Introduction to Radiotherapy RAD 461 Dr. Ahmed Al Far, M.D. Professor of Radiation Oncology
Introduction Isodose curves: prepared under Standard Conditions • The homogeneity of the medium (water) • The vertical incidence of the beam axis on surface • The flatness of the irradiated surface • No obstructing part of the radiation field Are conditions in clinical practice always standard ? • Radiation Beam Modifiers • Beam shaping devices • Tissue compensators • Wedge filters
Beam Shaping Shielding Blocks Uses:protect some parts laying within the field ? Thickness:5 HVL can absorb 95% of radiation Made of:lead or Cerrobend (lead + bismuth) Place:over a tray mounted in the treatment head Sides:have same degree of beam divergence Distance:minimum 15 cm from skin A standard set: is supplied with treatment machine How canyou construct a Cerrobend block?
Shielding Block Thickness depends on: 1. Shielding material 2.Radiation energy
Tissue Compensators Uses: correct for irregular or curved surface (A) For Kilovoltage Radiation • Placement: directly on the skin surface (B) For Megavoltage Radiation • Placement: at distance from patient’s skin
In case of MVradiation Compensate for irregular surface A’E’C’) I) D’ B’ A’ I C’ E’ In case of KV radiation Compensate for irregular surface AEC (II) D B A II C E
Wedge Filters Uses:correct for inhomogeneous dose distribution by tilting parallel isodose line in the direction of the thin end of the filter. Made of:lead Place:over a tray mounted in the treatment head Distance:minimum 15 cm from skin Wedge angle:between 50% isodose line and a line at right angle to the central axis • Wedge Systems: • Individualized: a separate wedge for each field, telecobalt ? • Universal: a single wedge for all field sizes, L. accelerator ?
Wedge Filters Used in superficial tumors: e.g. parotid or maxillary sinus
The Hinge Angle A B Dose at point A is greater than that at point B ? Wedge angle = 90o – ( / 2)