Chapter 25. HPLC High-Performance Liquid Chromatography. Typical HPLC. Application Microdialysis. Monitor Aspirin in Blood. Increased Efficiency. In any method the object is to increase mobile phase / stationary phase interaction. Decrease particle size. Better packing – slower flow.

Download Presentation

Chapter 25

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Chapter 25 HPLC High-Performance Liquid Chromatography

  2. Typical HPLC

  3. ApplicationMicrodialysis

  4. Monitor Aspirin in Blood

  5. Increased Efficiency • In any method the object is to increase mobile phase / stationary phase interaction. • Decrease particle size. • Better packing – slower flow.

  6. How Does the Smaller Size Help

  7. Smaller is Better • N ~ 3500 L (cm) / dp (mm) • Smaller particle size leads to • Higher plate number • Higher pressure • Shorter optimum run time

  8. Size Considerations

  9. What is a bar? • A measure of pressure. • One bar = 1.01325 Atm • One bar is 100 000 Newtons/m2 • Under water you gain about one atm for each ten meters in depth.

  10. HPLC Columns

  11. Stationary Phase / Support • Support is the scaffolding that the stationary phase sits on. • Support - Microporous Silica • Solvent can get inside • Surface area of 100’s m2 / gram • Silica degrades above pH 8 so keep pH below. • Special supports have been developed for higher pH

  12. Active SitesGenerally Bad

  13. Active Sites Can Lead to Tailing

  14. We Can Modify the Surface

  15. Bonded Phases • SiOH + ClSi(CH3)2 - R = Si-O-(CH3)2 - R • Change surface polarity • Amino (CH2)3NH2 • Cyano (CH2)3C#N • Diol (CH2)2OCH2CH(OH)CH2OH • Octadecyl (CH2)17CH3 • Octyl (CH2) 7CH3 • Phenyl (CH2)3C6H5

  16. New TechnologyMonolithic Silica Columns

  17. Micrographs

  18. Elution • Competition for the surface. • Which substance has the greater surface affinity – a solvent with a higher affinity will push the analytes along.

  19. The Eluotropic Series

  20. Isocratic and Gradient Elution • Isocratic - same composition • Gradient - composition changes over time • Method Development. • A mixture of compounds. Separated with acetonitrile (B) and aqueous buffer (A) • 1) benzyl alcohol 2) phenol 3) 3’,4’-dimethyloxyacetophenone 4) benzoin 5) ethyl benzoate 6) toluene 7) 2,6-dimethoxytoluene 8) o-methyoxybiphenyl

  21. Too Long - Over Two Hours! • We can do a gradient. Examine when we get the best separations and then change composition over time.

  22. Separation Design

  23. Separation Design

  24. Practical Issues • Solvents • Must be very pure, lack UV absorbing species • Should be filtered • Guard columns should be used to protect column from strong absorbing compounds • Solvents should be degassed – bubbles and oxygen (sparging) • Normal phase solvents should be 50% saturated with water • Gradients in reversed phase can require 10 – 20 column volumes to return to starting conditions • Add 3% 1-propanol to each solvent and you will cut this to 1.5 empty volumes

  25. Reducing Waste Solvent(Save some money – be a hero) • Shorter columns with smaller particles • Switch from 4.6 mm to 3.0/2.0 mm id columns • In isocratic systems – use an electronic recycler.

  26. Quality Assurance • Inject a QC sample each day to insure that you have consistent peak shapes and retention times. • Keep a log of your column performance

  27. Symmetric Band-shape • Asymmetry Factor A/B should rarely be worse than 0.9 to 1.5. • Tailing a bigger issue and fronting generally. • Amines interacting with support active sites. • Add 30 mM TEA • Acidic compounds • Add 30 mM Ammonium Acetate • Or for mixes or unknowns – 30 mM triethylammonium acetate. • Persistent problem – add dimethyloctylamine or dimethyloctylammonium acetate These take a long time to wash on on changing mobile phases.

  28. Other issues • Voids can develop at the column inlet. • Repack with fresh stationary phase to get rid of this problem. (Might want to get new column) • Columns should be washed to get rid of salts and strongly adsorbed compounds • Frits should be cleaned. Back wash or replace • Samples should be dissolved in mobile phase or a weaker solvent.

  29. Overloading • Care should be taken not to overload the injection on the column. • Inject 10x less and see if peaks look any better. • Reversed Phase can deal with 1 to 10 mg sample per gram of silica. (About 10 cm on a 0.46 mm column) • Injection volume • < 15% of the peak volume at baseline

  30. Minimize Dead Volume • Minimize connection tubing • Ensure that fittings are proper matches.

  31. Injection and Detection • Pumps • Piston Type under program control. Up to 400 bar (40 MPa or 6000 psi). Gradients made by proportioning valves.

  32. Injection • Sample loop filling either done manually or by and autoinjector

  33. DetectorsMass or Concentration Types

  34. UV- Vis (Photometric)Variable Wavelength Detector

  35. Photodiode ArrayAll wavelengths at once

  36. Refractive IndexUniversal Detection with major issues.

  37. Evaporative Light Scattering • Solutes less volatile than mobile phase • Light scattered from mass of analyte. • Poor linearity – polynomial calibration • Good with gradients. No solvent front. • Same buffers as used with mass spec. Acetic acid, formic acid and TFA, ammonium acetate, diammonium phosphate, ammonia or TEA.

  38. Electrochemical • Analytes that can be oxidized or reduced • Phenols, aromatic amines, peroxides, mercaptans, ketones, aldehydes etc. • Can be very sensitive but are difficult to work with.