520 likes | 1.29k Views
X-RAY Wor X Three Rivers Technical Conference August 3 rd & 4 th , 2010 Elk River, MN. X-RAY WorX GmbH Siemensstraße 26 D-30827 Garbsen Germany info@x-ray-worx.com http://www.x-ray-worx.com. Company Presentation Microfocus X-Ray tube technology Rod anode technology.
E N D
X-RAY WorXThree Rivers Technical Conference August 3rd & 4th, 2010Elk River, MN X-RAY WorX GmbH Siemensstraße 26 D-30827 Garbsen Germany info@x-ray-worx.com http://www.x-ray-worx.com
Company Presentation • Microfocus X-Ray tube technology • Rod anode technology
X-RAY WorXCompany Presentation X-RAY WorX GmbH Siemensstraße 26 D-30827 Garbsen Germany info@x-ray-worx.com http://www.x-ray-worx.com
Words of WelcomeAlfred Reinhold … X-RAY WorX is a group of highly experienced and competent experts, partly working in this business from the very beginning of microfocus technology. As the founder of the former company Feinfocus, that introduced the first microfocus X-ray system for industrial application with focal spot sizes below 5µm, I highly appreciate the pioneering spirit of the young team of X-RAY WorX… Alfred Reinhold,Pioneer of Microfocus X-ray Technology (February 2010) Source: http://www.alf-reinhold.de/
Vision • Sole independent supplier of high resolution micro- and submicrofocus X-ray tubes • Preferred supplier for OEMs and building up of long-term relationships • Development of innovative microfocus technology • Joint development with technical universities and scientific institutions
History • February 2010: Foundation of X-RAY WorXby Thorsten Fröba and Holger Behnsen • March 2010: First orders of OEM customers • March 2010: Finishing of the first high resolution microfocus X-ray tube of the new XWT-series • June 2010: Presentation of 160kV rod anode on ECNDT conference in Moscow • July 2010: Delivery of the first 160kV and 225kV dual-head microfocus X-ray tube
The Founders • Thorsten Fröba • 15 years experience in the development of open micro- and submicrofocus X-ray tubes • Excellent personal relationships to many prospective OEM customers • Holger Behnsen • 8 years experience in international marketing and sales of microfocus X-ray systems and tubes • Very trustful contact to OEMs, end customers and local representatives
Organisation • Thorsten Fröba, Technical GM (15) • Andreas Kösterke, Production & QA (24) • Benjamin Dostal, SW Development (5) • Mario Fiedrich, HW Development (2) • Holger Behnsen, Commercial GM (8) • Dr. Jens Peter Steffen, Sales (4) • ZoranZecevic, Service & Training (10) • Matthias Stoeber, Service & Training (2) • Marc Fisker, Service & Sales Administration (3) Altogether more than 73 years of experience in the business with microfocus X-ray tubes and systems
Fields of Business X-RAY WorXMicrofocus X-ray Technology Development & Production Services Microfocus X-ray tubes and systems Microfocus X-ray tubes Maintenance Special X-ray tubes for CT Repair Training Rod anode tubes Special projects OEM+R&D Used systems
Products • Microfocus X-ray tubes 160/225kV • Transmission target • Reflection target • Exchangeable tube head • High resolution submicrofocus tubes 80/100kV • Special targets • Special cathodes • Special tubes for CT • Rod anode tubes
Microfocus tubes XWT-160-T XWT-160-TC XWT-160-SE XWT-225-T XWT-225-TC XWT-225-SE
Rod anode tubes • Widespread application in • Aerospace industry • Tank and pipeline construction • Casting industry • NDT at inaccessible places • Engines • Hollow chamber profile elements • Pipes and tanks • Different targets and beam geometries • Exchangeable radiation heads
Summary • Sole independent supplier of high resolution open micro- and submicrofocus X-ray tubes • Concentration on core competence:Future development and production of X-ray tubes • More than 73 years of experience in microfocus technology • Worldwide sales & service
X-RAY WorXMicrofocus X-ray tube technology X-RAY WorX GmbH Siemensstraße 26 D-30827 Garbsen Germany info@x-ray-worx.com http://www.x-ray-worx.com
Schematic view of a microfocus X-ray tube • Cross section of a transmission X-Ray tube Turbo pump Vacuum gauge Tube hinge Objective coil Transmission target HT-Flange Objective aperture Electron beam Deflection unit Grid cap RTU-Filament HT-Receptacle
Schematic view of a microfocus X-ray tube • Centering of the electron beam Target current [µA] Centering current [mA] Insulated target
Schematic view of a microfocus X-ray tube • Cross section of an electron optic (reflexion) Solid target Objective aperture Objective coil Deflection unit Electron beam
Schematic view of a microfocus X-ray tube • Tube head cooling • High efficent cooling concept • Target direct cooling design • Two by-pass cooling channel • Two parallel cooling channel
Schematic view of a microfocus X-ray tube • Tube head cooling Cooling liquid outlet Cooling liquid inlet
Interaction electron - target • Reflexion target – acc. voltage low Electron beam : Focus size / shape:
Interaction electron - target • Reflexion target – acc. voltage high Electron beam : Focus size / shape:
Interaction electron - target • Transmissiontarget Low tube current Low acc. voltage High tube current Low acc. voltage Electron beam focused Target layer Substrate layer Electron beam : Focus size / shape:
Interaction electron - target • Transmissiontarget Low tube current High acc. voltage High tube current High acc. voltage Electron beam focused Target layer Substrate layer Electron beam : Focus size / shape:
Interaction electron - target • Electron penetration depth Penetration depth 20,5µm @ 100kV - Cupper Penetration depth 12µm @ 100kV - Tungsten
Spherical Aberration • Spherical Aberration Lens system Focal length Focal spot range Focal plane Aperture
Resolution and dose rate • Resolution and dose rate are depending on • Accelaration voltage • Tube current • Saturation of the filament current • Modulation of the electron beam • Material and thickness of the target layer • Aperture diameter – spherical aberration of the electro-magnetic lens system • Design of the pole shoe (focal length) • and so on….
X-RAY WorXRod anode technology X-RAY WorX GmbH Siemensstraße 26 D-30827 Garbsen Germany info@x-ray-worx.com http://www.x-ray-worx.com
Rod anode technology • Widespread application in • Aerospace industry • Tank and pipeline construction • Casting industry • NDT at inaccessible places • Engines • Hollow chamber profile elements • Pipes and tanks • Different targets and beam geometries • Exchangeable radiation heads
Rod anode technology • Standard rod anode • Target cooling with external cooling liquid feeding • Inconvenient handling • Insufficient cooling of the O-Ring sealing • Unstable vacuum during continuous operation
Rod anode technology • Rod anode with internal cooling • Internal cooling channel • Effective cooling of the O-Ring sealing up to 100W • Easy handling
Rod anode technology • Application advantage • External cooling • Limited penetration depth • Internal cooling • Optimized usage of the penetration depth
Application advantage • Radiography of only one wall thickness • Better localisation of material defects • Better detail recognition
Thank you for your attention • Any questions?