ASSESSMENT OF SUPPORT SURFACES (STATIC & MIXED) FOR NEONATES. FP-50002 EWMA/GNEAUPP 2014. M. M. Sánchez-Lorente 1 ; P. García-Molina 2-3 ; C. Vergara-Hernández 2 ; E. Sanchis-Sánchez 2 ; J. Martos-Torres 2 ; A. Álvarez-Ordiales 4 ; & E. Balaguer-López 2-3.
E N D
ASSESSMENT OF SUPPORT SURFACES (STATIC & MIXED)FOR NEONATES FP-50002 EWMA/GNEAUPP 2014 M. M. Sánchez-Lorente1; P. García-Molina2-3; C. Vergara-Hernández2; E. Sanchis-Sánchez2; J. Martos-Torres2; A. Álvarez-Ordiales4; & E. Balaguer-López2-3 Conselleria de Sanitat, Valencia (SPA). University of Valencia, Valencia (SPA). Hospital Clínico Universitario, Valencia (SPA). Hospital Universitari i Politècnic “La Fe”, Valencia (SPA). Presenting author: Mr. Pablo García-Molina
FP-50002 EWMA/GNEAUPP 2014 Background Existence of several neonatal supportsurfaces however… Lack of evaluation regarding… • Interface pressure & pressureredistribution • Efficacy in thepreventionortreatment of PU
FP-50002 EWMA/GNEAUPP 2014 Aim 5 2 1 • To assess the interface pressure and pressure redistribution of five devices of constant low pressure (static and mixed support surfaces). 3 4
FP-50002 EWMA/GNEAUPP 2014 Methods Study design • Laboratory study with repeated measures design. Sample • Eight neonatal simulation dummies with different characteristics regarding head circumference, weight and height.
FP-50002 EWMA/GNEAUPP 2014 Methods Supportdevices
FP-50002 EWMA/GNEAUPP 2014 Methods Variables • Mean pressure (mmHg). • Maximum pressure (mmHg). • Mean pressure in head region (mmHg). • Area of contact (cm2). Equipment • XSENSOR X3 PX100:40.40.002 device. Procedure • See: E-poster no. 50006.
FP-50002 EWMA/GNEAUPP 2014 Methods Statistical analysis • Descriptive statistics. • Due to a small sample size, a non-parametric test (Friedman test) was carried out to assess differences among groups. If a significant difference was found, post hoc comparisons among groups were performed using the Wilcoxon-Nemenyi-McDonald-Thompson test. • The analysis was carried out by means of the R environment (The R Foundation for Statistical Computing). • Data set and code employed for the analysis can be found at www.figshare.com (*). *http://dx.doi.org/10.6084/m9.figshare.977866
FP-50002 EWMA/GNEAUPP 2014 Results Lowest values! Highest values!
FP-50002 EWMA/GNEAUPP 2014 Results
FP-50002 EWMA/GNEAUPP 2014 Results Post hoc analysis: mean pressure (p-value)
FP-50002 EWMA/GNEAUPP 2014 Results Post hoc analysis: maximum pressure (p-value)
FP-50002 EWMA/GNEAUPP 2014 Results Post hoc analysis: pressure area (p-value)
Conclusion • The “T” device presents the best values of pressure relief. • The polyester fiber device presents the best values of pressure redistribution. • Adding an upper layer of polyester fiber to the “T” device could result in a better support surface. • Next step: studying the properties of the new support surface in healthy subjects under clinical conditions.