690 likes | 1.92k Views
Nuclear Medicine: Planar Imaging and the Gamma Camera. Katrina Cockburn Nuclear Medicine Physicist. Methods of Analysis. Once tracer has traced – need some method of analysing distribution. Imaging Gamma Camera, PET Camera. Compartmental Analysis Sample Counter. Radiation Detectors.
E N D
Nuclear Medicine: Planar Imaging and the Gamma Camera Katrina Cockburn Nuclear Medicine Physicist
Methods of Analysis • Once tracer has traced – need some method of analysing distribution Imaging Gamma Camera, PET Camera Compartmental Analysis Sample Counter
Radiation Detectors • Converts incident photon into electronic signal • Most commonly used detectors are scintillation • Photon interacts with crystal to convert incident photon into light photons • PMT changes light into electrical signal • Electrical signal recorded and analysed
Imaging Equipment • The Gamma Camera • Basic principle hasn’t changed since 1956!
Scintillation Imaging • Administration of Isotope
Scintillation Imaging • Localisation and Uptake
Scintillation Imaging • Localisation and Uptake
Scintillation Imaging • Localisation and Uptake
Scintillation Imaging • Localisation and Uptake
Scintillation Imaging • Localisation and Uptake
Scintillation Imaging • Localisation and Uptake
Scintillation Imaging • Localisation and Uptake Enhanced contrast between Organ of Interest and rest of body
Scintillation Imaging • Imaging distribution Gamma-rays emitted by radiopharmaceutical Collimator ‘selects’ only those rays travelling at right angles to face of camera Scintillation events in crystal recorded
Components of a Modern Gamma Camera • The components of a modern gamma camera Lead Shield Electronics PMTs Lightguide Crystal Collimator
The Collimator • The collimator consists of: • a lead plate • array of holes • It selects the direction of the photons incident on the crystal • It defines the geometrical field of view of the camera
The Collimator Detector Detector Patient Patient • In the absence of collimation: • no positional relationship between source – destination • In the presence of collimation: • all γ-rays are excluded except for those travelling parallel to the holes axis – true image formation
Types of Collimators • Several types of collimator: • Parallel-Hole • Converging • Diverging • Pin-Hole
The Scintillation Crystal • First step of image formation • Photon detected by its interaction in the crystal • γ-rays converted into scintillations
Can be thought of as “partial ionisation” Electrons excited and gain energy As electrons fall back to ground state, photons emitted Use of doping (eg NaI:Tl) creates smaller gaps Scintillation
Scintillation Crystal Properties • High stopping efficiency • Stopping should be without scatter • High conversion of γ-ray energy into visible light • Wavelength of light should match response of PMTs • Crystal should be transparent to emitted light • Crystal should be mechanically robust • Thickness of scintillator should be short
Properties of NaI(Tl) Scintillator • The crystal – NaI(Tl) • emits light at 415 nm • high attenuation coefficient • intrinsic efficiency: 90% at 140 keV • conversion efficiency: 10-15% • energy resolution: 15-20 keV at 150 keV
Disadvantages of NaI(Tl) crystal • NaI(Tl) crystal suffers from the following drawbacks: • Expensive (~£50,000 +) • Fragile • sensitive against mechanical stresses • sensitive against temperature changes • Hygroscopic • encapsulated in aluminium case
Lightguide and Optical Coupling • Lightguide acts as optical coupler • Quartz doped plexiglass (transparent plastic) • The lightguide should: • be as thin as possible • match the refractive index of the scintillation crystal • Silicone grease to couple lightguide, crystal and PMT • No air bubbles trapped in the grease
The Photomultiplier Tube • A PMT is an evacuated glass envelope • It consists of: • a photocathode • an anode • ~ 10 dynodes
The Photomultiplier Tube • Photocathode of PMT emits 1 photoelectron per ~ 5 – 10 photons • Photoelectronaccelerated towards first dynode • Dynode emits 3 – 4 secondary e- per photoelectron • Secondary e- accelerated towards next dynode • Multiplication factor ~ 106 • Output of each PMT proportional to the number of light photons
PMT Properties • The photocathode should • be matched to blue light • have high quantum efficiency • High stability voltage supply: ~1kV
Positional and Energy Co-ordinates • PMT signals processed • spatial information –X and Y signals • energy information – Z signal • Z signal – the sum of the outputs of all PMTs • proportional to the total light output of the crystal • Light output proportional to the energy of incident gamma • Pulse height analyser accepts or rejects the pulse
Pulse Height Analysis • Z-signal goes to PHA • PHA checks the energy of the γ-ray • If Z-signal acceptable • γ-ray is detected • position determined by X and Y signals • 20% window stillincludes 30% ofscattered photons
Image Acquisition Techniques • Static - (Bones, Lungs) • Dynamic - (Renography) • Gated - (Cardiac) • Tomography • SPECT • PET • List Mode - (Cardiac)
Static Imaging • Camera FOV divided into regular matrix of pixels • Each pixel stores number of gamma rays detected at corresponding location on detector • Typical Matrix Sizes: 2562, 1282, 642 1 Camera Computer Memory Image Display
Static Imaging • Camera FOV divided into regular matrix of pixels • Each pixel stores number of gamma rays detected at corresponding location on detector • Typical Matrix Sizes: 2562, 1282, 642 1 1 Camera Computer Memory Image Display
Static Imaging • Camera FOV divided into regular matrix of pixels • Each pixel stores number of gamma rays detected at corresponding location on detector • Typical Matrix Sizes: 2562, 1282, 642 1 1 1 Camera Computer Memory Image Display
Static Imaging • Camera FOV divided into regular matrix of pixels • Each pixel stores number of gamma rays detected at corresponding location on detector • Typical Matrix Sizes: 2562, 1282, 642 1 1 1 1 Camera Computer Memory Image Display
Static Imaging • Camera FOV divided into regular matrix of pixels • Each pixel stores number of gamma rays detected at corresponding location on detector • Typical Matrix Sizes: 2562, 1282, 642 2 1 1 1 Camera Computer Memory Image Display
Static Imaging • Camera FOV divided into regular matrix of pixels • Each pixel stores number of gamma rays detected at corresponding location on detector • Typical Matrix Sizes: 2562, 1282, 642 2 1 1 1 1 Camera Computer Memory Image Display
Static Imaging • Camera FOV divided into regular matrix of pixels • Each pixel stores number of gamma rays detected at corresponding location on detector • Typical Matrix Sizes: 2562, 1282, 642 3 1 1 1 1 Camera Computer Memory Image Display
Static Imaging • Camera FOV divided into regular matrix of pixels • Each pixel stores number of gamma rays detected at corresponding location on detector • Typical Matrix Sizes: 2562, 1282, 642 3 1 1 1 1 Camera Computer Memory Image Display
Dynamic Imaging • Series of sequential static frames • E.g. 90 frames each of 20s duration • Image rapidly changing distribution of activity within the patient • Used in Renography
Dynamic Imaging Analysis Split Renal Function Curves showing changing renal activity over time ROIs
Gated Imaging • Several frames acquired covering the cardiac cycle • Acquired over many cycles