1 / 31

Chromatography Integration and Trouble-shooting

Chromatography Integration and Trouble-shooting. Presented by: Marlene Moore mmoore@advancedsys.com October 2007. Agenda. Problems Tools Solutions. Problems. Baseline Disturbances Irregular Peak Shapes, Sizes Retention Time Shifts

markku
Download Presentation

Chromatography Integration and Trouble-shooting

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Chromatography Integration and Trouble-shooting Presented by: Marlene Moore mmoore@advancedsys.com October 2007

  2. Agenda • Problems • Tools • Solutions

  3. Problems • Baseline Disturbances • Irregular Peak Shapes, Sizes • Retention Time Shifts • Loss of Separation or Resolution • Quantitation Difficulties • Rapid Column Deterioration • Ghost Peaks • Broad Solvent Front

  4. Baseline Disturbances spiking noise wander drift offset

  5. Spiking • Particulate Matter Passing Through Detector • Loose Connections • Cables • Circuit boards

  6. Noise • Contaminated Injector/column • Physical Defect Detector • Defective Detector Board • Carrier Gas Leak at Septum or Column • FID, NPD, FPD • Column in Flame • Incorrect Combustion Gases or Flow Rate/leak • ECD, TCD • Air Leak

  7. Wander • Contaminated Carrier Gas • Isothermal Conditions • Contaminated Gas Chromatograph • Poor Control of Carrier Gas or Detector Gas Flows • Poor Thermal Control of Detector

  8. Drift • Septum Bleed • Upward • GC or column contamination • Damaged stationary phase • Downward • Incomplete conditioning of column • Unequilibrated detector

  9. Offset • Contamination: • Injector, column, detector • Carrier gas - change/install impurity traps • Column Inserted Into Flame (FID, NPD, FPD) • Malfunctioning or Improperly Set Recording Device

  10. Peak Shape Peak Shape tells the story of your chromatography

  11. Irregular Peak Shapes tailing fronting split reduced rounded negative

  12. No Peaks • Plugged Syringe • Broken Column • Injecting Sample Into Wrong Injector • Column Installed in Wrong Detector • Integrator Reading Wrong Signal or Not Connected • Detector Not on • Temperature • Oven not heated • Column temperature too low • Injection port temperature too low • Detector Gases Improperly Set or Not on • Very Low or No Carrier Gas Flow

  13. Reduced Peaks • Plugged Syringe • Change in Injection Technique • Large Leak - Injector • Split Ratio Too High • Too Short of Purge Activation Time for Splitless Injections • Very High Septum Purge Flow • Too Low Injector Temperature • Column Temperature Too Low • Sample Components Absorbed by Column or Inlet Liner

  14. Reduced Peaks • Initial Temperature Column Too High for Splitless or On-column Injections • High Background Signal - • Contamination • Excessive column bleed • Check auto zero function and setting • Improperly Operated Detectors • Impurities Carrier Gas or Detector Gas • Detector Compound Mismatch • Excessive Attenuated Integrator Signal • Sample Concentration or Integrity Problems

  15. Tailing Peaks • Active Injector Liner or Column • Contaminated • Injector liner • Column • Dead Volume Caused by Poorly Installed Liner, Column or Union • Poorly Cut Column End • Polarity Mismatch - Stationary Phase, Solute or Solvent • Cold Spot in the Flow Path

  16. Tailing Peaks • Solid Debris in Liner or Column • Poor Injection Technique - Too Slow • Too Low of a Split Ratio • Overloading on a PLOT Column • Decrease sample amount • Amines and carboxylic acids tail on most columns • Need ph modified stationary phase • Some Peaks Always Exhibit Tailing

  17. Rounded or Flat Topped Peaks • Overloaded Detector • Exceeding Range of Integrator

  18. Split Peaks • Detector Overloaded • Poor Injection Technique • Poorly Installed Column Into Injector • Column Temperature Fluctuations • Coelution of Two or More Compounds • Mixed Sample Solvent for Splitless or On-column Injections.

  19. Negative Peaks • All Peaks Negative - Polarity Check Connectors • TCD - Select Peaks • Thermal conductivity of compound greater than carrier gas • ECD - After Positive Peak • Dirty or old ECD cell

  20. Retention Time Shifts • Different Column Temperature • Different Carrier Gas Flow or Linear Velocity • Leak in Injector - Septum • Contaminated Column • Change in Sample Solvent

  21. Loss of Separation/Resolution • Contaminated Column • Damaged Stationary Phase • Different Column Temperature, Carrier Flow Rate, Column • Large Changes in Sample Concentrations • Improper Injector Operation

  22. Quantitation Difficulties • Injection Techniques • Split Discrimination • Different Purge Activation Time for Splitless Injection • Baseline Disturbances • Improper Integrator Settings • Inconsistent Detector Gas Flows or Temperatures • Column or Liner Activity

  23. Rapid Column Deterioration • Exposure of Column to Air at Elevated Temperatures • Exceeding Upper Temperature Limit of Column • Chemical Damage • Contamination of Column With High Molecular Weight Materials • Column Breakage

  24. Ghost Peaks • Contamination • Injector • Column • Septum Bleed • Dirty Syringe • Sample Too Large • Previous Run Terminated Too Soon BOO!

  25. Broad Solvent Front • Poorly Installed Column • Leak in the Injector • Solvent Interacting with Detector • Too Low • Split Ratio (Flow) • Injector Temperature • Too Long Purge Activation for Splitless Injections • Large Injection Volume • Low Column Temperature and High Boiling Solvent • High Column Temperature and Low Boiling Solvent

  26. Tools • Flowmeter • Syringe (NEW) • Methane or Nonretained Compound • Septa, Ferrules, Injection Liners (NEW) • Leak Detector • Column Test Mixture (Reference) • Checkout Column • Instrument Manuals

  27. Solutions Quickest Solution ---- Improve Chromatography --- Do Manual Integration --- Get Results out Fast !! NO! WRONG ANSWER!!!

  28. Solutions • Preventive Maintenance • Clean detector • Clean injector • Clean and repair parts • Clean and tighten connections • Follow instrument manual suggestions • Allow Time for Optimizing System • Integration parameters • Programs

  29. Does anyone have a question? Question ! Where do you document your trouble shooting?

More Related