An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. PACS Thermal Behaviour TestsT. Müller, H. Feuchtgruber, A. Contursi,Cs. Kiss, Sz. Csizmadia PACS Thermal Behaviour

  2. Available Reports • Thermal Behaviour Tests During EQM IMT: SPEC & PHOT , • PACS-ME-TR-067, issue 1.0 • Blue Spectrometer Temperature Sensor Variation – EQM IMT • PACS-ME-TR-066, issue 1.0 • Comparison of the nominal and the PACS/SPIRE parallel mode cooler • Recycling procedures • Cooler Holdtimes of the Photometer Focal Plane Unit, PICC-KO-TR-004, • Draft 0.1 • As part of the PACS ILT report: • Cooler Recycling Procedure of the Photometer Focal Plane Unit • AVM/CQM ILT • Thermal Behaviour Tests during CQM ILT: SPEC & PHOT PACS Thermal Behaviour

  3. PACS EQM Cryostat Environment To be compared with IST and TV/TB Environment PACS Thermal Behaviour

  4. FM IST Cryostat Environment Closed ! (drifting) Tilt may affect Red Ge:Ga strap Full PACS Thermal Behaviour

  5. FM TV/TB Cryostat Environment Tilt may affect Red Ge:Ga strap Empty PACS Thermal Behaviour

  6. In-Flight Cryostat Environment Empty PACS Thermal Behaviour

  7. Thermal Behaviour Test: Spectroscopy (1) • Problems: • He leak • Bugs in the script • Blue detector temperature sensors where not functional • Remarks: • - Two cover heat-up activities took place during the test, • causing some impact on L1 temperatures • - General upward drift of L1 temperatures, probably due to • cryostat operations • - Significant increase of Red detector temperature in second • half of the test probably due to the He-leak effects and • combined with the cryostat operations. • - The internal calibration sources were not switched-on due to • a bug in the command script PACS Thermal Behaviour

  8. Thermal Behaviour Test: Spectroscopy (2) PACS Thermal Behaviour

  9. Thermal Behaviour Test: Spectroscopy (3) PACS Thermal Behaviour

  10. Blue Detector Heating in Spectroscopy Iheater=200µA After Recovery of Temperature Sensor: DECB_TS_2_3 Iheater=100µA Input to thermal shimming for FM PACS Thermal Behaviour

  11. Thermal Behaviour Test: Photometry (1) • Problems: • He leak • Bugs in the script • excessive dissipation by bolometer group 5 (PACS-ME-NCR-163) • Timing FPGA parameters problem DMC (PACS-ME-NCR-159) • DPU CFM generated corrupted data • Remarks: • - Executed 4 times (2 with leak, 1 without, 1 parallel mode) • - Ideally, more time would be required to see stabilization of CSs • The PACS/SPIRE parallel mode worked from the thermal point of view, • at the He-mass flow given during EQM IMT. PACS Thermal Behaviour

  12. Thermal Behaviour Test: Photometry (2) PACS Thermal Behaviour

  13. Thermal Behaviour Test: Photometry (3) without Bias group 5 problem, during parallel mode with Bias group 5 problem PACS-ME-NCR-163 PACS Thermal Behaviour

  14. Cooler hold Time with He-leak End of 2h10m Recycling Procedure PACS Thermal Behaviour

  15. Cooler hold Time without He-leak End of 2h10m Recycling Procedure only little detector operation during cooler hold time ! ~40 hours PACS Thermal Behaviour

  16. Thermal Behaviour Conclusions • Do not run a test with a He-Leak. Consequence: Non-Nominal He- • flow scheme due to main tank leak • Absolute L1 He-flow more than a factor 10 larger than specified for FM • [spec. (IID-A) = 2.1 mg/s ; actual >25 mg/s] • L1 temperatures became stable only towards the end of the test phase • Photometry: bias group 5 problem caused anomalous heat dissipation • Overall thermal behaviour of PACS was within expectations • and comparable to ILT • Cooler hold time too short, even with little bolometer operations: • Recy. Opt. + improved cooler filling; Cover-T ? L1-T ? Straps ? Leak ? • Cryo-cover, L2, L1 and L0 temperatures should be kept constant during • thermal behaviour tests • The malfunction of at least one blue spectroscopy detector could be • recovered to obtain the required FM design information. • Stabilized value of DM_CAL_SRC temperature not visible due to time • constraint • Steep learning curve for implementation of related commanding scripts PACS Thermal Behaviour