531 likes | 946 Views
RES 131. Pulmonary Function Testing. Laboratory Instrumentation. TWO TYPES OF SPIROMETERS AVAILIBLE IN THE PFT LAB. VOLUME-DISPLACING Sometimes known as “volume collecting” FLOW-SENSING. VOLUME-DISPLACING SPIROMETERS. COLLECTS EXHALED AIR ACTS AS A RESEVOIR FOR INHALED AIR.
E N D
RES 131 Pulmonary Function Testing
Laboratory Instrumentation • TWO TYPES OF SPIROMETERS AVAILIBLE IN THE PFT LAB. • VOLUME-DISPLACING • Sometimes known as “volume collecting” • FLOW-SENSING
VOLUME-DISPLACING SPIROMETERS • COLLECTS EXHALED AIR • ACTS AS A RESEVOIR FOR INHALED AIR.
WATER SEAL SPIROMETER • CONSISTS OF THREE CYLINDERS • ONE CYLINDER WITH OPENING ON TOP • SECOND CYLINDER WITH OPENING ON BOTTOM • THIRD CYLINDER BETWEEN 1ST AND 2ND WITH OPENING ON BOTTOM
BELLOWS SPIROMETER • USUALLY CONSISTS OF : • SOFT FLEXIBLE BELLOWS • LATEX OR RUBBER • RIGID OR SEMI-RIGID PANELS ON EITHER SIDE OF THE BELLOWS
DRY ROLLING SEAL • COMPRISED OF A PISTON IN A CYLINDER • USES A SOFT SILICONE BASED ELASTIC MATERIAL AS A SEALING COMPONENT • OLD SYSTEMS USED A LEATHER ROLLING SEAL
DIAPHRAGM SPIROMETER • CONSISTS OF: • HOUSING • UPPER AND LOWER • DIAPHRAGM • PUSHER PLATE
VOLUME DISPLACEMENT SPIROMETER'S • GENERALLY, BEFORE TESTING STARTS, THESE SPIROMETER'S ARE EMPTY • SOME TYPES ALLOW PRESETTING OR FILLING BEFORE TESTING
DESIRABLE CHARACTERISTICS • DIRECTLY MEASURE VOLUME • LOW COST • EASE OF OPERATION • WATER-SEAL SPIROMETER IS CONSIDERED TO BE THE “GOLD STANDARD”
UNDESIRABLE CHARACTERISTICS • SOME VERY LARGE AND BULKY • LESS PORTABLE • WATER MUST BE CHANGED • LEAKS • MANUAL CALCULATION REQUIRED IF NO MICROPROCESSOR
FLOW-SENSING SPIROMETER'S • VOLUME IS CALCULATED BY MULTIPLYING: • FLOW BY TIME • THIS IS KNOWN AS “INTEGRATION” • REQUIRES A COMPUTER OR MICROPROCESSOR
FOUR TYPES OF FLOW SENSING SPIROMETER'S • PNEUMOTACHOGRAPH • HOT-WIRE ANEMOMETER • TURBINE DEVICE • VORTEX DEVICE
PNEUMOTACHOGRAPH • ALSO KNOWN AS “DIFFERENTIAL PRESSURE DEVICE” • TUBE WITH FIXED RESISTANCE • CONTAINS A BUNDLE OF CAPILLARY TUBES OR FINE MESH SCREEN(S)
PNEUMOTACHOGRAPH • ACCURACY BASED ON FIXED RESISTANCE • CHANGES IN RESISTANCE AFFECT ACCURACY • WATER VAPOR • SECRETIONS • MANY ARE HEATED TO PREVENT CONDENSATION
HOT-WIRE ANEMOMETER • FINE WIRE IN CENTER OF TUBE • AIR MOVING THROUGH THE TUBE COOLS THE WIRE • ELECTRICAL CURRENT IS PASSED THROUGH WIRE TO MAINTAIN TEMPERATURE • CURRENT NEEDED TO MAINTAIN TEMP IS MEASURED
HOT-WIRE ANEMOMETER • UNABLE TO MEASURE TURBULENT FLOWS ACCURATELY • LONG TUBE REQUIRED TO CREATE LAMINAR FLOW • FRAGILE
TURBINE DEVICE • TURBINE INCREASES SPEED WITH INCREASING FLOW • REVOLUTIONS ARE MEASURED • INERTIA CREATES INACCURACY
TURBINE DEVICE • NOT AFFECTED BY: • TURBULENT FLOW • MOISTURE • GAS COMPOSITION
VORTEX DEVICE • STRUTS IN AIRFLOW CAUSE TURBULENCE RESULTING IN A VORTEX • VORTEXES ARE COUNTED
VORTEX DEVICE • NOT SENSITIVE ENOUGH TO MEASURE LOW FLOWS
Flow Sensing Spirometers • DESIRABLE CHARACTERISTICS • SMALLER AND MORE PORTABLE • COMPUTERIZED, NO MANUAL CALCULATIONS • MANY PROVIDE QUICK REFERENCE VALUES • CAN MEASURE BI-DIRECTIONAL FLOW
Flow Sensing Spirometers • UNDESIRABLE CHARACTERISTICS • MORE KNOWLEDGE REQUIRE TO OPERATE COMPUTER • FREQUENT AND CAREFUL CALIBRATION • MOISTURE AND SECRETIONS CAUSE INACCURACY • GAS COMPOSITION CAN AFFECT RESULTS • MAY NOT SENSE LOW FLOWS
CHOOSING A SPIROMETER • RECOMMENDATIONS FOR SPIROMETER'S ARE PUBLISHED BY THE AMERICAN THORACIC SOCIETY • PUBLISHED IN 1979, 1987 AND THE LATEST IN 1995
SPIROMETER'S FOR FORCED EXPIRATORY MANEUVERS • ABLE TO MEASURE VOLUMES OF AT LEAST 8 LITERS • FLOWRATES (INSPIRATORY AND EXPIRATORY) BETWEEN 0 AND 12 LITERS PER SECOND
SPIROMETER'S FOR FORCED EXPIRATORY MANEUVERS • ABLE TO COLLECT OR MEASURE VOLUMES FOR 15 SECONDS • IF THE SPIROMETER IS USED TO MEASURE THE FEV1, THE BACK EXTRAPOLATION METHOD SHOULD BE USED
SPIROMETRY DISPLAYS • TWO WAYS TO DISPLAY THE SPIROGRAM • VOLUME-TIME • FLOW-VOLUME
VOLUME-TIME DISPLAY • VOLUME IS DISPLAYED ON THE Y AXIS (VERTICAL) • TIME IS DISPLAYED ON THE X AXIS (HORIZONTAL)
VOLUME-TIME DISPLAY • USEFUL IN EVALUATING THE LENGTH OF A TEST MANEUVER • ALLOWS ASSESSMENT OF VOLUME PLATEAU DURING THE TERMINAL PORTION OF THE TRIAL
FLOW-VOLUME DISPLAY • FLOW IS DISPLAYED ON THE Y AXIS (VERTICAL) • VOLUME IS DISPLAYED ON THE X AXIS (HORIZONTAL)
FLOW VOLUME DISPLAY • MOST USEFUL IN EVALUATING THE START OF TEST • TYPICALLY, THE GRAPH IS SCALED SO THAT THE FLOW REPRESENTATION IS TWICE THAT OF THE VOLUME
CALIBRATION OF PULMONARY FUNCTION INSTRUMENTATION • Without a conscientiously applied calibration program, pulmonary function instruments may generate erroneous information
CALIBRATION • Calibration of volume collecting and flow sensing SPIROMETER'S require the use of a calibration syringe
Calibration of volume collecting SPIROMETER'S • Patient testing tubing should be attached • Inject the entire calibration syringe volume • Observe the pen or graphic display
Calibration of volume collecting SPIROMETER'S • If the display or line does not travel in a straight line, there is a leak • If no leak is present, the indicated volume should be equal the volume of the calibration syringe within +/-3%
Calibration of volume collecting SPIROMETER'S • Calculation of acceptable calibration error • Multiply the syringe volume by 3% (.03) • 3.0 x .03 = .09 • Add and subtract the calculated value from the known value • 3.0 L + .09 = 3.09 L • 3.0 L - .09 = 2.91 L
Calibration of flow- sensing SPIROMETER'S • This type of spirometer of ten has a menu option for calibration • At least one injection from the calibration syringe should be with a reported value within +/- 3% • Usually several injections are completed
Calibration of flow- sensing SPIROMETER'S • ATS recommends varying the flow rate of the injections • Once at approx. 3 L/sec (1 second) • Once at approx. .05 L/sec (6 seconds) • Once at a rate between the previous two
Calibration • If either type of spirometer cannot meet the accuracy standards, the spirometer may not be used. • Calibration must be completed at least daily or whenever accuracy is questioned
Calibration • ATS standards require the documentation of the calibration • A log book should be kept, documenting the calibration values obtained
Quality Control • Activities used to maintain Quality • Laboratory Standards • Quarterly testing of known subjects with stable pulmonary conditions • Recorder time sweep • Check the accuracy of the recorder by stop watch