An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. The CATANA Protontherapy Beam Delivery System Cuttone G., Cirrone G.A.P., Lojacono P.A., Lo Nigro S., Raffaele L., Rovelli A.,Sabini M.G., Salamone V. Laboratori Nazionali del Sud (INFN) Via S. Sofia 44 Catania, ITALY

  2. LNS Superconducting Cyclotron is the unique machine in in Italy and South Europe used for protontherapy Treatment of the choroidal melanoma In Italy about 300 new cases for year

  3. PRESENT TREATMENT ROOM • 0 ° respect the switching magnet • 80 meter after extraction • 3 m proton beam line LAYOUT OF LNS

  4. WITH A BARE BUT VERY STABLE TABLE AND A SUPERCONDUCTOR CYCLOTRON SUITABLE FOR62 AMeV PROTON BEAM 1999 OUR HISTORY STARTED THREE YEARS AGO......

  5. FIRST OF ALL WE TOOK IN CONSIDERATION TWO PARAMETER FOR THE DEVELOPMENT OF OUR PROTON BEAM DELIVERY SYSTEM: In order to obtain ranges up to 30 mm E.T. at isocenter in air • ENERGY In order to reach a proton beam irradiation field: In terms of lateral dose distribution in air • OMOGENEITY • At least 30 mm in diameter • Omogeneus within  2.5 % • With lateral penumbra of about 1mm

  6. SIMULATION TOOLKIT : TRIM2000 UNDER NT OS GEANT3.0 UNDER LINUX RED HAT 6.1 OS SINGLE SCATTERER FOIL 300 m Tantalum DOUBLE SCATTERER FOIL 80 m + 20 m Tantalum DOUBLE SCATTERER FOIL WITH CENTRAL STOPPER 15 m + 25 m + 7 mm thick copper beam stopper WE DECIDED FOR A PASSIVE SCATTERER SYSTEM AND STUDIED THREE DIFFERENT EXPERIMENTAL CONFIGUARATIONS

  7. SIMULATION RESULTS

  8. EXPERIMENTAL RESULTS OUR BEAM PROFILE WITH A 50 mm DIAMETER FINAL COLLIMATOR SIMULATION VS. EXPERIMENTAL RESULT

  9. EXPERIMENTAL RESULTS LNS PURE BRAGG PEAK

  10. Modulator & Range shifter Ligth field Scatterer system Monitor chambers Laser

  11. JUST AFTER THE FIRST SCATTERER WE EVALUATED THE PROTON BEAM LATERAL DISTRIBUTION ALONG OUR BEAM LINE

  12. 2. AFTER THE SECOND SCATTERER

  13. 3. AT ISOCENTER WITH A 25 mm COLLIMATOR

  14. IN COLLABORATION WITH CCO WE PRODUCED A PRELIMINARY SET OF MODULATORS It is interesting to note we didn’t find any difference in the experimental results, varing this distance from 15 cm up to 33 cm The actual distance beetwen modulator and range shifter is 33 cm

  15. ACTUALLY THE BEAM LINE IS READY 1999 2002, February

  16. WE ACKNOWLEDGE FOR THEIR SUPPORT AND ENCOURAGEMENT: CCO A. Kacperek M. Sheen PSI E. Egger TRIUMF E. Blackmore MGH B. Gottshalk AND LAST BUT NOT AT LEAST Prof. G. Biti University of Florence For his support in clinical authorization