Developments in Radiation Detection Systems

Developments in Radiation Detection Systems. By Alex Rak 03/17/2011. Presentation Outline. About Me Introduction to Scintillation Detectors Scintillation Detector System Mechanical Design Test Procedure Final Results Introduction to Gamma Cameras Gamma Camera System Theory

jasper
Download Presentation

Developments in Radiation Detection Systems

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Developments in Radiation Detection Systems By Alex Rak 03/17/2011

  2. Presentation Outline • About Me • Introduction to Scintillation Detectors • Scintillation Detector System • Mechanical Design • Test Procedure • Final Results • Introduction to Gamma Cameras • Gamma Camera System • Theory • Test Procedure • Results • Summary

  3. About Me Bardeen Engineering Fellowship Candidate Age: 23 Hometown: Burlington, IL Education: • B.S. in Mechanical Engineering from Bradley University • M.S. in Mechanical Engineering from Bradley University Expected in May

  4. About Me Continued Work Experience: • Graduate Researcher, Bradley University (May-Dec. 2010) • Graduate Assistant, Bradley University (Aug.-Dec. 2009) • Drafting Intern, Flexible Steel Lacing Co. (Summer 2007) Other Experience: • President, Sigma Phi Epsilon Fraternity (2006-2007) • Vice-President, Inter-Fraternity Council (2008-2009) • Senator, Student Senate (2006)

  5. Scintillation Detectors 2 Key Components: • Scintillation Crystal • Can be Crystal, Plastic, Liquid, or Glass • Photomultiplier Tube • Contains Photocathode, Several Dynodes, and an Anode

  6. How Scintillation Detectors Work • Scintillation Light is Released when Struck by Ionizing Radiation • Converts Gamma Rays to Visible Blue Light • The Light is Converted into an Electric Pulse by the Photocathode • The Signal is then Amplified in the Photomultiplier Tube by Passing Through a Series of Dynodes Until it Reaches the Anode

  7. Senior Project • Funded by Los Alamos National Laboratory (LANL) • Est. 1943 in Los Alamos, NM 87545 • LANL wanted an Automated Omni-Directional Sensor to Locate Radioactive Material • Contacts at LANL • Dr. Gregory Dale • Dr. Michele DeCroix • Mr. Robert Wheat

  8. System Requirements • Use Standard 3 in. Right Circular Cylinder NaI Detector • Detect and Locate Source Within 10 Minutes • Transportable Pieces Weighing Less than 60 lbs. • Locate the Source Within 1 ft. of Actual Location at a Distance of 5 ft. from the Detector • Scans Continuously Without User Feedback • Scan an Area 90⁰ Left and Right of Center and 60⁰ Down

  9. System Design • Motorized Linear Slide was Used to Move Scintillation Detector in and out of a Collimator • 1.7 in. Thick Lead Cylinder Used for the Collimator • Second and Third Motor Used to Tilt and Rotate the Entire System • 100:1 Gearbox Used with Both Motors • Ball Bearings were Used to Smooth the Movement of the Device

  10. Motor Mount Analysis • Made from A36 Hot Rolled Steel Plate • Yield Strength of 400 MPa • Max Stress – 180 kPa • Max Deformation - 3.3 x 10-4 mm

  11. Scintillation Detector Arm Analysis • Built from ¾ in. OD Mild Steel Tubing • Yield Strength of 500 MPa • Max Stress – 44.8 MPa • Max Deformation – 0.5 mm

  12. Base Analysis • Constructed out of 6061 Aluminum • Yield Strength of 275 MPa • Max Stress – 53.9 MPa • Max Deformation – 0.8 mm

  13. A-Frame Analysis • Modeled with a Truss Analyzer • Built from ¾ in. OD Mild Steel Tubing • Yield Strength of 500 MPa • Max Stress – 1.2 MPa • Buckling Factor of Safety – 2031 Forces: Buckling: ∑Fx = 0 σact= F1-2/A ∑Fy= 0 σallow = Pcrit/A Pcrit= π2EI/(KL)2 n =σact /σallow

  14. Search Algorithm • A C++ Compiler was Used to Determine the Next Step in the Search Algorithm • Based on Concepts of Windowing and Interpolation • There were 4 Stages to the Algorithm • Every Scan was Taken for 30 sec. • The Locations of the Sections, the Positions of the Motors, and the Depth of the Scintillation Detector were Calculated and Set in the C++ Code • Poisson Statistics were Used to Determine if the Scintillation Detector saw a Source of Radiation in a Particular Section

  15. Statistics • Scanning with the Scintillation Detector Yields a Value with Units of Counts/Second • The Background Radiation, or B, with no Source Present was Recorded for the Testing Room • Each Scan Yields a Total Counts, T • The Source Counts/Second, or S, is Found Using the Following Equation S=T-B • The Standard Deviation of the Source Counts of the Initial Scan was Determined Following the Rules of Poisson Statistics σinitial=√(Ti+B)

  16. Statistics Continued • A Threshold of 3 was Chosen to Determine if the Source was Within a Window • This is Represented in the Equation Below S/σinitial≥3 • If the Threshold Condition is met by the Counts in a Particular Window, the Source is Determined to be in that Window • The Scan then Moves to the Next Stage and Continues from the Window with the Source

  17. Stages 1 and 2 • Stage 1 is the Initial Search and Scans the Entire Search Area • By Increasing the Detector’s Depth in the Collimator, its Circular Viewing Area was Decreased • Positioning was Adjusted so the Circular Viewing Area Encompassed the Entire Rectangular Window Stage 2 - Windowing Detector Depth = 1.29”

  18. Stages 3 and 4 Stage 4 - Interpolation Detector Depth = 5.26” Stage 3 - Windowing Detector Depth = 3.59”

  19. Stage 4 Continued • The Interpolation Stage Records the Values in All 4 Quarters and Plugs them into the this Equation • In this Equation, x is the Angle Off the Center Axis of the Scintillation Detector and y is the Counts/Second • The Angle Off the Center Axis is Used as the Radius of a Circle that is Centered in the Middle of Each Quarter • It is Known that the Quarter with the Highest Recorded Counts Contains the Source • This Quarter will have the Smallest Value for the Angle Off the Center Axis and Therefore the Smallest Circle

  20. Final Source Location • The Intersections of the Circles Provide a More Accurate Location • The Black Dots are the Intersections of the Circles and the Red Dot is the Experimental Source Location

  21. Test Setup • A 10 μCi Source of Cesium-137 was Used • 30 Scans were Performed with the Source at Random Locations • 4 of the Sources were Placed Outside of the Required Search Area to Test for False Positives • The Sources were Placed at Different Distances from the Scintillation Detector Ranging from 1.1 to 8 ft. • These Distances Corresponded to the Limits of the Detector

  22. Results • The 4 Sources Placed Outside the Search Area were not Found as Expected • All 26 Sources Placed Within the Area were Found Within the Tolerance Required • The Average Distance from the Experimental Source to the Actual Source was 4.5 in. • The Entire System was Designed to Break Apart into 3 Pieces Each Weighing 54 lbs. • One Full Scan Took About 9.5 Minutes to Run

  23. Scintillation Detector System

  24. Gamma Cameras Components: • Large, Thin Sheet of Scintillation Crystal • Can have a Collimator in Front of the Crystal • Array of Small Photomultiplier Tubes • Complex Electronics to Interpret Signal and Send to Computer

  25. Gamma Cameras Continued • The Size of the Crystal can Vary Depending on the Application • The More Photomultiplier Tubes in the Array, the Better the Spatial Resolution • Position of the Incident Gamma Ray is Calculated Using the Signal Measured in the Surrounding Photomultiplier Tubes

  26. Gamma Camera System Goals • Design a System that can be Easily Integrated with a Standard Gamma Camera • Develop a more Accurate Method for Determining the Location of a Source of Radiation • Design a System that can be Used for Different Radioactive Sources • Eliminate the Need for Moving Parts to Reduce the Chances of Failure

  27. System Theory • By Suspending a Lead Bar in Front of the Gamma Camera in a Known Location, a Portion of the Incident Radiation is Blocked • When the Data Recorded by the Camera is Converted into an Image the Portion that was Partially Blocked will Appear as a Shadow • The Direction of the Origin of the Radiation can be Found from this Image by Using some Transforms and Geometry

  28. Lead Bar Design • Lead Bar was Sized at 0.4 x 2 x 10 in. • 0.4 in. was the Thickness Required to Stop 99% of the Radiation from the 10 μCi Cesium-137 Source Determined Using Lambert’s Law I(x)=I0e-αx where α = linear attenuation coefficient x = length of material and I0 = initial intensity • The Length and Height were Chosen Because they Matched Standard Sizes from the Manufacturer, NucLead

  29. Lead Bar Apparatus • Standard Plumbing Pipes and Fittings were Used to Keep the Costs Low • Thin L-Brackets were Used to Hold the Bar in Place to Minimize the Interference with the Final Image • The Bar was Suspended 2 in. Away from the Camera Screen, with its Longest Side Horizontal, and with its Thinnest Face Oriented Up and Down • This Configuration Provided the Best Image

  30. Apparatus Image • The Metal Fittings in the Apparatus Blocked a Negligible Amount of the Radiation

  31. Gamma Camera Procedure • Measure the Background Radiation in the Test Room with no Source Present with a 600 sec. Scan • Place the Lead Bar in the Apparatus and the Cesium-137 Source in a known Location • Initiate a 600 sec. Scan to Record the Incident Radiation from the Source • Import the Background and Source Scans into the Wolfram Mathematica Code • Import the Mathematica Output File into the MATLAB Code as an Image and Determine Experimental Source Location

  32. Scanning Software Syngo: • Software Installed on Siemens Workstation that is Used to Setup the Type and Length of a Scan • Outputs a List of Energies and Positions of all Gamma Events Siemens e.soft: • Directly Connected to the Gamma Camera • Transmits the Information Recorded by the Camera to the Syngo Workstation

  33. Scanning Hardware Siemens e.cam Gamma Camera: • Non-Production Test Model • 3/8 in. Thick NaI(Tl) Crystal • 16 x 22 in. Viewing Area • Array of 59 Photomultiplier Tubes • 53 with a 3 in. Diameter • 6 with a 2 in. Diameter

  34. Mathematica Code • Originally Developed by Dr. John Engdahl in 2009 • Creates a List of Gamma Events • Subtracts Background Radiation • Removes Markers Placed in the List by e.soft • Eliminates Values Beyond 1 Standard Deviation of the Mean • Outputs a Comma Separated Value File that Can be Opened as a Matrix or Image

  35. MATLAB Code • Performs a Mean and Median Filter to Image − 3 x 3 Filter Kernel • Imports the CSV File as an Image

  36. MATLAB Code Continued • Performs a Hough Transform to Find the Location of the Edges • Finds Edges Using Built-in Canny Edge Finder

  37. Hough Transform • Take all the Points in an Image and Transforms them into Parameter Space • Each Point is Assigned a Line Equation Containing a Sine and Cosine Term xicosθ + yisinθ = ρ • These Lines are Plotted Over a Specified Range of the Parameters ρ and θ • The Intersections of the Sine and Cosine Lines Correspond to Line in Real Space • The Endpoints are Saved as Variables to be Used Later

  38. Other Operations • Saves Values for the Known Position of the Lead Bar • Adjusts the Values of the Endpoints of the Lines so they are Relative to the Center of the Gamma Camera • Runs through Multiple If, Elseif, and Else Loops to Determine which Points Correspond to which Value and what Side of the Screen the Shadow is on • The Hough Transform Assigns the Values by their Order in Parameter Space and not Real Space • This Allows the same Naming System to be Used no Matter where the Shadow is Projected

  39. Source Trajectory • Using the Endpoints Found through the Hough Transform, the Center of the Shadow is Determined • This Point is Connected with a Point at the Center of Mass of the Bar to Create a Line • The Experimental Source Location is at some Distance Along the Line • This Line can be Extended as far as Needed

  40. Sample Result • The Images on Previous Slides came from a Scan with the Source Located at (12”, -4”, 60”) Relative to the Center of the Gamma Camera • After the Analysis was Finished the Experimental Source Trajectory Passed the Actual Location by 11.6” • Need to Take more Scans to see how the Source Location Affects the Result • Gamma Camera was Moved so it Needs to be Powered on and Calibrated

  41. Ideal Test Results • Sketches were Created in Pro-E so Tests could be Performed Using Images with Ideal Edges • 12 Mock Source Locations were Used and Images were Created by Projecting Lines from the Source • The Same Code was Used to Analyze the Ideal Images • The Average Distance Away from the Actual Source Location was 2.4 in.

  42. Example Usage • Shipping Container is Suspected of Containing Radioactive Material • Someone could Search the Entire Container with a Geiger Counter by Hand • Could Take a Long Time and Puts People in Danger • A Van Containing the Gamma Camera System could Park Next to the Container and Scan it • This would Narrow the Search Area Within the Container and Determine the Radiation Level of the Source

  43. Improvements • Modify the Image Processing to get Better Edges from the Real Scan Images • Find an Experimental 3D Point Source Location Instead of a Trajectory through Space • Add Computer Codes to Completely Automate the Process • Output the Average Energy Level of Gamma Rays so the Type of Radioactive Material can be Determined

  44. References • Gamma Cameras, David S. Graff Ph.D. • Attenuation of Radiation in Matter, Peter B. Siegel Ph.D. • The Gamma Camera, UBC Physics & Astronomy • Interaction-Detectors, University of Waterloo • Detectors of Light, UC Davis University of California • Digital Image Processing, Rafael C. Gonzalez, Richard E. Woods

  45. Acknowledgments • Senior Project Team • Keegan Roach • Derek Blunier • Eugene Kim • Dr. John Engdahl • Bradley University • Fermilab • Elaine G. McCluskey • Shelley A. Krivich • Bardeen Fellowship Committee

  46. Summary • About Me • Introduction to Scintillation Detectors • Scintillation Detector System • Mechanical Design • Test Procedure • Final Results • Introduction to Gamma Cameras • Gamma Camera System • Theory • Test Procedure • Results