340 likes | 644 Views
A Rapid Analysis of Actinides In Urine Samples Using Eichrom’s Resin Cartridges. Anil H.Thakkar Eichrom Industries, Inc.. Darien, IL. Objective. To perform analysis for Am, Pu and U in urine samples using Eichrom’s new chromatographic configuration of UTEVA and TRU Resin.
E N D
A Rapid Analysis of Actinides In Urine Samples Using Eichrom’s Resin Cartridges Anil H.Thakkar Eichrom Industries, Inc.. Darien, IL
Objective • To perform analysis for Am, Pu and U in urine samples using Eichrom’s new chromatographic configuration of UTEVA and TRU Resin
CurrentEichrom Method • UTEVA and TRU Resin Column • Industry Standard • Flow Rate Limitation
Precedent: Actinides in Water • 1 liter water samples • Coprecipitate with calcium phosphate • Dissolve in load solution • Load on UTEVA and TRU columns in tandem
U T E V A T R U Strip Strip U Pu Am Tandem Column Arrangement Load Discard
Eichrom UTEVA Resin Column Separation Load: 3M HNO3, 1.0M Al(NO3)3 to TRU Resin column 0.01M HCl U
Eichrom TRU Resin Column Separation Load: 3 M HNO3, 1.0M Al(NO3)3 to waste 9m HCl & 4M HCl Am 0.1M ammonium bioxalate Pu
Decontamination Factors (DF) • To determine the ruggedness of the method • To determine the degree of separation of the analyte from potentially interfering matrix components
Experiment for DF • Approximately 100 pCi (3.7 Bq) of each actinide was added to a load solution and run through the method. • DF in uranium, plutonium and americium fractions were calculated.
Equation for Decontamination Factor DF = Ao Af Where; Ao = Original activity of the analyte added Af = Final activity of analyte in the fraction
Criteria for Acceptance • For actinides, a DF value of > 1000 is acceptable • Tracer Recovery: 70%-105%
New Configuration:Resin Cartridge • Packed with 2 mL of resin • Cartridges with Luer lock • Ease of tandem arrangement for sequential analysis
Eichrom Vacuum Box System • 24 Hole clear polycarbonate vacuum box • Inexpensive cartridge connection • Yellow outer tip provides support • White inner tip provides excellent seal • Vacuum gauge • Plastic rack for twenty four 50 mL C Tubes • Eichrom has limited stock on hand
Decontamination Factors (with Cartridge)
Sample Preparation • 1.2 liter aliquots of acidified urine samples were spiked with Am, Pu and U • Samples were heated to near boiling for 2 hours • A Ca3(PO4)2 precipitation was performed • Decant supernatant, centrifuge and collect the precipitate • Perform a wet digestion on the precipitate until white ash
Load Solution for UTEVA and TRU Resin Columns • Dissolve precipitate in 3M HNO3/1M Al(NO3)3 solution • Add ferrous sulfamate and ascorbic acid • Load on UTEVA and TRU Resin cartridge in tandem
Am/Cm, Pu and U Analyses Calcium phosphate precipitation Wet ash digestion Eichrom U/TEVA Resin U Eichrom TRU Resin Am/Cm, Pu Preparation for counting Alpha spectrometry
Flow rate • Sample load: Typically 1 mL/minute • Rinse solution: 2-3 mL/minute • Strip solution: 1 mL/minute • Achievable with vacuum pump
Summary • Eichrom Resin cartridges provide fast separation for actinides • Good, reproducible and consistent recoveries achievable • Rapid turn-around possible
Eichrom Cartridges • Available in TRU, TEVA, UTEVA and Sr • Other Resins available by special order • 2 mL Resin volume • Standard S grade, 50-100 micron beads • Can be stacked using Luer connections • Standard syringe barrel used for reservoir • Individually labeled w/ lot number
Future Objectives • Evaluate effect of faster flow rate on chemical recovery • To test other matrices such as; fecal, soil and vegetation for actinides using Eichrom Resin cartridges • To evaluate separation of Sr using Eichrom Sr Resin cartridges