An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. First SIM² - WORKSHOPThermal analysis (TGA/DSC):practical aspectsTGA – Thermogravimetric Analysis DSC – Differential Scanning Calorimetry Jacek Chwast jacek.chwast@ees.kuleuven.be Geo-Instituut 23.01.2013

  2. Measurement conditions Netzsch STA 409 PC Luxx®: • T [25 – 1000 °C] • heating rate 10 °C/min • N2 atmosphere (60 ml/min) • alumina crucibles • simultaneous measurement of mass change (TG) and heat flow (DSC) • 1 analysis (25100025 °C) = 3 hrs furnace microbalance

  3. Measurement conditions furnace Netzsch STA 409 PC Luxx®: • T [25 – 1000 °C] • heating rate 10 °C/min • N2 atmosphere (60 ml/min) • alumina crucibles • simultaneous measurement of mass change (TG) and heat flow (DSC) • 1 analysis (25100025 °C) = 3 hrs sample + reference microbalance

  4. Thermal analysis (TGA/DSC) in a nutshell Examples of thermal reactions resulting in mass change, measured by TG analysis: • loss of free water • loss of bound water • decomposition TG applications: • identification and quantification of sample components • thermal stability studies TG analysis – mass (T) DSC analysis – heat flow (T) • Examples of thermal reactions resulting inheat flow, measured by DSC analysis: • crystallization • melting • glass transitions • DSC applications: • thermodynamic characterization of pure substances • quality control: sample purity • thermal stability studies

  5. Sample compatibility Requirements: • no reaction with alumina crucible • no expansion or creep during thermal decomposition Compatible materials (KUL expertise): • clays and other geological materials • cements • slags Limitations: • polymers can be hazardous, as they can foam at high temperatures, • compatibility assessment is necessary for new materials

  6. Evaluation of new materials Compatibility assessment: • by providing a reference to a thermal analysis or • by burning the sample in a furnace at the conditions foreseen for the measurement: • sample composition • temperature range (max. 1000 °C) • inert atmosphere • alumina crucible

  7. Sample preparation • Sample form: • fine powders • compact solids • films, fibers • Ensure good thermal contact between sample and heat flux-sensor: • powders: evenly distributed at the bottomof the sample crucible, gently tamped • Always use the same sample mass (~ 10-30 mg)

  8. Analysis of results Software: NETZSCH Proteus®  (Marsh procedure) Quantification of portlandite (Ca(OH)2) content in cement ~430°C: Ca(OH)2 -> CaO + H2O↑

  9. Analysis of results Software: NETZSCH Proteus®  (Marsh procedure) Quantification of portlandite (Ca(OH)2) content in cement ~430°C: Ca(OH)2 -> CaO + H2O↑ mass loss

  10. Analysis of results Software: NETZSCH Proteus®  (Marsh procedure) Quantification of portlandite (Ca(OH)2) content in cement ~430°C: Ca(OH)2 -> CaO + H2O↑

  11. Analysis of results Software: NETZSCH Proteus®  (Marsh procedure) Quantification of portlandite (Ca(OH)2) content in cement ~430°C: Ca(OH)2 -> CaO + H2O↑

  12. Contact information and reservations Contact / Training: Jacek Chwast jacek.chwast@ees.kuleuven.be tel. +32 16 3 27593 or +32 16 3 27580 Responsible professor: Jan Elsen TG /DSC instrument: Geo-Instituut Celestijnenlaan 200c Room 01.83 Reservations: - fill in the reservation list in the room 01.83

  13. Thank you for your attention Questions?