An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Blood Establishment Computer Software (BECS) Michael Gorman, BA, MT(ASCP)SBB Consumer Safety Officer, CBER, OBRR, DBA September 15, 2009

  2. Outline • What is a BECS? • Regulations for Manufacturers • Manufacturers: Medical Device 510(k) Clearance • Regulations for Users • Validation • User Reporting of Implementation • CBER 510(k) Cleared BECS Website List • References

  3. What is a BECS? • Medical Device: Defined under the Federal Food, Drug, and Cosmetic Act (FD&C Act) • FD&C Act applicability to BECS: • Device intended for use either in the: • Diagnosis of disease or other conditions OR • Cure, mitigation, treatment, or prevention of disease in man or other animals

  4. What is BECS?(cont.) • Part of a computer system used by Blood and Source Plasma establishments, which also includes: • Hardware • SOPs and other documentation • Staff

  5. BECS Examples Software Applications Used For: • Receiving and storing data for later use during the manufacturing process - examples: • Test results • Donor suitability/eligibility • Blood/blood component suitability – release for transfusion or manufacture into injectable or noninjectable products • Computer-assisted self-interview (CASI) donor questionnaires • Performing compatibility testing and other transfusion service related functions

  6. Regulations for BECS Manufacturers • Medical Device Regulations: • 21 CFR 800 to 898 • Quality System Regulation (QSR) • 21 CFR 820 • Registration: Establishment Registration • 21 CFR 807 • Listing the Device: Medical Device Listing • 21 CFR 807 • Medical Device Reporting (MDRs) • 21 CFR 803 • Labeling • 21 CFR 801

  7. BECS Manufacturers: Medical Device 510(k) Clearance • Premarket Notification (510k) required under the FD&C Act • Allows a medical device, BECS, to be introduced into interstate commerce: • Sold or given away (gratis) interstate • FDA also currently considers interstate data access or transmission to be interstate commerce

  8. Regulations for BECS End Users • Blood Establishments • 21 CFR 211.68 • 21 CFR 606.100(b)(15) • 21 CFR 606.160(b)(5)(i) • 21 CFR 601.12 • Source Plasma Establishments • 21 CFR 606.100(b)(15) • 21 CFR 606.160(b)(5)(i) • 21 CFR 601.12 • Voluntary Medical Device Reporting

  9. BECS End Users User Site Validation Overview • Done outside developer’s controlled environment • 21 CFR 211.68 (Blood Establishments Only) • 21 CFR 606.160(b)(5)(i) • 21 CFR 606.100(b)(15) • General Principles of Software Validation; Final Guidance for Industry and FDA Staff  (1/11/2002) • Can be performed by: • End user • Third party (consultants) • Developer/manufacturer

  10. BECS End Users User Site Validation Overview(cont.) • Perform at locations where the software is implemented (installed) • Follow a pre-defined written plan • Retain documented evidence of all testing and procedures

  11. BECS End Users Validation SOPs Must be Developed • 21 CFR 606.160(b)(5)(i): Records shall be maintained for calibration and standardization • 21 CFR 606.100(b)(15): Equipment maintenance and calibration schedules and procedures

  12. User Site Validation Should: * • Use actual hardware and software that will be part of the installed system • Be performed within the context in which the software is intended to be used • Take into account any site specific issues *General Principles of Software Validation; Final Guidance for Industry and FDA Staff  (1/11/2002)

  13. User Site Validation Should: *(cont.) • Ensure all system components are exercised and are the specified versions • Be done throughout the full range of operating conditions • Be carried out for sufficient time so the system encounters a wide range of conditions

  14. User Site Validation Should: *(cont.) • Encompass: • High volumes of data • Heavy loads or stresses • Security • Error messages • Faults • etc.

  15. Reporting of BECS Implementation to FDA • BECS: Annual Report - 21 CFR 601.12 • CASI: Discussed in CASI presentation

  16. CBER 510(k) Cleared BECS Website List http://www.fda.gov/BiologicsBloodVaccines/BloodBloodProducts/ApprovedProducts/SubstantiallyEquivalent510kDeviceInformation/ucm134987.htm

  17. CBER 510(k) Cleared BECS Website List (cont.) • List of BECS is only for versions cleared by FDA • List does not include versions that did not require a new 510(k) submission • Not all software changes require a new 510(k) • List is not updated if a manufacturer discontinues distribution and support

  18. References: General • FD&C ACT: http://www.fda.gov/RegulatoryInformation/Legislation/FederalFoodDrugandCosmeticActFDCAct/default.htm • CFR on-line access - 1996 to 2009: http://www.access.gpo.gov/nara/cfr/cfr-table-search.html#page1

  19. References: Reporting BECS Implementation to FDA • 21 CFR 601.12: Changes to an approved application • Guidance for Industry: Changes to an Approved Application: Biological Products: Human Blood and Blood Components Intended for Transfusion or for Further Manufacture  (8/07/2001)

  20. References: User Site Validation • General Principles of Software Validation; Final Guidance for Industry and FDA Staff  (1/11/2002) • Draft Guidance for Industry: Blood Establishment Computer System Validation in the User's Facility  (10/26/2007) • Draft guidance comments to the docket are under review • Guidance Link: FDA website A to Z Subject Index - G Page; Click Guidance Documents – Biologics • http://www.fda.gov/SiteIndex/ucm148505.htm