaStartup T=19deg, RH set = 15%, Flow ~330 m3/h SSD T Stable ~19.9 – 20.1 deg Dessicaalarm test Flow  250 m3/h, restfixed Flow  300 m3/h, restfixed RH  10%, restfixed Fisairmachine

  18. Conclusions • SSD conditionsignificantlyimproved: • RH stable at the requestedvalues • Criticalhalf-ladderscurrentdecrasing (a complete stabilizationneeds more workingweeks at the proper RH conditions) • Furtherinterventions • Leakcheckthenredo the flow measurements • Pipe insulation (togain in termsof air temperature at the detector), • Dessicasettingstobekept: 250 m3/h, RH 10% (1.35 g/kg)

  19. BackUp

  20. SSD HumiditySensor 04 September – 11 September

  21. Humidity in CR5

  22. SSD – C side (detector entrance) DessicaStartup T=19deg, RH set = 15%, Flow ~330 m3/h Dessicaalarm test Flow  250 m3/h, restfixed Flow  300 m3/h, restfixed RH  10%, restfixed Fisairmachine

  23. SSD – A side (detector exit) DessicaStartup Flow  250 m3/h, restfixed Flow  300 m3/h, restfixed RH  10%, restfixed Fisairmachine

  24. Flow Measurements

  25. FISAIR FLOW Sep 2010 - FISAIR flow 250 m³/h Jan 2012 - FISAIR flow 250 m³/h

  26. DESSICA FLOW

  27. CleanessMeasurements

  28. FISAIR machine 2009 FISAIR – DESSICA machine 2012

  29. SSD ReferenceCurrent