Cryogen free automated gas chromatography for the measurement of ambient volatile organic compounds. Journal of Chromatography A, 896 (2000) 31-39. Jia-Lin Wang*, Wei-Li Chen, Yen-Hung Lin, Cheng-Hsiung Tsai. experimental. HP6890-Dual traps, FID PLOT for C3-C7 30m × 0.32mm ; 8 μ m.

khan
Download Presentation

Journal of Chromatography A, 896 (2000) 31-39

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Cryogen free automated gas chromatography for the measurement of ambient volatile organic compounds Journal of Chromatography A, 896 (2000) 31-39 Jia-Lin Wang*, Wei-Li Chen, Yen-Hung Lin, Cheng-Hsiung Tsai

  2. experimental HP6890-Dual traps, FID PLOT for C3-C730m × 0.32mm ; 8 μm. DB-1 for C6-C1260m × 0.32mm ; 1 μm. Uncoated column 5 m × 0.25 mm

  3. C3-C7 NMHCs in Taiwan EPA monitoring station 1th 250th 673th

  4. preconcentration system

  5. Minimizing longitudinal diffusion 0.5mm ID × 1.6mm Non- Cooling GC oven 0.5mm ID × 1.6mm Cooling GC oven

  6. 0.5mm ID × 1.6mm Cooling GC oven 0.25mm ID × 0.8mm Non-Cooling GC oven

  7. Reproducibility

  8. Linearity Propane from PLOT column Flow rate = 50ml / min Benzne from PLOT column Flow rate = 50ml / min

  9. (a) CO by NDIR (b) Benzene by automated system

  10. R2> 0.92 R2> 0.25

  11. conclusion This work was to build an automated GC system for measuring VOCs either in the lab or the field without using cryogen in preconcentration and column separation. Two traps were coupled to their designated columns to form suitable pairs to provide adequate resolution without cryogen use.