Image Processing for HTS SQUID probe microscope. Advanced Image Processing Seminar Colin Bothwell 0570063. General Discussion of Paper. Use High Temp SQUID probe microscope Probe allows high spatial resolution of measurement of samples even at room temp Aim to produce magnetic field images
E N D
Image Processing for HTS SQUID probe microscope Advanced Image Processing Seminar Colin Bothwell 0570063
General Discussion of Paper • Use High Temp SQUID probe microscope • Probe allows high spatial resolution of measurement of samples even at room temp • Aim to produce magnetic field images • Improve images using Image Processing
Theory of SQUIDs • Superconducting QUantum Interference Device • Measure tiny magnetic fields (threshold is around 10-14 T where 1T=10000Gauss. ) • Magnetic field of heart=10-10T
Theory of SQUIDs • To understand SQUIDs need to know basic principles • Superconductivity-no resistance • Josephson Effect- current flow carried by cooper pairs
Theory of SQUIDs • Josephson Junction-a junction of 2 S/C materials divided by non-S/C material (oxide) • With these principles, SQUID was designed
Theory of SQUIDs • SQUID has two Josephson Junctions • Apply current to SQUID • Voltage will oscillate • Oscillations depend on magnetic flux
Experimental Set-Up • Developed HTS SQUID probe microscope using fine permalloy needle as flux guide • Images the magnetic field by means of raster scan
Experimental Set-up • Separation of probe and SQUID adjusted by viewing through glass window • PC used to control SQUID
Magnetic Image-Deficiencies • Resultant images are noisy and unclear because of • Background Noise • Drifting • Data Jumping • So…
Image Processing! • Background Noise • Use image processing algorithm • Calculate mean and standard deviation of power spectrum • Threshold value=Pm + αPs • Apply this to each scan line in same way
Drift between Lines • Drift of flux bias point of SQUID • Treated with another algorithm • Drift calculated as difference in the mean value of magnetic field for each scan line
Data Jumping • Artifacts due to a jump in the flux bias point in SQUID by flux trapping or unexpected noise • Jump causes change dynamic range of magnetic field to become larger
Data Jumping • Small changes can no longer be detected • IP detects line with jump by derivation of line data (higher derivation considered to contain jump) so is interpolated with neighbouring lines
Conclusion • Used high-permeability needle to enable high spatial resolution measurement of samples. • Developed IP algorithms for magnetic field data obtained from SQUID probe microscope. • Algorithms can remove problems of background noise, drifting and jump data.