1 / 24

The Forefront in Image Processing: PET/Molecular Approaches

The Forefront in Image Processing: PET/Molecular Approaches. Joel Karp University of Pennsylvania. Sixth Annual NCI-Industry Forum Quantitative Oncologic Imaging April 7-8, 2005. Issues of Performance, Image Processing, Quantification.

kaden
Download Presentation

The Forefront in Image Processing: PET/Molecular Approaches

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. The Forefront in Image Processing: PET/Molecular Approaches Joel Karp University of Pennsylvania Sixth Annual NCI-Industry Forum Quantitative Oncologic Imaging April 7-8, 2005

  2. Issues of Performance, Image Processing, Quantification • Performance of current-generation PET scanners Global effects - data correction Local effects - image reconstruction Statistical and count-rate effects • Self-consistency: instrument performs same day-to-day • Cross-consistency: all instruments produce same result • Comparing images (PET and CT) from different patients, different instruments, and different institutes

  3. What is Measured with PET b Random coincidence (~2t. singles2) True coincidence a Scattered coincidence Coincidence? • Yab= Nab(AabTab+Sab+Rab) Record event Trues What is measured Normalization Attenuation Scatter Randoms

  4. Reconstruct image from line-of-response (LOR) projection data If there are N counts in the image, SNR ≠ N / (N)1/2 Signals from Different Voxels are Coupled  Statistical Noise Does Not Obey Counting Statistics

  5. x(0) Iterative reconstruction Data A x = y y Back-projection c(k) Differ ence d(k) Correction for Attenuation, Scatter, Randoms Start here Update Forward projection y(k) x(k) ^ Image

  6. Data Flow DETECTOR Philips Allegro: 616 x 29 crystals RAWVIEW (52 bytes/event) For A,B side of event 26 PMT energies/zone (26 bytes) 100M events = 5200 Mbytes DIGITIZER POSITION CALCULATOR LISTVIEW (8 bytes/event) For A,B side of event 2D position (3 bytes) timestamp Energy (1 byte) TOF (1 byte) 100M events = 800 Mbytes BINNER SINOGRAM (80 Mbytes/frame) R, Phi (295x161x2 = 95 Kbytes) Slice (29^2 = 841) 100M events = 560 Mbytes (7 frames) COMPUTER IMAGE X,Y (128x128 = 16 Kbytes) 100M events = 4 Mbytes (250 slices) reconstruction PACS archive

  7. True Scatter Low Scatter and Randoms High Scatter and Randoms 2D (septa) vs. 3D (no septa) 2D Imaging 3D Imaging S T R Low geometric sensitivity High geometric sensitivity Scatter decreases with high energy threshold - depends on energy resolution

  8. Singles FOV 3D mode Out-of-field activity increases randoms in 3D Problem increases as bore size increases -> less shielding • Randoms ~ 2t. Singles2 • decreases with narrow timing window (2t) • decreases with high energy threshold • estimated (and subtracted) with 2nd (delayed) timing window

  9. Count-rate Performance 10 mCi dose 70-cm long x 20-cm diameter NEMA 2001 (body) Noise Equivalent Count-rate NEC = T/(1+S/T+R/T) NEC ~ SNR2 Philips Allegro

  10. PET Imaging Performance • Spatial resolution -> partial volume effect intrinsic: 4-6 mm reconstructed: >10 mm • Scatter fraction -> noise and bias (after correction) 2D: 10-20% SF 3D: 30-60% SF • Sensitivity and count-rate capability -> statistical quality 25 - 100 kcps or 5 M - 20 Mevents per 3 min frame

  11. A B Scatter Correction Before Scatter correction After Scatter correction Single Scatter - Model based correction Calculate the contribution for an arbitrary scatter point using the Klein-Nishina equation

  12. Attenuation correction with radioisotope transmission scan 20 mCi 137Cs source - 662 keV A = 1 / e -md d = length of chord through tissue m = attenuation coefficient

  13. Attenuation correction for PET Types of transmission images Single photon Cs-137 (662 keV) lower noise 5-10 min scan time some bias lower contrast X-ray (~30-140kVp) no noise 1 min scan time potential for bias high contrast Coincident photon Ge-68/Ga-68 (511 keV) high noise 15-30 min scan time low bias low contrast

  14. Attenuation/Scatter correction No AC or Scatter Corr AC and Scatter Corr University of Pennsylvania PET Center Philips Allegro

  15. Fore-FBP 3D Ramla Fully 3D Iterative Reconstruction improves image quality How about quantification?

  16. 10 mm 37 mm 13 mm 28 mm 22 mm 17 mm NEMA NU2-2001 Image Quality Phantom foam Out-of-field Activity

  17. Partial Volume Effect

  18. NEMA IEC Phantom LOR RAMLA reconstruction Vary relaxation parameter l from 0.00025 (top left) to 0.075 (bottom right)

  19. Contrast vs. Noise Iterative - Ramla Filtered Backprojection (FBP) 1.7 cm hot sphere 2.8 cm cold sphere

  20. Image processingFilters for restoring the spatial frequency components Low (left) - Maximum gain = 2.5 Medium(middle)- Maximum gain = 3.5 High gain (right) - Maximum gain = 4.5 k - parameter describing the Gaussian roll-off fcut - cutoff frequency K, fcut -were bracketed from an analysis of phantom data WF(f) = 1/MTF(f) for f<fcut WF(f) = 1/MTF(fcut) exp-kf 2 for f>fcut

  21. Lesion contrast improves with filtering Profile through the lesion no low med high

  22. 5Mcts 5Mcts TOF 1Mcts TOF 1Mcts no TOF 300 ps TOF Time-of-Flight : list-mode iterative reconstruction 1 Mcts 5 Mcts 10 Mcts

  23. Challenges in comparing images • Spatial resolution differences partial volume - simple (approximate) correction spatial recovery in reconstruction model adds noise • Reconstruction algorithm local convergence depends on algorithm and activity • Accuracy of corrections - randoms, scatter, attenuation depends on patient size and activity distribution • Imaging protocol scan acquisition time and delay post-injection • Quantification - typically based on simple cylinder QC - monitor and correct daily drifts Activity calibration - counts/voxel/min -> nCi/ml Count-rate corrections - dead-time

  24. Challenges in comparing images • Instrumentation in PET is constantly evolving performance of new scanner >> older scanner • Image data size is large - data transfer and archiving PET: 4 Mbyte (with 4 mm3 voxels) CT: 64 Mbyte (with ~1 mm3 voxels) • DICOM quantification (SUV) requires PT format (not NM) manufacturers workstations still most practical • Data analysis tools must be standardized and validated region-of-interest • Image processing behavior must be understood - difficult to standardize

More Related