ICP-MS. ICP – Inductively Coupled Plasma MS – Mass Spectrometry. Perkin-Elmer Elan 6100 DRC. Thermo-Finnegan Element 2. Quad. Element. ICP-MS outline. Introduction Instrumentation Sample introduction Ionization in plasma Mass separation and ion detection Analytical procedures

Download Presentation

ICP-MS

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. ICP-MS ICP – Inductively Coupled Plasma MS – Mass Spectrometry Perkin-Elmer Elan 6100 DRC Thermo-Finnegan Element 2 Quad Element ICP-MS Training - Cees-Jan De Hoog

  2. ICP-MS outline • Introduction • Instrumentation • Sample introduction • Ionization in plasma • Mass separation and ion detection • Analytical procedures • Calibration • Interference corrections • Detection limits • Tuning • Data reduction • Sample preparation • Lab safety and etiquette ICP-MS Training - Cees-Jan De Hoog

  3. ICP-MS Training - Cees-Jan De Hoog

  4. Instrumentation Quad low vacuum ~3x10-3 bar intermediate vacuum ~1x10-7 bar high vacuum ~5x10-10 bar ion beam ICP-MS Training - Cees-Jan De Hoog

  5. Element Sample introduction ICP-MS Training - Cees-Jan De Hoog

  6. Sample introduction: fluids Samples are introduced with nebulizer which aspirates sample with high velocity Ar forming a fine mist Peristaltic pump Sample solution Aerosol passes into spray chamber where large droplets are removed via a drain. Only 2% of original mist enters the spray chamber. Processes yield droplets small enough to be vaporized in the plasma torch ICP-MS Training - Cees-Jan De Hoog

  7. Sample introduction: solids • Alternative sample introduction: laser ablation resolution: 25-300 µm depth profiling: ~1 µm/s d.l.s: 1ppm-1ppb • Some applications: • Trace elements in minerals • Growth zones in shells • Forensic and archeologic • ‘fingerprinting’ ICP-MS Training - Cees-Jan De Hoog

  8. Argon plasma generates ions ICP-MS Training - Cees-Jan De Hoog

  9. Ionization in plasma <15% ionization 15-30% ionization 30-70% ionization 25 He 70-90% ionization >90% ionization Ne 20 Ar F 15 Kr Cl Xe Hg ionization potential (eV) Sr As Ba 10 Se 5 K Rb Cs 0 0 20 40 60 80 100 atomic number (z) first second ICP-MS Training - Cees-Jan De Hoog

  10. Ion extraction interface ICP-MS Training - Cees-Jan De Hoog

  11. Mass separation and detection Quadrupole works as a mass filter Separates ions based on mass/charge ratio Quad Typical measurement time per mass unit: 10-20 ms Analyzes more than 40 elements in less than 1 second! ICP-MS Training - Cees-Jan De Hoog

  12. Mass separation and detection Magnetic sector • Magnet works also as a mass filter • Allows higher resolution Element ICP-MS Training - Cees-Jan De Hoog

  13. Isotopes - Isobars What we want to know What we measure ICP-MS Training - Cees-Jan De Hoog

  14. ICP-MS spectrum (low resolution) 55Mn 63Cu 79Br 66Zn 65Cu 66Zn 58Ni 68Zn 60Ni ICP-MS Training - Cees-Jan De Hoog

  15. Isobaric interferences • Isobaric interferences 40Ar+ on 40Ca+ (measure 44Ca+, only 2.1% abundance) 204Hg+ and 204Pb+ • Polyatomic interferences (oxides and archides) 40Ar16O+ on 56Fe+ (measure 57Fe+, only 2.2% abundance) 40Ar35Cl+ on 75As+ (monoisotopic) 40Ar23Na+ on 63Cu+ (measure 65Cu+,but watch 49Ti16O) 137Ba16O+ on 153Eu+ and 135Ba16O+ on 151Eu+ • Double-charged ions 90Zr++ on 45Sc+, 14N2+ on 7Li+ • Matrix dependent • Most elements have relatively interference-free isotopes (but they might be low in abundance) ICP-MS Training - Cees-Jan De Hoog

  16. Low vs. higher mass resolution Quad Element (MR setting) 56Fe + 40Ar16O+ 56Fe 40Ar16O+ 55.92 55.98 55 56 57 55.95 ICP-MS Training - Cees-Jan De Hoog

  17. Dry vs. wet plasma 56Fe 16O40Ar ‘Wet’ ‘Dry’ 16O40Ar 56Fe The Element has the option of a desolvator (called the Aridus). It is a spray chamber at high T (110°C) and has a membrane to separate water molecules from the sample. Reduces oxides more than 10x, increases sensitivity 5-10x. Drawbacks: increased memory effects, more difficult to tune, unreliable for volatile elements and As, Si. ICP-MS Training - Cees-Jan De Hoog

  18. Blank substraction • Even pure water gives a signal, so we need to correct raw data • Most significant for elements z<82 • Mostly results from molecular species from Ar gas and sample solution, e.g., ArO+, ArH+, ArOH+, ArC, CO+, CO2+ etc. • Match blank with sample matrix (‘sample without sample’) • Samples dissolved in 2% HNO3, then we use 2% HNO3 • Seawater: distilled water with 3.5% NaCl (and then dilute!) • Sequential extractions: the various extraction agents (e.g., HAc, NaAc, MgCl2, etc.) • Also corrects from small contributions from material deposited in sample introduction system (tubing, spray chamber, torch, cones) ICP-MS Training - Cees-Jan De Hoog

  19. Quantification • Correction for signal drift and signal suppression because of sample matrix: Internal Standardization • Blank correction • Calculation of element concentrations (external standardization) • Isobaric interference (overlap) correction • Data quality check (reference standards, duplicates) ICP-MS Training - Cees-Jan De Hoog

  20. Signal Drift One solution is to run standards often. But this is time-consuming, and what if you have many standards for many different elements? ICP-MS Training - Cees-Jan De Hoog

  21. Internal Standardization The assumption is that all elements behave similarly to In (1+2 from previous slide) • Also corrects for signal suppression due to matrix effects • Common internal standards: 115In, 185Re (45Sc, 6Li, 209Bi) ICP-MS Training - Cees-Jan De Hoog

  22. External Standardization blank level We mostly use multi-element standards: nitric acid solutions that contain 10-20 elements Typical precision: 5% ICP-MS Training - Cees-Jan De Hoog

  23. Other method 1: Standard Addition • Useful from samples with strong or unknown matrix • A spike with known concentration is added to sample • Precision similar to ES, but more accurate • Drawbacks • approximate concentration needs to be known • Labour intensive ICP-MS Training - Cees-Jan De Hoog

  24. Other method 2: Isotope Dilution • Addition of enriched isotope spike to sample • For ultimate precision and accuracy (better than 1%) • No internal standard necessary • Drawbacks • Approximate concentration needs to be known • Does not work for mono-isotopic elements • Labour-intensive and spikes often expensive natural ratio 2 1 2 1 ICP-MS Training - Cees-Jan De Hoog

  25. Elemental or isotope ratios • When you want to know the ratio of elements relative to one other element • Mg/Ca, Sr/Ca or other elements/Ca in forams, speleothems • When you want to do isotope ratios • Pb isotopes 207Pb/206Pb and 208Pb/206Pb • High precision needed • Sample bracketing (standard – sample – standard) • Preferably with standard composition close to samples • Standard and samples with similar concentrations • The ratio element or isotope serves as internal standard, does not need to be added ICP-MS Training - Cees-Jan De Hoog

  26. Detection limits • L.O.D. limit of detection (3x s.d. of the blank) • Measure a blank sample 6-10 times during run (e.g., 10000 cps for isotope X) • Calculate standard deviation (e.g., 150) • L.O.D. is 3x150 = 450 cps • Calculate cps/conc factor from standard for X (e.g., 10000 cps/ppb X) • L.O.D. is 450 / 10000 = 0.045 ppb ICP-MS Training - Cees-Jan De Hoog

  27. ICP-MS detection limits better on Element than Quad ICP-MS Training - Cees-Jan De Hoog

  28. Reference standards + blanks • Matrix often more complex than calibration standards • Unexpected interferences Measure something with known composition and similar matrix (Reference standards) • Sample digestion procedures involves many steps of adding acids, drying, etc., each will introduce some contamination Measure something without sample that has gone through whole dissolution procedure (digestion blanks) ICP-MS Training - Cees-Jan De Hoog

  29. Data processing • (SHOW EXCEL SHEET) • Normalize to internal standard (check for outliers which may indicate spiking errors • Calculate calibration line (Excel ‘slope’ function) from normalized intensities on standards and concentrations • Calculate concentrations of calibration blank + standards and check if ok • Calculate concentrations of samples incl. digestion blanks (include dilution factors) • Substract digestion blanks from samples • Correct for isobaric overlap (interferences) if measured • Evaluate precision and accuracy with duplicates and reference standards ICP-MS Training - Cees-Jan De Hoog

  30. Internal precision Precision on sample during analysis External precision Precision on several digestions of sample Accuracy Result on reference standard RSD vs. RSE RSE (=RSD/√n) improves with number of repeats, RSD does not Significant digits Should reflect precision E.g., rsd=1%: 0.304 (±0.003) rsd=10%: 0.30(±0.03) Analytical accuracy and precision High accuracy Low precision Low accuracy High precision ICP-MS Training - Cees-Jan De Hoog

  31. Instrument tuning • Torch position (x-y) if cones have been cleaned • Nebulizer gas flow (most important variable in wet plasma) • on Quad: dependent on autosampler that is used (0.9-1.1 L/min) • on Element: varies little (1.05-1.15 L/min) • Aridus gas flows (if used, Element only) Ar (2-2.5 L/min) +N2 (3-9 mL/min) • Check intensities • Quad: 115In > 20 000 • Element: 115In > 3 000 000 (Aridus) or > 400 000 (wet plasma) • Check oxide interferences (CeO+/Ce+) • <3% in wet plasma, <0.5% on Aridus (Element only) • Check doubly-charged ions (Ba2+/Ba+) • <3% in wet plasma, <6% on Aridus (Element only) ICP-MS Training - Cees-Jan De Hoog

  32. Optimizing the nebulizer gas flow Called ‘plasma gas’ on Element ICP-MS Training - Cees-Jan De Hoog

  33. Methods and Sequence/Batch • Software on Quad and Element are different, but principles are the same • Make a method • Select isotopes you want from a periodic table • Decide how long you want to measure each isotope • Decide how many repeats or runs • Set up a sequence (Element) or batch (Quad) • Put all blanks, standards, samples in a list, and indicate their position in the autosampler • Define which method you want to use • Indicate uptake and rinse times • Start! ICP-MS Training - Cees-Jan De Hoog

  34. Lab work, cleanliness & safety • Keep a lab book! • A record of sample handling • If something goes wrong you can track back what you did • Work clean • Wear gloves when handling samples • Clean equipment, vials, pipette tips, etc. before use (acid leach) • Lab coat (full body) in clean labs, incl. hood • Keep open samples in fume hoods if possible • Work safe • Very careful with strong acids and especially HF • Wear gloves and protective clothing • Clean up after use! ICP-MS Training - Cees-Jan De Hoog

  35. Check list • SAMPLE PREPARATION • Decide how to dissolve your samples • Keep a lab book • Work clean & safe • include digestion blanks + (if possible) reference standards • SAMPLE ANALYSIS • Book time in advance (a week at least, check with me) • Prepare samples and standards for analysis • Are the samples properly diluted? Keep signals <100 ppm! • Set up sample run in ICP-MS software • Tune instrument, start run • Clean up after use • Data processing and evaluation ICP-MS Training - Cees-Jan De Hoog