Strength evaluation of prosthetic check sockets, copolymer sockets, and definitive laminated sockets

Strength evaluation of prosthetic check sockets, copolymer sockets, and definitive laminated sockets . Maria J. Gerschutz, PhD; Michael L. Haynes, MS; Derek Nixon, BS; James M. Colvin, MS . Aim

etoile
Download Presentation

Strength evaluation of prosthetic check sockets, copolymer sockets, and definitive laminated sockets

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Strength evaluation of prosthetic check sockets, copolymer sockets, and definitive laminated sockets Maria J. Gerschutz, PhD; Michael L. Haynes, MS; Derek Nixon, BS; James M. Colvin, MS

  2. Aim • Assess static failure loads for check sockets (CSs), copolymer sockets (CPs), and definitive laminated sockets (DLs). • Compare facilities, fabrication material and methods, and socket thickness. • Relevance • While International Organization for Standardization (ISO) standards exist for most lower-limb prosthetic components, they do not include prosthetic sockets.

  3. Methods • We used modified ISO 10328 to measure static failure loads of prosthetic sockets. • Compared them with criteria set by this standard for other components. Static socket testing configuration.

  4. Results • CS Static Strength • Influenced by thickness, material choice, fabrication method. • CP Static Strength • Depended on thickness, fabrication method. • DL Static Strength • Influenced by material choice, fabrication method. • Ductile loading criterion: • Most CSs and all CPs failed in a ductile mode. • Brittle loading criterion: • Most DLs failed in a brittle mode.

  5. Conclusions • Our results provide: • Foundation for understanding quality of prosthetic sockets. • Insight into possible routes for improving current care delivered to patients. • Comparative basis for future technology.