An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Mammography Regulatory Issues Bruce Matkovich Radiation Safety Section Michigan Dept. Of Consumer & Industry Services

  2. Outline • Mammography overview • Regulatory history • Inspection results • The future • Questions CIS – Radiation Safety Section

  3. Mammography Overview • Mammography is radiography of the breast • Need to use appropriate energy spectrum • Need high image resolution • Specialized equipment CIS – Radiation Safety Section

  4. MammographyMachine X-Ray Tube Compression Paddle Film Holder Foot Peddles CIS – Radiation Safety Section

  5. Regulatory History • 1988 – American College of Radiology’s Mammography Accreditation Program • Voluntary • 1989 – Michigan Public Act 56 • Essentially required ACR accreditation • Outlawed general purpose machines • 1991 – Annual state inspections begin CIS – Radiation Safety Section

  6. Regulatory History – Cont. • 1992 – Federal Mammography Quality Standards Act (MQSA) • Nationwide mammography standards • 1993 – State of Michigan mammography rules • Standards for technologists and physicists • QC and equipment standards • 1994 – Michigan Public Act 100 • Physician qualifications • Interpreters must be board certified CIS – Radiation Safety Section

  7. Regulatory History – Cont. • 1994 – MQSA interim regulations • 1995 – MQSA inspections begin • 1999 – MQSA final regulations CIS – Radiation Safety Section

  8. Inspections • Inspections involve: • Equipment checks • Personnel credentials review • Quality control records review • Image processing evaluation CIS – Radiation Safety Section

  9. Inspections – Cont. • Inspections are scheduled in advance • Typical inspection of a one-machine facility is 4 to 5 hours • MQSA inspection fee is $1549 for a one-machine facility • Michigan inspection fee is $127.88 per tube CIS – Radiation Safety Section

  10. Inspection Results – Equipment • Common problems: • Collimation problems (typically problems found with alignment of collimator light to x-ray field) • Automatic exposure control (AEC) tracking (maintaining consistent film darkness with varying patient thickness and density) CIS – Radiation Safety Section

  11. CollimationTest CIS – Radiation Safety Section

  12. AEC Test CIS – Radiation Safety Section

  13. Inspection Results –Personnel Credentials • Common problems: • Lack of documentation to verify compliance with continuing experience or continuing education requirements • Occasionally find personnel that are not qualified CIS – Radiation Safety Section

  14. Inspection Results –Quality Control • Common problems: • Failure to perform testing at required intervals • Failure to take corrective action when out of limits CIS – Radiation Safety Section

  15. Personnel and QC Records Review CIS – Radiation Safety Section

  16. Inspection Results –Image Processing • Common problems: • Under processing – determined using the Sensitometric Technique for the Evaluation of Processing (STEP) • STEP compares the facility’s processor to an optimized processor CIS – Radiation Safety Section

  17. Phantom Image and Dose • Mammography phantom image quality and mean glandular dose are important aspects of inspection • Phantom image and dose are obtained simultaneously using clinical techniques for average patient • Phantom image quality and dose are usually found to be in compliance CIS – Radiation Safety Section

  18. Phantom and Dose Testing Dosimeter Mammography Phantom CIS – Radiation Safety Section

  19. PhantomImage Fibers Speck Groups Masses CIS – Radiation Safety Section

  20. Phantom Image Failure Rates CIS – Radiation Safety Section

  21. Mean Glandular Dose • 200 millirads per view is the State of Michigan dose limit for an average-size patient • 300 millirads per view is the federal and ACR dose limit for an average-size patient CIS – Radiation Safety Section

  22. Mean Glandular Dose – Cont. • Changes in imaging technology are driving doses up • Michigan’s low dose limit is controversial • Michigan does not want to obstruct image quality improvement CIS – Radiation Safety Section

  23. Average Mammography Doses CIS – Radiation Safety Section

  24. The Future • Digital mammography • Only one FDA approved manufacturer currently • Must be used in a facility that has a screen-film unit • Digital advantages • Digitally enhance images • No film or wet chemistry processing CIS – Radiation Safety Section

  25. The Future – Cont. • FDA regulations that take effect October 28, 2002: • AEC must track to within ±0.15 OD of mean OD (currently ±0.30 OD) • Radiation output must be 800 mR/sec at 28 kVp (currently 513 mR/sec) CIS – Radiation Safety Section

  26. The Future – Cont. • More FDA regulations that take effect October 28, 2002: • Compression device must have fine adjustment • Maximum compression force for the power drive must be at least 25 lbs. and not more than 45 lbs. • Medical physicist must evaluate system resolution CIS – Radiation Safety Section

  27. The Future – Cont. • Proposed Michigan statute revisions to improve consistency with federal mammography standards: • “Medical director” to become “lead interpreting physician” • Physician initial experience and continuing experience requirements same as MQSA CIS – Radiation Safety Section

  28. The Future – Cont. • Proposed Michigan statute revision for a surety requirement to insure that patient records are available even if a facility closes. • Dose limit in the Ionizing Radiation Rules would probably be changed CIS – Radiation Safety Section

  29. For More Information Radiation Safety Website http://www.cis.state.mi.us/bhs/rss FDA’s Mammography Website http://www.fda.gov/cdrh/mammography/ CIS – Radiation Safety Section