680 likes | 2.09k Views
Troubleshooting Part I. Objectives:. Understand the four basic steps used to solve troubleshooting problems Identify ECG abnormalities that result from pacing system malfunction and pseudomalfunction Recognize data and resources available to aid in troubleshooting pacing system anomalies
E N D
Objectives: • Understand the four basic steps used to solve troubleshooting problems • Identify ECG abnormalities that result from pacing system malfunction and pseudomalfunction • Recognize data and resources available to aid in troubleshooting pacing system anomalies • Discern pacemaker functions that can affect patient hemodynamics • Describe the causes of pacemaker system anomalies and propose a potential solution
The Steps Used in Troubleshooting Are Simple and Remain the Same for Each Type of Problem • Define the problem • Identify the cause of the problem • Correct the problem • Verify the solution
Potential Problems Identifiable on an ECG Can Generally Be Assigned to Five Categories: • Undersensing • Oversensing • Noncapture • No output • Pseudomalfunction
Undersensing • An intrinsic depolarization that is present, yet not seen or sensed by the pacemaker P-wavenot sensed Atrial Undersensing
Undersensing May Be Caused By: • Inappropriately programmed sensitivity • Lead dislodgment • Lead failure: • Insulation break; conductor fracture • Lead maturation • Change in the native signal
Oversensing • The sensing of an inappropriate signal • Can be physiologic or nonphysiologic ...Though no activity is present Marker channel shows intrinsic activity... Ventricular Oversensing
Oversensing May Be Caused By: • Lead failure • Poor connection at connector block • Exposure to interference
Noncapture is Exhibited By: • No evidence of depolarization after pacing artifact Loss of capture
Noncapture May Be Caused By: • Lead dislodgment • Low output • Lead maturation • Poor connection at connector block • Lead failure
Less Common Causes of Noncapture May Include: • Twiddler’s syndrome • Electrolyte abnormalities – e.g., hyperkalemia • Myocardial infarction • Drug therapy • Battery depletion • Exit block
No Output • Pacemaker artifacts do not appear on the ECG; rate is less than the lower rate Pacing output delivered; no evidence of pacing spike is seen
No Output May Be Caused By: • Poor connection at connector block • Lead failure • Battery depletion • Circuit failure
Pseudomalfunctions Pseudomalfunctions are defined as: • Unusual • Unexpected • Eccentric ECG findings that appear to result from pacemaker malfunction but that represent normal pacemaker function
Pseudomalfunctions May Be Classified Under the Following Categories: • Rate • AV interval/refractory periods • Mode
Rate Changes May Occur Due to Normal Device Operation: • Magnet operation • Timing variations • A-A versus V-V timing • Upper rate behavior • Pseudo-Wenckebach; 2:1 block • Electrical reset • Battery depletion • PMT intervention • Rate response
Magnet Operation • Magnet application causes asynchronous pacing at a designated “magnet” rate
A to A = 1000 ms A to A = 1000 ms AV = 150 AV = 200 V-A = 850 V-A = 800 A to A = 1000 ms A to A = 950 ms A to A vs. V to V Timing A-A Timing Atrial rate is held constant at 60 ppm AV = 150 AV = 200 V-V Timing V-A = 800 V-A = 800 Atrial rate varies with intrinsic ventricular conduction
Upper Rate Behavior • Pseudo-Wenckebach operation will cause a fluctuation in rate
Upper Rate Behavior • 2:1 block operation will cause a drastic drop in rate
Electrical Reset and Battery Depletion • Reset may occur due to exposure to electromagnetic interference (EMI) – e.g., electrocautery, defibrillation, causing reversion to a “back-up” mode • Rate and mode changes will occur • Device can usually be reprogrammed to former parameters • Elective replacement indicators (ERI) can resemble back-up mode • Interrogating device will indicate ERI (“Replace Pacer”)
PMT Intervention • Designed to interrupt a Pacemaker-Mediated Tachycardia
Rate Responsive Pacing • An accelerating or decelerating rate may be perceived as anomalous pacemaker behavior VVIR / 60 / 120
Rate Changes May Occur Due to Therapy-Specific Device Operation • Hysteresis • Rate drop response • Mode switching • Sleep function
Hysteresis • Allows a lower rate between sensed events to occur; paced rate is higher Hysteresis Rate 50 ppm Lower Rate 70 ppm
Rate Drop Response • Delivers pacing at high rate when episodic drop in rate occurs • Pacing therapy indicated for patients with neurocardiogenic syncope
Mode Switching • Device switches from tracking (DDDR) to nontracking (DDIR) mode
Lower Rate Rate Sleep Rate Sleep Function 30 mins. 30 mins. Wake Time Bed Time Time
AV Intervals/Refractory Periods May Appear Anomalous Due to: • Safety pacing • Blanking • Rate-adaptive AV delay • Sensor-varied PVARP • PVC response • Noncompetitive atrial pace (NCAP)
Ventricular Safety Pace Safety Pacing • Designed to prevent inhibition due to “crosstalk” • Delivers a ventricular pace 110 ms after an atrial paced event
Blanking DDDR / 60 / 125 / 200 / 225
Rate-Adaptive AV Delay • AV interval shortens as rate increases PAV delay with no activity: 150 ms PAV with activity: 120 ms
Sensor-Varied PVARP • PVARP will shorten as rate increases Long PVARP with little activity Shorter PVARP with increased activity
PVC Response • PVARP will extend to 400 ms DDD / 60 / 120 PVARP 310 ms
Noncompetitive Atrial Pace (NCAP) • Prevents atrial pacing from occurring too close to relative refractory period, which may trigger atrial arrhythmias
A Change in Pacing Modes May Be Caused By: • Battery depletion indicators (ERI/EOL) • Electrical reset • Mode switching • Noise reversion
Noise Reversion • Sensing occurring during atrial or ventricular refractory periods will restart the refractory period. Continuous refractory sensing is called noise reversion and will: • Cause pacing to occur at the sensor-indicated rate for rate-responsive modes • Cause pacing to occur at the lower rate for non- rate-responsive modes
Note: Adverse patient symptoms may occur as a result of any of the previously mentioned pacing system malfunctions and some pseudomalfunctions.
Management of Patient Symptoms May Be Necessary as a Result of: • Muscle stimulation • Palpitations • Pacemaker syndrome • Shortness of breath due to inappropriate rate response settings
Muscle Stimulation May Be Caused By: • Inappropriate electrode placement near diaphragm or nerve plexus • Break in lead insulation • Unipolar pacing
Palpitations May Manifest From: • Pacemaker syndrome • Pacemaker-Mediated Tachycardia (PMT)
Pacemaker Syndrome “An assortment of symptoms related to the adverse hemodynamic impact from the loss of AV synchrony.”
Pacemaker Syndrome Symptoms include: • Dizziness • Presyncope • Chest tightness • Shortness of breath • Neck pulsations • Apprehension/malaise • Fatigue
Pacemaker Syndrome May Be Caused By: • Loss of capture, sensing • A-V intervals of long duration • Onset of 2:1 block • Single chamber system • Absence of rate increase with exercise
Pacemaker-Mediated Tachycardia (PMT) • A rapid paced rhythm that can occur with atrial tracking pacemakers
PMT is the Result of: • Retrograde conduction • Tracking fast atrial rates (physiologic or non-physiologic)
Retrograde Conduction May Be Caused By: • Loss of A-V synchrony due to: • Loss of sensing/capture • Myopotential sensing • Premature ventricular contraction (PVC) • Magnet application