1 / 12

Release of WinRiver II Version 2.03

Release of WinRiver II Version 2.03. OSW Webinar: November 25 and December 2, 2008 Mike Rehmel. Topics Covered. WinRiver II Version 2.03 Improved Stability Change in Side-lobe Cutoff Other Fixes and Changes SxS Changes All users should upgrade. Improved Stability.

laszlo
Download Presentation

Release of WinRiver II Version 2.03

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Release of WinRiver II Version 2.03 OSW Webinar: November 25 and December 2, 2008 Mike Rehmel

  2. Topics Covered WinRiver II Version 2.03 • Improved Stability • Change in Side-lobe Cutoff • Other Fixes and Changes • SxS Changes • All users should upgrade

  3. Improved Stability • Some computers were very unstable with previous versions • Significant effort was placed on improve stability in this release • 2.03 has greatly improved in all cases • If after upgrading to 2.03 you still have any crashes let us know!

  4. Change in Side-Lobe Cutoff • Old side lobe was calculated as 6% of the mean beam depth for each ensemble • If one beam struck something, it was possible for that beam to report a velocity past the bottom and this velocity be used for the bin velocities below the depth of the shallowest beam

  5. Change in Side-Lobe Cutoff • Side lobe is now calculated from 6% of the shallowest beam

  6. Change in Side-Lobe Cutoff • New side lobe cutoff off may result in • Larger bottom Q and smaller measured Q • Fewer good bins near edges Old Side Lobe New Side Lobe

  7. Change in Side-Lobe • Overall affect on total discharge • Reviewed 70 measurements so for • Mean depth range: .65 to 13 ft • Q range: 7.5 to 18100 cfs • Average change in discharge < 1% (Median + 0.3%) • 3 out of 70 measurements had > 5% change in Q and had mean difference of beam depths for individual ensembles between 17 and 32% • Largest changes in Q likely in measurements with few bins per ensemble with uneven bottoms (depths varying on average > 15% per ensemble) • Reprocessing old measurements currently not recommended. If you have measurements with significant depth variations (in individual ensembles) contact OSW • Additional guidance will be provided in an upcoming memo

  8. Other Fixes and Changes • Enabled diagnostic test for StreamPro • Lost ensembles labeled correctly now • Changed locking mechanism(no longer file permission) • Fixed total Q not including ending edge in composite tabular during data collection • Fixed lock-up in wizard with certain drafts • Fixed counting Pass-Fail in RGTest • Changed default filename prefix to station number

  9. Other Fixes and Changes • Fixed Q Measurement Summary • Sdev, incorrect when in English units • %Bad Ens label • Width label • WT Pings label • Remarks not wrapping correctly • Improved ability to collect data if crash occurs • Fixed problem with raw data files being recorded in multiple directories • Fixed multiple ASCII-out issues

  10. SxS Mode Changes • Can now set ADCP draft for each section • Can edit section location in playback • Added new velocity magnitude graph for evaluating extrapolation • Can mark section not used in playback • Can use screening thresholds • Can set extrapolation per section (only ice, bidirectional, normal, wind) • Ability to reload .mmt file and collect data • Has not been tested by OSW

  11. Summary • Much improved Stability • New side lobe cutoff may result in changes in discharge • Should be small < %1 for most measurements • Shallow cross sections with uneven depths (avg > 15% variation in beam depths with an ensemble) may have larger changes in Q. • Other minor issues fixed • SxS has been modified to include many OSW recommendations, but has not been fully tested • All users should upgrade

  12. Questions? WinRiver II version 2.03 available at: • http://hydroacoustics.usgs.gov • http://rdinstruments.com

More Related