170 likes | 319 Views
Institute of Chemical Technology Prague , Department of Analytical Chenistry , Technická 5, 166 28 Prague 6, Czech Republic. LiquiScint 2004. Determination of low b -activity of 99-technetium in the bentonite leachate. H. Vinšová, P. Večerník , V. Jedináková- Křížová. LiquiScint 2004.
E N D
Institute of Chemical Technology Prague, Department of Analytical Chenistry, Technická 5, 166 28 Prague 6, Czech Republic LiquiScint 2004 Determination of low b-activity of 99-technetium in the bentonite leachate H. Vinšová, P.Večerník, V. Jedináková-Křížová
LiquiScint 2004 • Radionuclide of 99Tc (T1/2 = 2.1·105 years) fission product of spent nuclear fuel • Aerobic conditions TcO4- • Anaerobiccondtions TcO2 or TcO(OH)2 Fig 1.: Eh-pH diagram of Tc
2FeS + 5H2O + 1/2O2 2Fe(OH)3 + 2H2S (1) 2TcO4- + 7H2S Tc2S7 + 8H2O (2) LiquiScint 2004 Suitable modificators for the uptake of 99TcO4- FeS reduction of TcO4- to insoluble TcO2.nH2O anaerobic conditions aerobic conditions • Fe • reductive conversion of Tc(VII) to the lower oxidation state Tc(IV) • sorption of the reduced species onto hydroxide surfaces • precipitation of Tc with iron hydroxides on surfaces • Activated carbon => sorption
GOAL LiquiScint 2004 Liquid scintillation method Sorption of technetium on the bentonite mixtures
LiquiScint 2004 Sorption of 99TcO4- on the bentonite mixtures Solid phase: bentonite R (0.315 - 0.8 mm) + Fe (powder) bentonite R (0.315 - 0.8 mm) + FeS (powder) Liquid phase: bentonite water (14 days leaching of bentonite in deionized water on the s:l phase 1:10 ratio) • Sorption experiment: • bentonite R 0.7 g + modificator (Fe, FeS) + 7 mL bentonite water + 50 mL 99TcO4- (c=0.0111 mol/L, A= 740 kBq/mL) • shake (70 period/min.) • centrifugation (20 min./2000rpm) • bentonite leachate
LiquiScint 2004 • Bentonite leachate: • Eh, pH,specific conductivity • b-activity 99Tc • b-scintillation detector NaI (Tl), SPF 35(evaporation residue - 1 mL of bentonite leachate) • liquid scintillation counter • ( 2 mL of bentonite leachate) Fig. 2.:b-scintillation detector NaI (Tl), SPF 35 Fig. 3.: Triathler
Liquid scintillation method LiquiScint 2004 Scintillation cocktail study Scintillation volume and vials study Determination of uncertainty of sample dilution
Scintillation cocktail study Aqua Light Ultima GOLD XR Table 1.: Comparison of selected scintillation cocktails; reaction mixture: 2 mL of scintillation cocktail + 2 mL of bentonite leachate labelled by 99Tc • Next experiments => scintillation cocktail Aqua Light LiquiScint 2004 ? Scintillation cocktail
Scintillation volume study Table 2.: The values of b-activity 99Tc measured in the different vials, Sample: b-activity of 99Tc in the aqueous solution 8 Bq/mL. • Optimum reaction mixture => 2mL of scintillation cocktail + 2 mL of bentonite leachate labelled by 99Tc LiquiScint 2004 • Suitable vials and amount of scintillation cocktail ? Fig. 4.: Tested vials for the measurement of studied samples
Determination of uncertainty of sample dilution (A0-A).V KD= A.m LiquiScint 2004 Initial value of b-activity 37 kBq/7mL => sample diluted 122.5 times A0 - total activity, A - activity of liquid phase, V - volume of liquid phase (mL), m - weight of solid phase(g) Expanded uncertainty => 16.7 mL Liquid scintillation counter expanded uncertainty of measurement =>2% b-scintillation detector NaI (Tl) expanded uncertainty of measurement =>5%
Liquid scintillation method LiquiScint 2004 Figure 5.: Example of liquid scintillation b-spectrum of 99Tc, reaction mixture: 2mL of Aqua Light + 2 mL of aqueous solution labelled by 99Tc
Kinetics of sorption 99TcO4- • 2 days of uptake => sorption yield over 99% • low value of b-activity 99Tc => LS method LiquiScint 2004 Figure 6.: Sorption kinetics of 99TcO4- on the different bentonite mixtures A) solid phase: bentonite mixture with Fe; mass ratio bentonite:Fe 10:1 B) solid phase: bentonite mixture with FeS; mass ratio bentonite:FeS 10:1.
Kinetics of sorption 99TcO4- LiquiScint 2004 Table 3.: Comparison of sorption yield and KD value on the different solid phases obtained by b-scintillation detector NaI (Tl) and LS cocktail Aqua Light for the sorption kinetics of 99TcO4- A) solid phase: bentonite mixture with Fe; mass ratio bentonite:Fe 10:1, B) solid phase: bentonite mixture with FeS; mass ratio bentonite:FeS 10:1. A) B)
Effect of ratio bentonite:additive on 99TcO4- sorption LiquiScint 2004 Figure 7.: Dependence of sorption yield of 99TcO4- for the different bentonite:additive ratios A) Solid phase - bentonite:Fe ratios 4:1 - 150:1, B)Solid phase - bentonite:FeS ratios 5:1 - 100:1, contact of both phases 8 days • bentonite:Fe 4:1 – 150:1 and bentonite:FeS 5:1-20:1 ratios => sorption yield of 99TcO4-> 90%
Effect of ratio bentonite:additive on sorption 99TcO4- LiquiScint 2004 Table 4.: Comparison of sorption ylield and KD value for the different solid phases obtained by b-scintillation detector NaI (Tl) and LS cocktail Aqua Light - study of the effect of different bentonite:additive ratios A) solid phase: bentonite mixture with Fe; B) solid phase: bentonite mixture with FeS. A) B)
CONCLUSIONS LiquiScint 2004 Liquid scintillation method Scintillation cocktail Aqua Light Optimum compositon of reaction mixture 2 mL of scintillation cocktail + 2 mL of bentonite leachate labelled by 99Tc Sorption experiments Suitable additives for the uptake of 99TcO4- are FeS and Fe Optimum ratios bentonite:FeS and bentonite:Fe 10:1, contact of both phases 1 day => sorption yield > 90%
LiquiScint 2004 Thank you for your attention !