1 / 11

PCD Infusion Pump WG Containment / Mode / Status Discussion (rev. 3)

PCD Infusion Pump WG Containment / Mode / Status Discussion (rev. 3). Updated: 2012.05.02 Wednesday. DS-IP-xyz Containment/Mode/Status. Discussion Scope: Review proposed containment models for : DS – Infusion Pump – LVP DS – Infusion Pump – Syringe DS – Infusion Pump – PCA

marcos
Download Presentation

PCD Infusion Pump WG Containment / Mode / Status Discussion (rev. 3)

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. PCD Infusion Pump WGContainment / Mode / Status Discussion (rev. 3) Updated: 2012.05.02 Wednesday

  2. DS-IP-xyz Containment/Mode/Status • Discussion Scope: • Review proposed containment models for : • DS – Infusion Pump – LVP • DS – Infusion Pump – Syringe • DS – Infusion Pump – PCA • Define role and values for pump Mode parameter • Define purpose and values for pump Status parameter

  3. DS-IP-LVP Containment Model Spreadsheet => RTMMS => ICSGenerator • LVP VMD is paired with a generic pump delivery channel • Source Channels package semantics that define / drive the behavior of a Delivery Channel • Multiple source channels can be “active” at any point in time – thus the addition of a “bolus” channel • Source & Delivery Channel definitions consistent across pump types.

  4. DS-IP-SYR Containment Model Spreadsheet => RTMMS => ICSGenerator • Syringe-specific parameters are contained in the Syringe Channel • Source & Delivery Channels are the same as those specified in the General Purpose pump specification.

  5. DS-IP-PCA Model Spreadsheet => RTMMS => ICSGenerator • PCA specific parameters are packaged in a PCA Channel • PCA Channel can be combined with Source Channels as needed • Syringe-based PCA would also include a Syringe Channel (next slide) • PCA Channel may operate exclusively, sequentially or concurrently with one or more Source Channels.

  6. DS-IP-PCA-SYR Model Spreadsheet => RTMMS => ICSGenerator • PCA capability added to a syringe pump • Only difference from the generic PCA pump is the addition of the Syringe Channel parameter package.

  7. Pump Mode & Status Parameters • Discussion Points: • Operational Mode & Status contained in delivery channels • Mode reflects … • what source channels are actively determining the operation of their associated delivery channel • What the user would “see” as the configured “mode” • All source channels OR only currently active source channels? • Status reflects … • primarily what is observed in the distal line • … • ISSUE #1: What do we want these values to be for DS-IP-xyz? • Including active / concurrent source channel indication • ISSUE #2: Harmonization with the 11073-10101:2004 Definitions

  8. X73 Pump “Mode” Parameter • hRTM Group: “Operational Mode”: • pump-mode-nominal • pump-mode-drug-dosing • pump-mode-ramp-taper • pump-mode-multi-step • pump-mode-multi-dosing • pump-mode-bolus • pump-mode-loading-dose • pump-mode-multi-channel • pump-mode-pca • pump-mode-continuous • pump-mode-pca-and-continuous • pump-mode-piggyback • pump-mode-concurrent (from 11073-10101:2004, A.7.7.12) MDC_PUMP_MODE

  9. X73 Pump “Status” Parameter • hRTM Group: “Operational Mode”: • pump-status-ready • pump-status-infusing • pump-status-paused • pump-status-kvo • pump-status-delayed • pump-status-standby • pump-status-vtbi-complete • pump-status-off • pump-status-priming (from 11073-10101:2004, A.7.7.18) MDC_PUMP_STAT

  10. Proposed Resolution • Define two Operational Mode parameters + 1 status parameter • MDC_PUMP_MODE_SETTING (proposed) • Indicates the configured infusion “protocol” • Enumerations include: pump-mode-continuous / -piggyback / -pca / … • Multiple mode “bits” can be set • Example: pump_mode_continuous + pump_mode_pca + pump_mode_bolus • MDC_PUMP_MODE_CURRENT (proposed) • Indicates the actual operational – including simultaneous – mode components (tied to Source Channels) • Example: pump_mode_continuous + pump_mode_pca • MDC_PUMP_STAT (… can we make this _STATUS?) • Indicates what the pumping mechanism is doing • Does NOT include alarm status indication (separate parameter) • Enumerations include: pump-status-ready / -infusing / -paused / -delayed / -standby / -off / …

  11. PCD Rules!