420 likes | 514 Views
State of the art of indoor calibration of pyranometers and pyrheliometers. Main points. Most field pyranometers are calibrated indoors Many procedures for indoor calibration Not all well connected to ISO 98-3 GUM Industry requires straightforward approach. Industry.
E N D
State of the art of indoor calibration of pyranometers and pyrheliometers
Main points • Most field pyranometers are calibrated indoors • Many procedures for indoor calibration • Not all well connected to ISO 98-3 GUM • Industry requires straightforward approach
Industry • Meteorology - Solar renewable energy • Site assessment • Installation performance • Professionalisation / IEC
Conclusion • Points for discussion • Normal incidence calibration is preferred (diffuse dome not) • Uncertainty & accuracy of reference can be optimised • Pyrheliometer indoor calibration must be added
Myself • Kees VAN DEN BOS • Engineer Physics • Director / owner Hukseflux Thermal Sensors • Last 20 years sensor design
Founded 1993 Thermal sensors 15 employees 5 radiometry compensation pyrheliometer Hukseflux DR01 pyrheliometer
Background • Most pyranometers and pyrheliometers have indoor calibration • Exception: highest accuracy (BSRN, outdoor) • Exception: Japan, China (outdoor) • Cost, time, weather; outdoor calibration is unacceptable to industry
Present status (excerpt) • Eppley, US Weather Bureau: indoor integrating diffuse source • Kipp, Hukseflux: indoor normal incidence • EKO, JMA: outdoor tracker with collimation tube (changes) • China: outdoor
Background • Measurement uncertianty is a function of: • Characterisation / class • Calibration • Measurement & maintenance conditions • Environmental conditions
Background • Indoor calibration covered by ISO 9847 • Present ASME: “Indoor Transfer of Calibration from Reference to Field Pyranometers”
Hierarchy of Traceability • A: Reference calibration (uncertainty) • B: Correction of reference to indoor conditions (uncertainty) • C: Indoor calibration of field instrument (uncertainty) • Indoor calibration uncertainty estimate (A+B+C) • Field measurement uncertainty estimate
Hierarchy of Traceability • A: Reference calibration (uncertainty) • B: Correction of reference to indoor conditions (uncertainty) • C: Indoor calibration of field instrument (uncertainty) • Indoor calibration uncertainty estimate (A+B+C)
Hierarchy of Traceability • KNMI TR 235 "uncertainty in pyranometer and pyrheliometer measurements at KNMI in De Bilt".
Hierarchy of Traceability • A: Reference calibration (uncertainty) (conditions and class) • B: Correction of reference to indoor conditions (uncertainty) • C: Indoor calibration of field instrument (uncertainty) • Indoor calibration uncertainty estimate (A+B+C) • Field measurement uncertainty estimate (conditions & class)
Strange… • Errors in reference calibration re-appear in measurement errors • Counted double • At least systematic errors (Zero offset A and directional errors) can be avoided.
One step back • Present approach works well if calibrated instruments are used: • Outdoor / unventilated • Without applying GUM analysis • At same latitude
One step back • Present approach does not work well calibrated if instruments are used: • As indoor reference • Applying GUM analysis • At other latitudes • Ventilated
Typical calibration • Irradiance 800 W/m2 • 40 to 60 degrees angle of incidence, + / - 30 degrees azimuth • Zero offset A: -9 +/- 3 W/m2 (larger than ISO9060) • Directional: +/- 10 W/m2 @ 1000 W/m2 , now estimated +/- 5 W/m2
Typical calibration • PMOD specified uncertainty +/- 1.3% • Systematic error -1%? Type B.
Improved approach • Zero offset A: -9 +/- 3 W/m2 (larger than ISO9060) • Directional: +/- 10 W/m2 • Solution 1: ventilation • Solution 2: single angle of incidence
For consideration • Japanese collimated tube with tilt correction and ventilation
Diffuse sphere source • Uniformity of sphere top-edge (experimental -13%) • Weighing for non uniform source requires weighing of reference with source • Diffuse sphere: weighing requires weiging of field instrument with source. Complicated! • Normal incidence: weighing of field instrument is not necessary
Conclusion • Indoor calibration offers only acceptable solution for manufacturers and solar industry • “Normal incidence” calibration fits within ISO 98-3 GUM
Conclusion • Indoor normal incidence calibration is preferred (diffuse dome not) • Accuracy and precision of reference can be optimised (single angle, ventilated) • Pyrheliometer indoor calibration must be added
P.S. • Memo available via server