CsI Compton Veto Detector for A low Mass WIMP Experiment. He Dao KPS@Jeju University 23 Oct 2004. Content. Introduction of our low mass wimp experiment Structure of CsI detector Some test and conclusion plan. Status of WIMP detection. The thresholds are a few keV
E N D
CsI Compton Veto Detector for A low Mass WIMP Experiment He Dao KPS@Jeju University 23 Oct 2004
Content • Introduction of our low mass wimp experiment • Structure of CsI detector • Some test and conclusion • plan
Status of WIMP detection • The thresholds are a few keV • Corresponding to the ~10GeV mass limit measurable
CRESST: lowest threshold • Cryogenic bolometer • 262g Sapphire detector • The lowest threshold gotten until now: 500keV
Our Idea Using low-threshold detector to “find” low-mass WIMP
Our detector • 5g Ge detector Shielding
Background Control • Cosmic Ray: Underground lab • Material background: purest material: Ge • Other particles: Multi-layer shielding: Pb, Cu, CsI veto detector PSD(Pulse Shape discrimination)
Energy Threshold of Detector • Threshold? less than 100eV after PSD PSD Parameter: for every current shape 100eV
The Aim of Our experiment Our estimated results: Target mass: 5g Counts rate: 1 count/keV·kg·day Threshold: 100 eV Upgrade to 1kg Ge
CsI Crystal Structure • Top: • 150mm50mm • 29 degree slope from height 30mm • 76mm at top • Bottom: • 150mm30mm
CsI Crystal Structure • Wall (6 parts): • R=75 mm; • r=45 mm • 5 L=220 mm; • 1 L=171 mm; • Outlet hole (see pic before)
Photon Collection • PMT performance: • CR119, one, at the top of CsI crystal • low background, good stability • Minus HV: -1.5KV • Anode resistance : 10K • Current: < 1mA • Reflecting film covers all surface of crystal .
Compton Veto Function Shielding for high energy gamma.
Primary Performance Test • Place : Tsinghua, China • Time : 16-18,Aug,2004 • Source : Cs-137, Co-60 • Bias: 12V • HV: -1000V • Void inside
Co-60 test result • Resolution becomes better when the source moves form bottom to top. • Peak moves since amplifier threshold adjusted.
Cs-137 test result The result seems bad while source near the bottom. That is because of the energy which is not high enough.
Test Conclusion • Photon collection efficiency Efficiency becomes worse while source moving from top to bottom because only one PMT is used, at top. • Ray Energy Higher energy ray can be detected more easily than lower one, which is good for high energy gamma ray shielding in this experiment.
Plan • Now: • CsI detector was sent to Korea • Nov: • Installation in Y2L and advanced test • MC simulation • Dec: • Built up with HPGe detector