CEQ Sequencing and Fragment Analysis Troubleshooting. CEQ User Group Meeting Noreen Galvin Ph.D. Introduction. Run Method Summary Troubleshooting Tools Normal traces Current problems Raw data problems Mixed signals Difficult templates. Run Methods. Sequencing LFR-1 = 700bp
E N D
CEQ Sequencing and Fragment Analysis Troubleshooting CEQ User Group Meeting Noreen Galvin Ph.D.
Introduction • Run Method Summary • Troubleshooting Tools • Normal traces • Current problems • Raw data problems • Mixed signals • Difficult templates
Run Methods • Sequencing • LFR-1 = 700bp • LFR-a = >800bp • LFR-b = 600bp • LFR-c = 500bp • Seq-Test = Sequencing Test Sample/pUC only • Fragment Analysis • Frag1 • Frag2 • Frag3 = 400bp frags • Frag4 = 600bp frags • SNP-1 = SNP anaysis • Frag-Test = Fragment Test Sample only
Troubleshooting Tools • Sequencing Test Mix (Part 608070) • Instrument • Gel • Capillary • Method • pUC control and -47M13 Primer • Reagents • Technique • Samples • DNA & Primer These three tests rely on the analysis of Raw Data signal, Current profiles and Analysed Data integrity
Troubleshooting Tools • Sequencing Test Mix & pUC18dG/-47 • Method Seq-Test • >700bp @ >98% accuracy (2%N) • Align with pUC18dG
Current Problems • Current problems are associated with the electrokinetic injection of sample and the subsequent separation of the DNA sequencing fragments in the Linear Polyacrylamide (LPA) matrix • Electrokinetic Injection • All negative charged molecules loaded • DNA sequencing products • DNA template • RNA • Protein • Salts
Current Problems • Gradually decreasing current => short reads
Current Problems • Gradually decreasing current => short reads • Some excess of supercoiled DNA template • Re-inject sample • Reduce injection • Reduce [template] • Pre-heat treatment • 96°C 1 minute
Current Problems • “Crashed” current • Short/failed read • Inaccurate base calling
Current Problems • Crashed current • Reduce [template] • Pre-heat treatment • 96°C 1 minute • Cleave template • Restriction • Remove template after reaction • Magnetic beads
Current Problems • Optimizing • 96°C • 1 minute • 2 minutes • 3 minutes • 85°C • 5 minutes
Current Problems • Identical current crash in all eight capillaries • Air bubble in manifold • Manifold purge and re-run samples
Current Problems • Current too high • >10µA on LFR-1 • Data starts at 20 minutes • Poor resolution • Gel degraded • Excess hours on CEQ
Raw Data Signal Problems • Full available range • Min Signal > 3x baseline • Max signal < 130,000 units • High Background • Reduced sensitivity with lower signals • Signal too low • Short/no read • Minimum signal:noise > 3:1 • Signal too high • Base calling errors • Too many peaks • Multiple sequencing products in one tube
Raw Data Background • If >6000 rfu (relative fluorescence units) • Clean capillary window • MQ water • 70% Ethanol • Blow dry
Low Raw Data Signal • No Signal & no dye peaks • No signal but dye peaks present
Low Raw Data Signal • No Signal & no dye peaks • Lost pellet • (Dye degradation) • Instrument Problem • Centrifuge • CEQ • No signal but dye peaks present • Chemistry problem • [Template DNA] • Salts in sample • [primer] • Poor primer • Polymerase inhibitor • Instrument Problem • Thermal cycler
Low Raw Data Signal[DNA Template] • Too little DNA Template • Not enough reaction products • Add more DNA • Correct [DNA Template] • 50-100 femtomoles for dsDNA • 25-50 femtomoles for ssDNA • 10-50 femtomoles for PCR products
Quantitation of [Template DNA] • Agarose Gel • Fluorescence • Picogreen • Spectrophotometer
Low Raw Data Signal • Salt in template or primer • Competition for injection • Add less template • Purify DNA template • Re-synthesize primer
Low Raw Data SignalPrimer Problems • Low [primer] • Inefficient primer binding • Primer design • Cycling conditions • Cycle number • Annealing temperature
Primer Design Rev (Tm 79.8) • Ideal Primer length 22-25bp 50% G/C • Tm >55°C • Annealing temp + 5°C Fwd (Tm 43.5)
Primer Design • Primer Interactions • Hairpins • Primer dimers
Primer Quality • Poor quality primer • Degraded • Freeze-thawing • Store in MQ water or 10mM Tris • Poor synthesis • Check primer integrity on CE/HPLC • Poor purification • Minimum = desalting
Low Raw Data SignalPolymerase Inhibitors • Polymerase inhibitors • DEPC water • Phenol • EDTA • DNA binding proteins • Other…. • Purify Template • Spin columns • Store primer • Water or Tris
Low Raw Data SignalDye Degradation • Typically red/black dyes degrade first • Dyes stable in: • Reaction buffer • SLS under mineral oil • Bad formamide • Use CEQ kit SLS (part 608082) • Do NOT freeze-thaw • Heating pellets to dry them • Dry in vacuum or air • No mineral oil overlay in CEQ plate • Always overlay samples in SLS
Low Raw Data SignalDye Degradation • Typically red/black dyes degrade first
Low Raw Data SignalEthanol Contamination • Typically red dye affected the most Massive noise in the D1 channel Low sample signal
Instrument Issues • Thermal Cycler • Heated lid • Sample evaporation with plates/caps • Block problem (PCR may still work) • Centrifuge • Refrigerated • Forces > 1000 x g for plates, >10,000 x g for tubes • CEQ • Check with sequencing test sample • Resolution and accuracy • Capillary, gel, instrument
Data Truncated (off-Scale) Raw Data Signal Too High • When the signal is too high this causes the data to be “truncated” (off-scale) • This can result in the addition of erroneous peaks in the base calling
Raw Data Signal Too High • Reduce injection of samples • Reduce [DNA template] • Reduce cycle number • Check [DNA template] before reaction
Correct Raw Data SignalAnalysis Poor or Fails • Insertions Caused by: • Mixed Templates • PCR Primer Carryover • Primer Mis-priming
Correct Raw Data SignalAnalysis Poor or Fails • Template contamination
Correct Raw Data SignalAnalysis Poor or Fails • Template contamination • PCR products: • Multiple products • Check on agarose • Purify from agarose • Plasmids • More than one colony • Pick another colony • Re-plate
Low Level Contamination of PCR Template • Reduce injection • Reduce amount of template • Increase annealing temp in PCR reaction • Re design PCR primers
Correct Raw Data SignalAnalysis Poor or Fails • Primer contamination • PCR templates – primer carry-over • Purify by spin column or ExoSap • Primer degredation • N-1 peaks • Fresh primer
Primer Degradation • N-1 peaks • Pre-peak reduction in S/W • Resynthesize primer
Difficult Templates • Short PCR Products • Difficult to purify • G/C rich samples • Form hairpins • PolyA or polyT • Enzyme slippage
Short PCR Products (<200bp) • Low recovery on Spin columns
Short PCR Products (<200bp) • Low recovery on Spin columns • Exo-SAP purification • ExonuleaseI + Shrimp Alkaline Phosphatase • 0.05u SAP + 0.1u Exo per µl PCR reaction • 37°C for 60 minutes • Inactivate at >75°C for 15 minutes • Hold at 4°C
G/C rich templates • Range of Heat Treatments • From 1 to 5 minutes • Adjust thermal cycling conditions: • Hot start • 96 ° C for two minutes • Thirty cycles of • 96 ° C 30seconds • 60 ° C 4 minutes • Hold 4 ° C