FROST I Site Initiation Visit version 1.4. IVTM System. IVTM System. IVTM System. Control Buttons. To provide confirmation, each time a button is pressed, a “ key click ” sound is produced by the annunciator. Target Temperature
E N D
Control Buttons To provide confirmation, each time a button is pressed, a “key click” sound is produced by the annunciator. • Target Temperature • A screen that allows you to set the patient’s target temperature • RateDegree/Hr • Set a cooling/warming rate
Control Buttons • Standby/Run • Press to go between standby mode or run mode • An alarm can place the System into standby mode automatically • Press this button to return to run mode • Control Knob (Menu/Enter) • The dual-function control knob and pushbutton • Press the knob to display a menu screen • Turn the knob to scroll temperature trend graphs
SYSTEM SET UP • Check the level of coolant • If the level is below the bottom indicator line, add distilled water until the liquid is at the top indicator line • Plug in the power cord and turn the power switch on • The System performs a short self-test • At the end of the self-test there will be two short beeps
SYSTEM SET UP • Select the Bath Pre-Set: Pre-Cool • Choose the “Pre-Cool” option and press knob once to enter
SYSTEM SET UP • Are you starting a New Patient? • Always choose “NO” and press the knob once to enter the selection
SYSTEM SET UP • Select patient Target Temperature • Press the knob once to enter the selection
SYSTEM SET UP • Select Treatment Mode • Turn the knob to highlight the “Max Power” mode. Press the knob once to enter the selection • Wait until Self-Test is finished • If the System has not yet finished its self-tests, the Self-Test screen is displayed
System Sensor Check • Setup Sensor Checks • If the system finishes its self-test before you complete installing the Start-up Kit, you will see the following Check Screen • An item in RED requires your attention
Thermogard XP is now ready • When the Start-up Kit is loaded and primed, the system will enter STANDBY
Connection to the patient • Temporary disconnection from the Patient • Press the Standby/Run button to place the System in Standby mode • Disconnect the temperature probes from their cables • Leave the temperature probes in the patient • Using aseptic technique, disconnect the saline lines from the Quattro catheter and cap all lines or connect them to each other • Reconnecting after a temporary disconnection • Using aseptic technique, reconnect the saline tubing to the Quattro catheter Reconnect the temperature probes to their cables • Restart treatment by pressing the Standby/Run button
Ending Treatment • Press the Standby/Run button. • Using aseptic technique, disconnect both tubing connectors from the catheter. Cross connect the loose ends of the Start-up Kit • Disconnect the primary and secondary patient temperature probes from their cables
Ending Treatment • The cables connecting the device with temperature probe are reusable • Do not throw away cables for temperature probe • Remove the Quattro catheter from the patient • Leave the loose ends of tubing connectors from the Quattro catheter open, so that the saline solution can flow from the catheter when removing it from the vessel • Remove the patient temperature probe/s from the patient • Press the knob once. Select End Procedure