370 likes | 378 Views
This talk provides an overview of the SHMS PID performance during the Spring and Fall 2018 experiments, focusing on the calorimeter and Cherenkov cut efficiencies, pion contamination, and other issues.
E N D
PID Capabilities in Hall C Simona Malace Jefferson Lab Talk at the J/psi Collaboration Meeting, October 26 2018
Overview • SHMS PID performance during Spring 2018 (E12-10-002) • SHMS Calorimeter and NGC and HGC response • Pion contamination for standard cuts • Calorimeter and Cherenkov cut efficiencies • Issues • SHMS and HMS PID performance during Fall 2018 (Kaon LT experiment) • SHMS and HMS Calorimeter and HGC response • Pion contamination for standard cuts • Issues
SHMS PID during E12-10-002 (F2) – Spring 2018 Run 2788: SHMS at E’ = -2.1 GeV and theta = 33 deg All, just a delta cut HGC: C4F8O at 1 atm NGC: CO2 at 1 atm Pions will not produce Cherenkov light in neither gas for run 2788 + HGC > 2. + NGC > 2. Etottracknorm
SHMS PID during E12-10-002 (F2) Calorimeter is used here to select electrons mostly NGC NPE HGC NPE HGC mean NPE ~ 16 NGC mean NPE ~ 22
SHMS PID during E12-10-002 (F2): The Inefficiency in the “Middle” of Acceptance Calorimeter is used here to select electrons mostly NGC > 2. NGC > 4. delta • NGC: > 2. / > 4. npe cut leads to at most 2.5 % / 4 % additional inefficiency around delta ~ 0
SHMS PID during E12-10-002 (F2): The Inefficiency in the “Middle” of Acceptance Calorimeter is used here to select electrons mostly NGC > 2. HGC > 1. NGC > 4. HGC > 2. HGC > 4. delta delta • HGC: > 1. / > 2. / > 4. npe cut leads to at most 2 % / 5 % / 13% additional inefficiency around delta ~ 0 • NGC: > 2. / > 4. npe cut leads to at most 2.5 % / 4 % additional inefficiency around delta ~ 0
SHMS PID during E12-10-002 (F2): Pion Contamination (- Polarity) Run 2788: SHMS at E’ = -2.1 GeV and theta = 33 deg (ELREAL trigger) All NGC > 2. NGC < 2. NGC < 2. rescaled Etottracknorm
SHMS PID during E12-10-002 (F2): Pion Contamination (- Polarity) Run 2788: SHMS at E’ = -2.1 GeV and theta = 33 deg (ELREAL trigger) All NGC > 2. NGC < 2. NGC < 2. rescaled Pion contamination Etottracknorm NGC > 2. && Etottracknorm > 0.7 Pion Contamination: ~ 0.3 %
SHMS PID during E12-10-002 (F2): Pion Contamination (- Polarity) Run 2788: SHMS at E’ = -2.1 GeV and theta = 33 deg (ELREAL trigger) All NGC > 2. NGC < 2. NGC < 2. rescaled Pion contamination Etottracknorm Etottracknorm HGC > 1. && Etottracknorm > 0.7 Pion Contamination: ~ 0.5 % NGC > 2. && Etottracknorm > 0.7 Pion Contamination: ~ 0.3 %
SHMS PID during E12-10-002 (F2): Pion Contamination (- Polarity) Run 3028: SHMS at E’ = -4.4 GeV and theta = 25 deg (ELREAL trigger) At 4.4 GeV the pion will produce light in the HGC; pion contamination not estimated with an HGC cut * Disclaimer: the calorimeter distributions need to be better understood; I am applying a minimal set of cuts for this quick estimation of the pion contamination NGC > 2. && Etottracknorm > 0.7 Pion Contamination: ~ 0.4 %
SHMS PID during E12-10-002 (F2): Pion Contamination (+ Polarity) Run 2879: SHMS at +2.7 GeV and theta = 29 deg(ELCLEAN trigger) NGC > 2. && Etottracknorm > 0.7 Pion Contamination: ~ 6.5 %
SHMS PID during E12-10-002 (F2): Pion Contamination (+ Polarity) Run 3089: SHMS at +2.7 GeV and theta = 21 deg(ELCLEAN trigger) Run 2879: SHMS at +2.7 GeV and theta = 29 deg(ELCLEAN trigger) NGC > 2. && Etottracknorm > 0.7 Pion Contamination: ~ 6.5 % NGC > 2. && Etottracknorm > 0.7 Pion Contamination: ~ 5.4 %
SHMS PID during E12-10-002 (F2): Calorimeter Cut Efficiency • A“clean” sample of electrons is selected with a tight NG Cherenkov cut; only those electrons that went through the part of the trigger that did not involve the calorimeter are selected (ELLO without PRLO) • Then the effect of the Etottracknorm > 0.7 cut is tested on this sample • The cut efficiency is obtained per momentum setting by extrapolating to zero pion/electron ratio
SHMS PID during E12-10-002 (F2): Calorimeter Cut Efficiency • A“clean” sample of electrons is selected with a tight NG Cherenkov cut; only those electrons that went through the part of the trigger that did not involve the calorimeter are selected (ELLO without PRLO) • Then the effect of the Etottracknorm > 0.7 cut is tested on this sample • The cut efficiency is obtained per momentum setting by extrapolating to zero pion/electron ratio • Etottracknorm > 0.7 cut efficiency is high E’ = -4 GeV Efficiency at 4 GeV: 99.7 +/- 0.1 E’ = -2.7 GeV Efficiency at 5.1 GeV: 100.5 +/- 1.9 Plots by Fernando Araiza Gonzales (F2 student)
SHMS PID during E12-10-002 (F2): NG Cherenkov Cut Efficiency • A “clean” sample of electrons is selected with a tight calorimeter cut; only those electrons that have made it through the ELHI trigger leg (no Cherenkov input) are used • The NGC npe > 2. cut is tested on this sample • The cut efficiency is obtained by extrapolating to zero pion/electron ratio • The NGC cut efficiency is high Plots by AbishekKarki (EMCstudent)
SHMS PID during E12-10-002 (F2): NG Cherenkov Cut Efficiency • A “clean” sample of electrons is selected with a tight calorimeter cut; only those electrons that have made it through the ELHI trigger leg (no Cherenkov input) are used • The NGC npe > 2. cut is tested on this sample • The cut efficiency is obtained by extrapolating to zero pion/electron ratio • The NGC cut efficiency is high NGC cut efficiency vs x (SHMS E’ = -5.1 GeV, theta = 21 deg) 2-3 % cut inefficiency at x close to zero Plots by AbishekKarki (EMCstudent)
Reminder: The NGC and Its Gassy PMTs PMT 1 PMT 2
Reminder: The NGC and Its Gassy PMTs PMT 1 AdcTdcTimeDiff PMT 2 AdcTdcTimeDiff PMT 3 PMT 2 PMT 1 NGC NPE per PMT • PMT 2 should be replaced; timing cuts help a lot reduce high rate background from the gassy tube but some background still left
Reminder: The NGC Calibration • Ideally the calibration should be done via the identification of the single photoelectron (SPE) peak in the npe distribution • During Spring 2018 the gains of the PMTs were such that the FADC threshold (10 mV) was cutting into the SPE peak • It appeared that increasing the gains to push the SPE peak above 10 mV would lead to saturation of PMTs in beam conditions
Reminder: The NGC Calibration • Ideally the calibration should be done via the identification of the single photoelectron (SPE) peak in the npe distribution • During Spring 2018 the gains of the PMTs were such that the FADC threshold (10 mV) was cutting into the SPE peak • It appeared that increasing the gains to push the SPE peak above 10 mV would lead to saturation of PMTs in beam conditions • The calibration is more painful
Reminder: The NGC Calibration • Ideally the calibration should be done via the identification of the single photoelectron (SPE) peak in the npe distribution • During Spring 2018 the gains of the PMTs were such that the FADC threshold (10 mV) was cutting into the SPE peak • It appeared that increasing the gains to push the SPE peak above 10 mV would lead to saturation of PMTs in beam conditions • When NGC is used again this should be revisited; maybe one can find a way to preserve the SPE in the FADC distribution
SHMS PID during Kaon LT: the HGC • Since Spring 2018 the HGC mirror positions have been slightly adjusted* • We see a more than 30 % increase in the NPE yield in some regions of the acceptance (it is not a calibration artifact, the effect is clear when comparing amplitudes directly for Spring and Fall 2018 runs*) NPE
SHMS PID during Kaon LT: the HGC • Since Spring 2018 the HGC mirror positions have been slightly adjusted* • We see a more than 30 % increase in the NPE yield in some regions of the acceptance (it is not a calibration artifact, the effect is clear when comparing amplitudes directly for Spring and Fall 2018 runs*) • The inefficiency close to delta of zero, however, got worse: from run 4721 it’s a 25% relative* inefficiency for a npe cut > 2. delta NPE NPE delta
SHMS PID during Kaon LT: the HGC Calibration • Calibration done with cosmic data • PMTs are pretty well gain matched PMT 2 PMT 1 PMT 3 PMT 4 Amplitude (mV) Amplitude (mV)
SHMS PID during Kaon LT: the HGC Calibration • Calibration done with cosmic data • The calibration coefficients agree very well with those obtained by the Regina group (fitting a PMT response function to beam data) PMT 1 PMT 2 PMT 3 PMT 4 npe npe
SHMS PID during Kaon LT: Pion Contamination (- Polarity) Run 4721: SHMS at E’ = -2. GeV and theta = 20.5 deg (3/4 trigger) The momentum here was chosen so that the pions do not produce Cherenkov light * Disclaimer: likely overestimated; a more careful study is needed HGC > 2. && Etottracknorm > 0.7 Pion Contamination: ~ 6* %
HMS PID during Kaon LT: the Cherenkov • During Spring 2018 the HMS Cherenkov was a mess
HMS PID during Kaon LT: the Cherenkov • During Spring 2018 the HMS Cherenkov was a mess • Howard replaced the mirrors and readjusted the PMT positions w.r.t. mirrors • The HMS Cherenkov is in great shape now This is now fixed HMS Cherenkov performance during Fall 2018 npe delta delta
HMS PID during Kaon LT: the Cherenkov Calibration • The HMS Cherenkov PMTs have been gain matched and a calibration with cosmics has been done • Average npe yield now from electrons is ~12.5 npe Amplitude (mV)
HMS PID during Kaon LT Run 4721: SHMS at E’ = -3. GeV and theta = 24.4 deg (3/4 trigger) Etottracknorm • At these kinematics the pion contamination is negligible
Summary • The HMS PID are performing well • The SHMS NG and HG Cherenkovs have few issues
SHMS Trigger PID Legs: Fall 2018 elec eff. elreal: 5041 5041 1pion rej. elreal: 58127 205741 3.53951 Details: elec eff. prlo: 5019 5041 0.995636pion rej. prlo: 113502 205741 1.81266 elec eff. prhi: 4992 5041 0.99028pion rej. prhi: 53755 205741 3.82738