1 / 28

Supplementary Training Modules on Good Manufacturing Practice

Validation Part 4: QC-related validation. Supplementary Training Modules on Good Manufacturing Practice. Validation. Introduction Why is analytical monitoring necessary? What is the purpose of analytical validation?. Validation. Objectives To introduce the concepts of :

judith
Download Presentation

Supplementary Training Modules on Good Manufacturing Practice

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Validation Part 4:QC-related validation Supplementary Training Modules on Good Manufacturing Practice

  2. Validation Introduction • Why is analytical monitoring necessary? • What is the purpose of analytical validation?

  3. Validation Objectives To introduce the concepts of : • Protocol development • Instrument qualification • Analytical procedure • Extent of validation • Method transfer • Chemical and physical, biological, and microbiological test validation

  4. Validation Validation of analytical procedures requires: • Qualified and calibrated instruments • Documented methods • Reliable reference standards • Qualified analysts • Sample integrity

  5. Validation Validation protocol for analytical method • Statement of purpose and scope • Responsibilities • Documented test method • List of materials and equipment • Procedure for the experiments for each parameter • Statistical analysis • Acceptance criteria for each performance parameter

  6. Validation Qualification of the instrument • Make, model and maker’s manual • Modifications • Installation and operational qualification • Calibration programs • Maintenance schedules

  7. Validation Characteristics of analytical procedures (1) • Accuracy • Precision • Repeatability • Reproducibility

  8. Validation Relationship between accuracy and precision Accurate AND Precise

  9. Validation Characteristics of analytical procedures (2) • Ruggedness • Robustness • Variability caused by: • Day-to-day variations • Analyst-to-analyst • Laboratory-to-laboratory • Instrument-to-instrument • Chromatographic column-to-column • Reagent kit-to-kit • Instability of analytical reagents

  10. Validation Characteristics of analytical procedures (3) • Linearity and range • Specificity • Sensitivity • Limit of detection • Limit of quantitation

  11. Table of values (x,y) x y  # Reference material mg/ml Calculated mg/ml 1 0.0100 0.0101 2 0.0150 0.0145 3 0.0200 0.0210 4 0.0250 0.0260 5 0.0300 0.0294 6 0.0400 0.0410 Validation

  12. Validation Linearity Statistics • Intercept -0.0002 • Limit of Linearity and Range 0.005 – 0.040 mg/mL • Slope 1.0237 • Correlation coefficient • Pearson 0.9978 • Olkin and Pratt 0.9985 • Relative procedure standard deviation 3.4%

  13. Validation LOQ, LOD and SNR • Limit of Quantitation • Limit of Detection • Signal to Noise Ratio Peak BLOQ Peak ALOD noise Baseline

  14. Validation Different classes of analytical tests • Class A: To establish identity • Class B: To detect and quantitate impurities • Class C: To determine quantitatively the concentration • Class D: To assess the characteristics

  15. Characteristic A B quant. B Limit test C D Accuracy X X X* Precision X X X Robustness X X X X X Linearity and range X X X Specificity X X X X X Limit of detection X Limit of quantitation X Validation * A degree of bias may be allowed

  16. Validation Extent of validation • New methods require complete validation • Pharmacopoeial methods require partial validation (or verification) • Significant changes mean partial revalidation • equipment changes • formula changed • changed suppliers of critical reagents

  17. Validation Analytical method transfer • Method transfer protocol and procedure • precision • accuracy • ruggedness • Written and approved specific test method • Proficiency check • Formal acceptance by new laboratory

  18. Validation Chemical laboratory validation requirements (1) • Balances • Chromatography • HPLC, HPTLC, GC, TLC • Dissolution or disintegration apparatus • Karl Fischer moisture determination • Melting, softening or freezing point apparatus • Ovens, refrigerators, incubators

  19. Validation Chemical laboratory validation requirements (2) • pH meter • Polarimeter - optical rotation • Refractometer • Spectrophotometer UV/Vis, IR, FTIR, Raman, AA • Timers • Viscometer • Volumetric equipment

  20. Validation Typical validation of HPCL assay (1) • System suitability (performance check) • system precision • column efficiency • symmetry factor • capacity factor

  21. Validation Typical validation of HPLC assay (2) • Method validation • specificity • accuracy • precision • linearity • robustness

  22. Validation Biological assays • Can be difficult to "validate" • "Validity" on a case by case basis • Strictly adhere to the Biological Testing monographs in pharmacopoeias

  23. Validation Microbiological testing requiring validation • Microbial limit testing • Microbial count • Sterility testing • Preservative effectiveness testing • Environmental monitoring program • Biological testing

  24. Validation Validation of microbial test procedures (1) • Virtually impossible to completely validate test procedures for every microorganism • Neutralize /inactivate inhibitory substances, or dilute • Periodic media challenge • Media QC • Reliable methods

  25. Validation Validation of microbial test procedures (2) • Incubation temperature and time • Media may not grow all microorganisms • Variations in media may affect recovery • Inhibitory disinfectants or preservatives • Sample • procedures • handling, storage, transport

  26. Validation Microbiological viable count method validation (1) • Methods • pour plate / spread plate • membrane filtration • Most Probable Number • Sample size • Test dilution • Inoculation size

  27. Validation Microbiological viable count method validation (2) • Membrane filtration conditions • Incubation conditions • Acceptance criteria

  28. Validation Sterility testing validation requirements • Media growth promotion, sterility, pH • Product validation • Stasis testing • Environmental monitoring • Negative controls • Challenge organisms

More Related