1 / 29

Thijs Wijnands (TS/LEA), Richard Hall Wilton (TS/LEA), Markus Stockner (AB/BI)

BLMs - Radmon – BSCs – BCMs for steering the beams in collision during the engineering run at 450 GeV. Thijs Wijnands (TS/LEA), Richard Hall Wilton (TS/LEA), Markus Stockner (AB/BI). Input from : D. Kramer, C. Pignard, D. McFarlane, A. McPherson.

Download Presentation

Thijs Wijnands (TS/LEA), Richard Hall Wilton (TS/LEA), Markus Stockner (AB/BI)

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. BLMs - Radmon – BSCs – BCMs for steering the beams in collision during the engineering run at 450 GeV Thijs Wijnands (TS/LEA), Richard Hall Wilton (TS/LEA), Markus Stockner (AB/BI) Input from : D. Kramer, C. Pignard, D. McFarlane, A. McPherson Special Thanks : A. Prokofiev, C. Ekström, TSL laboratories Sweden W. Hajdas, R. Brun, PSI Villingen Switzerland

  2. BLM ionization chambers measure energy deposited by radiation in air • On line measurements • Dose rate • Dose • Resolution • Spatial : 4000 devices around rings (also UX) • Temporal : 90 ms – 100 s • Signal : 2.5 pA – 1 mA

  3. RadMon Radiation Monitors energy deposited by radiation in Si • On line measurements • Dose rate, Dose • Hadron flux (h > 20 MeV), fluence • 1 MeV eq. neutron fluence • Resolution • Spatial : 307 devices, 250 in tunnel • Temporal : 100 ms or more • Signals : • Dose, dose rate : 1 rad • Hadron flux (h > 20 MeV) : 1e5 cm-2 s-1 • 1 MeV eq. n fluence : 1e10 n cm-2

  4. Location BLMs and RadMons in tunnel RadMon under Q1 L8 6 x BLMs at Q1 L8

  5. BCM- Beam Condition Monitors energy deposited by radiation in C (PCVD) • On line measurements • MIPS/s • Dose (from degradation) • Resolution • Spatial : BCM1 (8 diamonds) BCM2 (12 diamonds) • Temporal : few ns – 40 ms • Signal : 10 pA or more A. McPherson, R. Hall-Wilton

  6. BSC- Beam Scintillator Counters energy deposited by radiation in plastic • On line measurements • MIPS/s • Dose rate • Resolution (CMS) • Spatial : in front of HFC • Temporal : 3-5 ns ? • Signal : ? Resolution Data to be confirmed …

  7. At 1.8 m : • BCM 1- 4 diamonds Spatial resolution BCMs – CMS • At 10.9m : • BSC1 - scintillators • At 25 m : • RadMon, BLMs IP • At 14.4m : • BCM 2 - 12 diamonds A. McPherson, R. Hall-Wilton … another ~20 RadMon on cavern wall UXC55

  8. Single bunch luminosities Reasonable goal for first collisions would be ~ 4 x 1010 protons/bunch From H. Burkhardt 26/01/07

  9. Nominal dose rate & particle flux in TAS-Q1 region IP1 and IP5, data from N. Mokhov LHC PR 633

  10. ATLAS particle flux • For L = 2.5 x 1027 cm-2s-1@ TAS • dose rate : ~1 rad/s • hadron flux : ~30 hadrons/cm/s M. Schupe

  11. CMS particle flux • For L = 2.5 x 1027 cm-2s-1 @ TAS • dose rate : ~3 rad/s • hadron flux : ~10hadrons/cm2/s M. Huhtinen

  12. Measured response data Proton data : • BLMs 1.8x104 gives 1 pA (60 MeV) • RADMON 1.0x105 gives 1count/s (60 MeV) • BCM 2.2x107 gives 1 pA (24 GeV) • BSC 1 count per proton ? Neutron data : • BLMs 1.6x104 gives 10 pA (174 MeV) • RADMON 5x104 gives 1count/s (174 MeV) • BCM ? • BSC ? Preliminary data !

  13. BLM and RadMon calibration 180 MeV neutrons A.V. Prokofiev, M.B. Chadwick, S.G. Mashnik, N. Olsson, and L.S. Waters. Journal of Nuclear Science and Technology, Supplement 2, pp.112-115 (2002) 1 m long collimator 180 MeV p Device 7Li(p,n)7Be neutron flux : 8.8 x 104 s-1cm-2

  14. Neutron Beam facility at TSL deflector magnet Li target stainless steel shielding RadMon device under test

  15. BLM- RadMon neutron irradiation LHC BLM chamber RADMON SPS BLM chamber D. Kramer, M. Stockner, C. Pignard, T. Wijnands

  16. BLM- RadMon neutron response D. Kramer, M. Stockner, C. Pignard, T. Wijnands

  17. Measuring luminosity at 450 GeV BRAN ?

  18. 450 GeV Commissioning with BSCs • Start with 4x1010 – single bunch – 1 beam • background from beam gas • alignment errors • symmetry U-D and L-R • commission DAQ • repeat for beam pipe 2 • Background low : • collide 2 single bunches • Background high : • increase bunch intensity • collide 2 single bunches • optimize lumi • Increase nbr of bunches • time resolution From Zeus - DESY

  19. Optimizing singe bunch luminosity at 450 Gev • Expressing the separation of the two beams in sigma.: • 1 sigma off in one plane, lumi drops by 22% • 1 sigma off in both planes, lumi drops by 39% • 2 sigma off in both planes, lumi drops by 87% • At least L = 1.1 x 1030 cm-2 s-1 is needed • 2 sigma seems feasible, 1 sigma depends on S/N ratio M. Lamont

  20. Summary • Radiation levels at L = 2.5 1027 cm-2s-1 will be extremely low : • Only BSCs will provide some signal • Commissioning Bran, BLMs, Radmons, BCMs, … • Preliminary background in entire LSS1,5 • Increase first bunch intensity to, say, Np = 8 x 1010 to • Increase S/N ratio in BSCs • Enable first lumi scans • Confirm MC simulations, background • Check S/N ratio from Bran, BLMs, Radmons, BCMs, … • If background sufficiently low : collisions with 2 or more bunches • Open issues to be addressed : • Resolution and response of BSC (CMS and ATLAS) • BSC data exchange rate (5 Hz was proposed) • Optimized layout BLMs, Radmons in LSS (from Q1 downstream)

  21. Space in front of TAS at IP5 – BLM, Radmon Q1 1 x Radmon 6 x BLMs + 6 x spare

  22. Picture gallery

  23. Example : Radiation Monitors in SPS 2006 Bending Magnet M2 line wrong

  24. N W E S Example : Radiation Monitors in CDF (1) antiprotons RADMON radiation monitor

  25. Example : Radiation Monitors in CDF (2) • Integrated values for High Luminosity runs Tevatron 2005

  26. Example : Radiation Monitors in CDF (3) ES seperator accident FNAL 21 Nov 2005 : • One of the electrostatic separators (ES) sparked • Magnet Quench, Vacuum lost • Several weeks downtime for repair ONW INW ISW OSW Time 011 009 167 388 Nov 21 16:12:51 2005 011 009 168 392 Nov 21 16:13:06 2005 corresponding fluence : 4x106 cm-2 (h > 20 MeV)

  27. Neutron response and resolution - BLM LHC BLM chambers : • linear response : 10 pA for 1.6 x 104 n/cm2/s • noise level : 0.5 pA (analog measurement) • resolution : 1 pA For h > 20 MeV D. Kramer, M. Stockner

  28. Neutron response and Resolution - RadMon RadMon monitors : • linear response : 2.6 counts/s @ 2.6 x 106 n/cm2/s • spread : +/- 1 counts per 40 counts • resolution 1 count D. McFarlane, C. Pignard, K. Cwalina, T. Wijnands