200 likes | 398 Views
BME 595 - Medical Imaging Applications Part 2: INTRODUCTION TO MRI Lecture 3 Multi-Dimensional Imaging Feb. 28, 2005. James D. Christensen, Ph.D. IU School of Medicine Department of Radiology Research II building, E002C jadchris@iupui.edu 317-274-3815. References.
E N D
BME 595 - Medical Imaging ApplicationsPart 2: INTRODUCTION TO MRILecture 3 Multi-Dimensional ImagingFeb. 28, 2005 James D. Christensen, Ph.D. IU School of Medicine Department of RadiologyResearch II building, E002C jadchris@iupui.edu317-274-3815
References Online resources for introductory review of MRI physics: • Robert Cox’s book chapters online • http://afni.nimh.nih.gov/afni/edu/ • See “Background Information on MRI” section • Mark Cohen’s intro Basic MR Physics slides • http://porkpie.loni.ucla.edu/BMD_HTML/SharedCode/MiscShared.html • Douglas Noll’s Primer on MRI and Functional MRI • http://www.bme.umich.edu/~dnoll/primer2.pdf • Joseph Hornak’s Web Tutorial, The Basics of MRI • http://www.cis.rit.edu/htbooks/mri/mri-main.htm Book covering Multi-Dimensional MRI: • E. Mark Haacke, et al. Magnetic Resonance Imaging: Physical Principles and Sequence Design, 1999.
2D Echo-Planar Imaging (EPI) Pulse Sequence K-Space Trajectory
3D Gradient-Echo Pulse Sequence Phase Spoiling: (RF or gradients) causes Incoherence in transverse magnetization
Fast-Imaging(Steady-State Incoherent pulse sequence) For a train of RF pulses repeated every TR seconds The steady-state equilibrium magnetization is where The transverse magnetization following the nth RF pulse is The Steady-State Incoherent signal intensity is proportional to M
Flip Angle (θ), TR & T1 in Fast Imaging where Steady-State Incoherent (SSI) signal intensity:
Homework Problem 1