570 likes | 908 Views
Evitherm September 15, 2003. Transient Plane Source Techniques for measuring Thermal Conductivity of Various Materials. Dr. Lars H älldahl Hot Disk AB. Hot Disk … or Transient Plane Source, TPS… or Gustafsson Probe… or G-Probe. The technique is based around :. -A unique sensor.
E N D
Evitherm September 15, 2003 Transient Plane Source Techniques for measuring Thermal Conductivity of Various Materials Dr. Lars Hälldahl Hot Disk AB
Hot Disk… or Transient Plane Source, TPS… or Gustafsson Probe… or G-Probe
The technique is based around: -A unique sensor • A new mathematical model for • thermal conductivity A Transient Plane Source in infinite material - Special electronics to collect data Designed for accuracy and precision
The SensorA double spiral made of Nickel metal foil, cladded by Kapton or Mica to provide electrical insulation from the sample. The diameter varies from 1 mm up to 60 mm. It connects to the electronics, which sends out a current during a selected time, while the voltage drop is recorded See in the following four slides the historical development of Hot Disk from Hot Wire – Hot Strip !
Hot Wire About 1960-ties Metal wire in contact with the sample OK for liquids, but not for solids contact resistance problems…
Hot Strip Development of Hot Wire the wire ….flattened out better contact on sample
From a Strip… …..to a spiral covers larger area on smaller sample
So….the story is … from Hot Wire to Hot strip and next step… Transient Plane Source This is Hot Disk
Sensor embedded in the sample. Heat can spread in all directions
A transient recording Power Send in power.. Time Voltage =resistance =temperature ….and measure at the same time Time For good conductors:high power, short time For good insulators: low power, long time
Heat dissipation in infinite sample Ideally, the sensor is surrounded by infinite sample in all direction. Can’t see the edge, the temperature increase doesn’t reach the edges. Both Conductivity and Diffusivity are measured, and from them Heat Capacity is calculated
Parameters of importance: POWER, TIME SENSOR SIZE Sample Sensor Sample The relation between Diffusivity, Time for measurement and radius of sensor is called Total to Characteristic Time x t r x r This value between 0.3 – 1.0 satisfies the basic assumption done for the Heat Conductivity equation: A Plane Source in Infinite Material =1
Variants of the TPS method By varying sensor size, power and time in the basic method, thermal conductivity over more than 4 orders of magnitude can be covered. To cover more applications, some variants have been developed……
Thin Film Method Samples like paper, textile, cloth, polymer films, etc. Thickness from ~ 10μm ~ 500μm Conductivity from ~ 0.01W/mK ~ 10W/mK Background material Sample Sensor Sample Background material Background material is good conductor, experiment parameters as if this is the sample: high power, short time
Thin Film often uses a square, naked Ni-sensor which is most sensitive, but also cladded sensors can be used
Slab Method For good conductors, >10 W/mK - Ceramic, metals, Si-wafer, SiC. Etc. Insulator Sample Sensor Sample Insulator Relation between Sample Thickness/Sensor radius 0.03 < t/R < 0.75 - Different radii to match different samples
Developing Slab Method -Air is very insulating
Anisotropic samples Axial direction Conductivity Diffusivity Radial direction Conductvity Diffusivity To solve the heat conductivity equation in this case: Heat Capacity Cp must be known!
Anisotropic sampels Tensile testing bars, polymer matrix with fiber reinforcement in the plan. Thickness 2 mm.
χmm2/s λ W/mK λradial W/mK χradial mm2/s λaxial W/mK χaxial mm2/s
Single sided method If you do not want to cut your precious sample… put the sensor on the surface… …..and back it up with a known sample. Tell the software the properties of the known, and run the analysis as before. In principal, the more similar, the better
In Transient Methods, like Hot Disk, the contact resistance does not cause any problem. This is not the case with Steady State Methods, where this is a limitation for the use. The following slides explains how Contact resistance is dealt with, and gives an illustrative example.
Ideally, the sensor is surrounded by infinite sample in all direction. Contact area Temp C Cut off here! Time s
Ideally, the sensor is surrounded by infinite sample in all direction…. If this should occur, the later part of the raw data are removed Cut off here! Temp C Time s
Measuring coated and uncoated ceramics Varistors made of ZnO ceramics surface coated by Al-metal, 60μm thick uncoated surface Measuring with same conditions – sensor, time, power,temperature
It seems like the Al-metal coating was insulating? A metal like Al ? In this case the sprayed metal was very porous. The conductivity was about 0.5 W/mK 60μm of porous metal added to the kapton coating increases the temperture about 4 times!
Some results GreenMetal Coated W/mK std % W/mk std% Sensor radius 6.675 mm 23 C 24.70 (0.2)24.69 (0.8) Mica sensor 300 C 12.67 (2.2)13.36 (0.8) Mica sensor Sensor radius 14.65 mm 23 C 24.71 (0.8)24.88 (1.0) Kapton sensor 300 C 12.19 (1.3)13.08 (0.4) Mica sensor
Measuring liquids Put it in a heating/cooling bath, and run at different temperatures -vertical sensor -low power -short time …to avoid convection - liquids with a viscosity as low as water and methanol can be measured
Pure water: Temperature Time Conductivity std Diffusivity std Heat Capacity std C s W/mK % mm2/s % MJ/m3K % 0 5 0.538 0.4 0.151 1.7 3.56 1.3 0 2.5 0.530 0.8 0.146 1.9 3.64 1 25 see A 2.5 0.619 2.4 0.167 6.5 3.71 3.9 25 see B 2.5 0.637 1.5 0.180 3. 7 3.55 2.3 40 2.5 0.711 0.7 0.207 1.6 3.45 0.9 80 1 0.904 2.9 0.280 6.8 3.24 4.1
Muffle furnace from RT to 700 C Measuring at high temperatures Sample Mica cladded sensor
How accurate is Hot Disk ? An article by professor Torbjorn Log* compares results from 5 different materials, measured with different techniques, with Hot Disk measurements. *T.Log, S.E.Gustafsson, Transient Plane Source (TPS) Technique for Measuring Thermal Transport Properties of Building Materials, Fire and Materials 19, 43-49 (1995)
These are the samples…….. Almost 4 orders of magnitude …with properties found with different techniques.
Results for diffusivity. Small deviations – but who is right?
Measuring on a NIST Standard Extruded Polystyrene, cert. No 1453 Values given for a range of density and temperatures for a thickness of 13.4mm Density from 38 kg/m3 to 46 kg/m3 at 295 K, the given Thermal Resistance (m2K/W), converted to Thermal Conductivity by using the thickness gives a range 0.0328-0.0338 W/mK
100 mm Density 43.8 kg/m3 40.4 kg/m3 13.4 mm 4 possible combinations to place the sensor Side combinations Results (W/mK) Std % 1-2 0.03394 0.12 2-1 0.03379 0.04 2-2 0.03451 0.08 1-1 0.03346 0.23
So….what Hot Disk can do in Evitherm.. measure most materials from 0.005 to 500 W/mK.. develop new applications… provide consultancy, measurements and equipment, as well as training, courses, seminars In Uppsala – development electronics, manufacturing, applications, testing service In Göteborg – development sensors/software applications, testing
That’s all folks ! Thank you for your attention.
Metal insert Vacuum feed through Sample holder and sensor
Outer lid with vacuum connection, gas valves, pressure valve and cables