Surviving the POCT Inspection

Surviving the POCT Inspection. Best Practices for Ensuring Quality and Meeting Regulatory Requirements. A Laboratory Perspective. Frederick L. Kiechle, MD, PhD Chairman, Department of Clinical Pathology Medical Director, Beaumont Reference Laboratory William Beaumont Hospital Royal Oak, MI.

fathia
Download Presentation

Surviving the POCT Inspection

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Surviving the POCT Inspection Best Practices for Ensuring Quality and Meeting Regulatory Requirements. A Laboratory Perspective. Frederick L. Kiechle, MD, PhD Chairman, Department of Clinical Pathology Medical Director, Beaumont Reference Laboratory William Beaumont Hospital Royal Oak, MI

  2. Outline • Compliance improvement with connectivity • Quality management program for unit use devices • Continuous glucose monitors: pre-analytical, analytical and post-analytical factors • Plastic capillary tubes

  3. *Post RALS Plus implementation with the operator lockout feature.

  4. Unauthorized operators on all 61 nursing units before and after connectivity: Costs

  5. Quality control failures: Costs

  6. Reduction in labor costs after interface of the Inform with the LIS: 3 month period

  7. Conclusion • Point of care connectivity reduces user error, increases program compliance and decreases POCC and nursing costs • Point of care connectivity resulted in a total annual cost saving of $119,092

  8. Quality Management Program The Quality Management Program is built around sources of error based on the: • Device • Operator • Staffing

  9. Quality Management for Unit-Use Testing Proposed Guideline: NCCLS Document EP-18-P release for review (about 8/99) • QC should be performed “periodically” to access: • Reagent storage conditions • Operator competency • Electronic QC should be performed when possible

  10. So – Here We Are! The continuous measurement of glucose for a subset of difficult to control insulin-treated diabetes in a hospital is very appealing in the face of a shortage of MTs and nurses to perform POCT glucoses. However, the current continuous measurement devices are dependent on capillary glucose values for calibration.

  11. MiniMed Continuous Glucose Monitoring System • Interstitiul fluid glucose; 40–400 mg/dL measures every 10 sec and averages over 5 min for 72 hour (288/24hr) • Calibration: 4 SMBG throughout the day retrospective) which compares glucose meter/CGMS sensor data pairs of results by linear recognition • Data downloaded to computer: cannot calculate area under curve • No alarms

  12. GlucoWatch Biographer • Transdermal extraction of interstitial fluid glucose; 40 – 400mg/dL using low-level electric current • Extracts for 3 min; measures glucose, 7 min • Cycle time between measurements: 20 min • Periodic calibration with SMBG • Alarm for perspiration +/or hypoglycemia • Glucose oxidase and amperometric sensor (hydrogen peroxide)

  13. Uses of CMGS – Type I DM • Determine the number of episodes of nonsystomatic nocturnal hypoglycemia/hyperglycemia • Reportable range 40 – 400mg/dL • Calibration: 4 comparisons with SMBG device throughout this range • Tightly controlled type I values do not vary enough for adequate calibration  falsely low CMGS results which may lead to inappropriate decrease in overnight insulin dose Diabetes Care 2002;25:1499-1503

  14. Uses of CGMS – Type I DM • Validate use of SMBG as a proxy for integrated blood glucose level Diabetes Care 2002;25:1203-6 • Good correlation with HgbA1c • Mean glucose for 3 days Ann Clin Biochem 2002;39:516-7 • Area under glucose curves for 3 days, Diabetes Care 2002;25:1840-4

  15. Preanalytical Factors • Arterial vs. venous vs. capillary blood - SMBG • Inadequate instrument cleaning - SMBG • Incorrect QC procedure - SMBG/Cont • Sweat on body temp extremes - Cont - nocturnal hyperemia (vasodilation) • Systolic bp < 80mm Hg - SMBG/Cont - CPR, ICU • ICU poor correlation in 1st 6hr due to stress Scand J Clin Lab Invest 2002;62:285-92

  16. Analytical Factors • Glucose extremes: <40; >400 mg/dL - SMBG/Cont • Hematocrit extremes - SMBG/Cont • Improper technique - SMBG/Cont • IV dopamine: inhibits GO Rx - SMBG/Cont • Low total fraction - SMBG/Cont • Oxygenation status (PO2) - SMBG/?Cont • Premature sensor failure with loss of data - Cont

  17. Analytical Factors (cont.) • Direct oxidation of electroactive - SMBG/Cont species - ascorbate, urate, acetominophen • Implantation side inflammation: - Cont decreased sensitivity of sensor – catalase/ myeloperoxidase from granulocytes • Protein coating sensor surface - Cont

  18. Postanalytical Factor • Data entry • Calculation errors

  19. Future • Internal calibration system which would detect potential interferences with direct oxidation of electroactive species at the amperometric sensors, inflammation at the implementation site and/or protein coating of the sensor surface – alarms • Wireless connectivity to LIS/HIS • Software to calculate area under the curve

  20. CAP Gen .71032 – Phase I • Has the laboratory discontinued the use of plain glass capillary tubes for specimen collection and specimen handling?

  21. Plastic capillary tubes Roche microsampler, 240 l RAM Scientific, 230 l

  22. POCT Future • Noninvasive techniques • Transcutaneous bilirubin • Pulse oximetry • Connectivity • Greater number of applications • Decrease in size of immediate response lab