US FUNDAMENTALS: B-MODE AND DOPPLER MACHINE SETTINGS

US FUNDAMENTALS: B-MODE AND DOPPLER MACHINE SETTINGS. MSUS BASIC. US: emission and reception of mechanical sound waves with a frequency greater than the human hearing frequency range ( 15-20 MHz).

ifama
Download Presentation

US FUNDAMENTALS: B-MODE AND DOPPLER MACHINE SETTINGS

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. US FUNDAMENTALS: B-MODE AND DOPPLER MACHINE SETTINGS

  2. MSUS BASIC US: emission and reception of mechanical sound waves with a frequency greater than the human hearing frequency range (15-20 MHz) Sound waves produced and captured by piezoelectric material located inside the transducer or probe US: 2-20 MHz

  3. MSUS BASIS The degree of reflection of an ultrasound wave depends on the body tissue being assessed. Each tissue has its own acoustic properties (density and resistance) which determine its ‘acoustic impedance’. Reflections occurs at interfaces between tissues of different acoustic interfaces.

  4. MSUS BASIS Gray scale (GS) ultrasound refers to the black and white images that are generated. Gray scale defines the morphology of a tissue. The whiter a structure appears, the more reflective it is. e.g. bone or calcification will appear white. The darker a structure is, the less reflective it is. As water is the best transmitter of sound, fluid e.g. blood, synovial fluid, cysts generally appear black on the image as sound is not reflected back and registered on the screen. Bone Fluid

  5. MSUS BASIC Doppler technique detects movement of red blood cells in vessels

  6. Machine set up The Clinical Application Specialist from the ultrasound company for your machine will assist in setting up appropriate settings for the assessment of synovitis (and other joint structures) as these will vary from machine to machine

  7. Frequency of transducer • The choice of transducer frequency is dependant on the joint, body habitus and age of a patient • A higher frequency transducer (e.g. 10-18 MHz) is preferable for the assessment of superficial lying anatomical structures e.g. for the MCPJ. Generally > 18MHz used for skin and very superficial structures. • A lower frequency transducer (e.g. 5-10 MHz) is preferably for deeper lying structures e.g. for the shoulder

  8. Footprint of transducer The footprint of the transducer is the surface area of the transducer that is physically in contact with the skin. An image this is acquired should contain all the information required to make a valued assessment For small structures e.g. MCPJ, this may be achieved with a smaller footprint (e.g. hockey stick style transducer) whilst larger structures e.g. shoulder and hip may require a larger footprint to capture the full field of view. Small transducers may be useful if there are deformities or for guiding interventional procedures

  9. Image optimization: General Grey scale Use a minimal number of focal points (on average 2 are used) Ensure focus points are at level of region of interest Adjust the depth so that region of interest at least half way down screen Adjust B-mode or 2D gain to optimize image brightness There are other modifiable parameters on some machines such as dynamic range. This ‘contrast control’ may be adjusted for personal preference but we would advise discussing this (and others) with clinical applications specialist according to your machine

  10. General Considerations Doppler Most scans for the assessment of joint inflammation are performed using Power Doppler or Colour Doppler displays After choosing either option, the Doppler box should be adjusted to cover the whole joint (or specific region of interest). It should encompass the joint proximally and distally and also reach the skin surface.

  11. Recommended settings for colour and power Doppler in rheumatology Doppler frequency Lowest or highest depending on machine Pulse repetition frequency Lowest possible* Colour priority All priority to colour Wall filter Lowest possible* Persistence Lowest possible Gain On the threshold to noise Focus Placed where highest sensitivity is required *Lowest possible where motion artifacts are avoided most of the time. From: Torp-Pedersen ST, Terslev L Settings and artefacts relevant in colour/power Doppler ultrasound in rheumatology. Ann Rheum Dis. 2008;67:143-9. .

  12. Image presentation – Always observe the full extent of joint space (capsule) proximally and distally

  13. Image presentation – Ensure colour box captures whole joint and that box reaches skin surface

  14. Scoring pathology – Always ensure pathology observed in both longitudinal and transverse views