1 / 15

Mohammad Peikari Advisor: Dr. Gabor Fichtinger

Characterization of Ultrasound Elevation Beamwidth Artefacts for Brachytherapy Needle Insertion. Mohammad Peikari Advisor: Dr. Gabor Fichtinger. Laboratory for Percutaneous Surgery School of Computing, Queen’s University, Canada. Motivation. Gall Bladder

jasontravis
Download Presentation

Mohammad Peikari Advisor: Dr. Gabor Fichtinger

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Characterization of Ultrasound Elevation BeamwidthArtefacts for Brachytherapy Needle Insertion Mohammad Peikari Advisor: Dr. Gabor Fichtinger Laboratory for Percutaneous SurgerySchool of Computing, Queen’s University, Canada

  2. Motivation Gall Bladder (arrow points to false artifact) • All US-guided procedure suffers from section thickness artifacts • Appearance of anatomy and localization of surgical tools affected • Motivating application is the transrectal ultrasound (TRUS) guided prostate brachytherapy Figure: http://www.gehealthcare.com/caen

  3. Achievements • Nominated for best master research award – IEEE Kingston section 2011 • Journal of Medical Physics 2012 (ISI impact factor = 3.25) • Medical Image Computing and Computer Assisted Intervention (MICCAI) conference 2011 (peer reviewed conference proceedings) • International Society for Optics and Photonics (SPIE) 2011 (nominated for best student paper award) • Patented a variation of the presented device by other members of the group

  4. Prostate Cancer • Second leading cause of cancer related death • Treatment options: • Prostatectomy • external beam radiation • Brachytherapy • Potential advantages of brachytherapy: • Outpatient treatment • Comparable to the other treatment options • Ability to target tumor and avoid healthy tissues • Potential disadvantages of brachytherapy: • Side effects may vary • Highest quality is hard to achieve

  5. Prostate Brachytherapy • Permanent implantation of radioactive seeds under live ultrasound (US) guidance Figure credited to C. Chao from Perk lab

  6. Ultrasound www.en.wikibooks.org and http://www.frca.co.uk

  7. Treatment Planning 100% isodose Margin: 3-6 mm www.oncoprof.net

  8. Ultrasound Guided Needle Insertion 1-Postate with target implant location 2-Needle insertion 3-Needle reaches the target http://www.eecs.berkeley.edu

  9. Treatment Validation and Plan Update E. Dehghan et al. “Prostate Implant Reconstruction from C-arm Images with Motion-Compensated Tomosynthesis”,Medical Physics, Vol. 38(10), pp. 5290 – 5302, 2011.

  10. Procedure http://www.prostatebrachytherapyinfo.net/PCT21.html

  11. Section Thickness Artifacts in TRUS • a) Main beam thickness • b) Side lobe energies Main lobe Side lobe

  12. Objectives • Characterization the ultrasound elevation beamwidth • Generate US beam profile • Compare main beam thickness and side lobe artifacts • Measure needle tip localization offsets • Recommendations to reduce the effects of these artifacts

  13. Prior Work on Beamwidth Artifacts • Goldstein (Ultrasound,1981) • Skolnick (Radiology, 1991) • Compared scan plane and section-thicknesses • Used an inclined surface and a phantom with multiple filaments 1cm apart in a vertical row • Difference compared to proposed method: Needs segmentation of the filaments • Richard (Radiology,1999 ) • Used several inclined surfaces located successively below each other in a phantom • Difference compared to proposed method: more complex phantom, results only at specific positions

  14. Prior Work on Side Lobe Artifacts • Laing (Radiology, 1982) • Illustrated the genesis of side lobe artifacts • Employed round plastic container filled with de-gassed water and a sponge • Compared the effects of main and side lobe artifacts • Barthez (Radiology and Ultrasound, 1997) • Reproduced the artifacts using metallic wires and wooden tongue depressor • Employed all sorts of US transducer • Shape and intensity varied with US transducer type

  15. Beamwidth Measurement 45⁰ 45⁰ Beamwidth Main beam thickness Main and side lobe beams thickness • CD≈Ultrasound beamwith • The US beamwidth is larger when side lobe energies present around the main lobe

More Related