MUHC Survey Meter Protocol Dr. Robert Corns Department of Radiation Oncology The Sir Mortimer B. Davis Jewish General Hospital
Radiation Safety • Availability of calibrated gamma survey meters legally required • Room and source surveys • License requirement • Patient survey • Contamination control • Emergencies
Do It Yourself? • Why bother developing a protocol? • External services available • $200 + shipping and handling • 10+ meters used in Medical Physics and Radiation Oncology • > $2000 per year
Regulatory Document R117(It’s the law) • Describes minimum requirements for survey meter calibrations • Pre-calibration checks • Known calibration sources (NIST) • Requirements on measurement setup • Records CNSC site at www.nuclearsafety.gc.ca
Calibration Goals • Annual calibration • Calibrate scales up to 10 mSv/hr (1 R/hr) • Typically 3 or 4 for scales per meter • Check scales beyond 10 mSv/hr • Scales ± 20% of true value • Measure at 25% and 75% of each scale’s maximum
d02 e–µd X = kA d2 But How to Do It? • Relate source activity to exposure rate in air • Combination of activities and distances to produce all desired exposure rates • Problem: Does not include scatter effects
X = Xprimary + Xscatter X ≈ Xscatter The Shadow Knows Shadow Block Technique
1 Cs-137 Yield 0 0 200 400 600 Energy (keV) 1 Ir-192 Yield 0 800 0 200 400 600 Energy (keV) A Source by Any Other Name… • Cs-137 • 3 mCi & 30 mCi • Simple spectrum • Ir-192 • 4 – 10 Ci • Complex spectrum • Well chamber • NIST traceable
1.2 Response 1.0 0.8 10 1000 100 Energy (keV) Energy Crisis? • Meters can have an energy dependence • Potential differences using Cs-137 and Ir-192 • Cs-137 used for low and intermediate scales • Possible to cross calibrate • Ir-192 at a large distance against the 30 mCi Cs-137 at a short distance • Reasonable agreement • Some meters are used to monitor Ir-192
425 149.9 Setup Distances
Optical Bench d dob
d??? Chamber Source X Marks the Spot! X • Where is d measured from? • A 2 cm chamber 40 cm from a source has a 10% gradient in the exposure rate • Most meters have their centers marked
Safety First! • Cs-137 needles are handled manually • Upto 8 mR/hr at 1 m • 1/2 hour exposure time per session • < 0.04 mSv body dose per session • Ir-192 source handled remotely • Upto 2 to 4 R/hr at 1 m • Meters read remotely • Camera or computer interface
Summary I • Annual calibration with a goal of ± 20% • 2 Cs-137 needle sources and an Ir-192 source used to calibrate • Exposure rate changed via source choice and distance • Shadow block method removes scatter
Summary II • Measured exposure adjusted to agree with calculated exposure • Nist traceable sources • Radiation Safety • Records and certificates