Electrical Impedance Tomography with Tomosynthesis for Breast Cancer Detection Jonathan Newell

Electrical Impedance Tomography with Tomosynthesis for Breast Cancer Detection Jonathan Newell. With: David Isaacson Gary J. Saulnier Tzu-Jen Kao Greg Boverman Richard Moore* Daniel Kopans*

Download Presentation

Electrical Impedance Tomography with Tomosynthesis for Breast Cancer Detection Jonathan Newell

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Electrical Impedance Tomography with Tomosynthesis for Breast Cancer DetectionJonathan Newell With: David Isaacson Gary J. Saulnier Tzu-Jen Kao Greg Boverman Richard Moore* Daniel Kopans* And: Rujuta Kulkarni Chandana Tamma David Ardrey Neha Pol RensselaerPolytechnicInstitute *Massachusetts General Hospital

  2. EIT electrodes added to mammography machine. • 1 : 2 : 4 : 2 : 1 is the ratio of the mesh thicknesses. • Only the center layer, III, is displayed in the results.

  3. EIT Instrumentation ACT 4 with Tomosynthesis unit Radiolucent electrode array

  4. Co-registration of EIT and Tomo Images To find the electrode position, display the slice containing the electrodes. Superimpose the mesh grid with correct scale. Slice 15 of 91 HS_14R Normal Then select the desired tomosynthesis layer. Slice 50 of 91

  5. Admittance Loci: format for summaries of EIS data Results of in-vitro studies of excised breast tissue.Jossinet & Schmitt 1999

  6. 120 EIS plots for a normal breast (HS14_Right)

  7. HS25_L: Invasive Ductal Carcinoma ROI 1 ROI 2

  8. LCM Image 700 Y Ym 0 compute for each voxel Linear Correlation Measure –LCM

  9. 700 0 LCM Image of invasive ductal CA (HS25_L) Gray scale image of LCM

  10. HS21_R: Fibroadenoma in the upper box ROI 1 ROI 2

  11. LCM Image of fibroadenoma (HS21_R) 350 0 Gray scale image of LCM

  12. HS10_L: Invasive Ductal CA“Proliferation is worrisome” ROI 1 ROI 2

  13. LCM Image of invasive ductal CA (HS10_L) 1300 0 Gray scale image of LCM

  14. LCM for 11 normal breasts 300 0 There are 120 EIS plots for layer 3 in each patient. The distribution of the LCM parameter in these plots is shown.

  15. LCM for the regions of interest in 4 patients Hyalinized Fibroadenoma Invasive ductal carcinoma Normal Normal The distributions of the LCM for the regions of interest identified. Note the LCM values are much larger for voxels associated with the malignant lesions.

  16. LCM on the same scale for a normal subject, a fibroadenoma and two carcinomas Normal Breast Fibroadenoma Invasive Ductal Carcinoma Invasive Ductal Carcinoma

  17. Conclusions • We have discovered a quantitative parameter derived from the admittance spectra that is markedly different in malignancy from normal breast tissues. • This parameter is location-specific to the tumor. • We need to study more patients with more benign lesions and tumors in order to evaluate the clinical potential of this parameter to improve the sensitivity and specificity of breast cancer screening. • We will present additional details about our system and its results in four posters at this meeting. See Kao, Boverman, Kulkarni and Tamma.

  18. Clinical details of the four patients.

  19. Summary of first 38 patients accrued • Total patient studied: 37-7 = 30 • First 7 patients were needed to develop correct operational procedures. • 19 breasts are underwent biopsy • 7 of these were reported as CA • 4 breasts of these were located outside of the electrode region. • 3 CA breasts are OK for analysis

  20. Threshold for Parameter, Straightness image • After preliminary study, we can set a threshold for normal or abnormal EIS plot. • 400 will be the first guess with these patient data. • The previous image will be show again with full scale set to 700.