50 likes | 121 Views
Reference Membrane Location. Reference Membrane. Span Check. Perform 8 minute count Compare mass value with ABS value ABS must be within 5% or BAM will log CAL error (D error code). CALIBRATION MODE. REF MBRN: <. COUNT (Io):. 634000. COUNT (I):. 556234. CAL MASS M:. 0.801 mg/cm2.
E N D
Span Check • Perform 8 minute count • Compare mass value with ABS value • ABS must be within 5% or BAM will log CAL error (D error code)
CALIBRATION MODE REF MBRN: < COUNT (Io): 634000 COUNT (I): 556234 CAL MASS M: 0.801 mg/cm2 START STOP EXIT Calibration Test Screen
Hands On Workshop – Reference Membrane • Remove Reference Membrane • Inspect Reference Membrane • Reinstall Reference Membrane • Inspect mechanical movement of Reference Membrane • Perform Reference Membrane Count/Mass Test in TEST/CALIBRATE Screen • Compare Mass value with ABS value