An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Nanosecond Fluorescence Lifetime Imaging using a directly gated interline transfer CCD detector by Photonic Research Systems Ltd Web: www.prsbio.com Material(c) Photonic Research Systems Ltd. 2005

  2. Experimental Details Sensor: 2/3” format monochrome interline CCD sensor, peltier cooled to -18oC Light Source: Diode pumped, frequency tripled solid-state pulsed laser with active Q-Switching. Pulse repetition rate was 5kHz, wavelength 355nm, pulse width at half maximum was 1.2 nsec, laser jitter was <1nsec. Samples: Two solutions of fluorescein were prepared one having twice the emission intensity of the other. The brighter sample was then quenched with potassium iodide until its intensity matched that of the lower intensity sample Web: www.prsbio.com Material(c) Photonic Research Systems Ltd. 2005

  3. Cuvettes of fluorescein illuminated by pulsed UV laser One cuvette holds fluorescein in buffer (4ns) while the other has a sample of twice the concentration but quenched with potassium iodide to match the dilute sample (2ns). This image was taken with the Imagex camera in steady state mode under laser illumination and with some ambient light. Web: www.prsbio.com Material(c) Photonic Research Systems Ltd. 2005

  4. Steady-State Fluorescence image of Fluorescein samples with no background light Web: www.prsbio.com Material(c) Photonic Research Systems Ltd. 2005

  5. Time-Resolved images of Fluorescein Samples quenched unquenched quenched unquenched quenched unquenched 2ns delay 4ns delay 6ns delay Delay following laser pulse Web: www.prsbio.com Material(c) Photonic Research Systems Ltd. 2005

  6. The next slide shows the resultant decay-weighted image when the appropriate pair of time-resolved images are divided. A psuedo-colour look-up table is used to show the lifetime differences. Note the very high signal-to-noise ratio, which is evident even in the uniform colour of the pixels representing very weak signals from reflections at the meniscus and base of the cuvettes quenched unquenched Web: www.prsbio.com Material(c) Photonic Research Systems Ltd. 2005

  7. Ratio ‘FLIM’ image of Fluorescein Samples N.B. the background ‘noise’ is due to the division of pixel values close to zero Notice how the lifetime ratios are clear even where the source signal is very weak. The low noise of the Imagex NanoCCD gives excellent dynamic range quenched unquenched Web: www.prsbio.com Material(c) Photonic Research Systems Ltd. 2005

  8. Combining Intensity and Lifetime information in a single image Intensity Image Intensity/FLIM Image quenched unquenched ‘Flim’ (Ratio) Image The Imagex Matrix Dialog allows the user to select 2 source images and adjust their contrast settings. The images are then combined into a single ‘Matrix’ image Web: www.prsbio.com Material(c) Photonic Research Systems Ltd. 2005

  9. Intensity-Lifetime ‘Matrix’ image of Fluorescein Samples Combining the Intensity and Flim Images in this way helps to suppress artifacts created by performing ratio operations on areas of very low intensity. quenched unquenched quenched unquenched Web: www.prsbio.com Material(c) Photonic Research Systems Ltd. 2005

  10. Future Work Investigation of smaller format and higher resolution CCD sensors Investigate Improved DPSS Light Source: 0.8 nsec FWHM with jitter <0.5nsec now available. Mount System on Microscope for real biological applications! Web: www.prsbio.com Material(c) Photonic Research Systems Ltd. 2005