INFN-LNS. DOSIMETRY COMMISSIONING OF THE LNS-INFN. PROTON THERAPY FACILITY THE DOSIMETRIC CHARACTERISTICS OF NARROW PROTON BEAMS USED IN EYE THERAPY HAVE BEEN DEFINED BY USING DIFFERENT RADIATION DETECTORS IONIZATION CHAMBERS ( parallel-plate, mini thimble chambers ) SILICON DIODES
E N D
INFN-LNS DOSIMETRY COMMISSIONING OF THE LNS-INFN PROTON THERAPY FACILITY THE DOSIMETRIC CHARACTERISTICS OF NARROW PROTON BEAMS USED IN EYETHERAPY HAVE BEEN DEFINED BY USING DIFFERENT RADIATION DETECTORS IONIZATION CHAMBERS (parallel-plate,minithimble chambers) SILICON DIODES RADIOCHROMIC AND RADIOGHRAPHIC FILMS THERMOLUMINESCENCE DETECTORS 1)CENTRAL AXIS DEPTH-DOSE DISTRIBUTIONS 2)LATERAL (TRANSVERSE) PROFILES 3)OUTPUT FACTORS (FSDF) TO PROVIDE INPUT TO THE TPSAND FOR M.U. CALCULATIONS. QUALITY CONTROL PROCEDURES DOSIMETRIC COMMISSIONING OF 60 MeV PROTON BEAMS PRODUCED AT LNS FOR EYE THERAPY SHAPED WITH 5 - 30 mm CIRCULAR BRASS COLLIMATORS.
PROTON DEPTH-DOSE DISTRIBUTIONS • THE PTW MARKUS PLANE-PARALLEL ION CHAMBER WAS USED IN THE COMMISSIONING AS THE REFERENCE DETECTOR FOR DEPTH-DOSE MEASUREMENT IN PROTON BEAMS (ICRU 59, IAEA 398) • • THE DESIGN OF THE MARKUS CHAMBER • • 1)Active volume=0.05 cm3, 2) Electrode separation=2 mm, 3) Collector diameter=5.4 mm • 4)tentrance window=2.3 mg/cm2 5) Exact Location of peff. • • PROVIDES DEPTH-DOSE DISTRIBUTIONS • WITH HIGH SPATIAL RESOLUTION AND HIGH PRECISION • ) PMMA CUP IS PROVIDED TO SEAL THE CHAMBER FOR USE IN WATER • )A SPACER IS PROVIDED FOR USE IN A SOLID PHANTOM, WHICH ALSO SERVES AS A HOLDER FOR A RADIOACTIVE SOURCE FOR STABILITY CHECKS. INFN-LNS
PTW MARKUSPLANE-PARALLER ION CHAMBER IN WATER PHANTOM INFN-LNS
CENTRE MAXIMUM RANGE (mm) EQUIVALENT ENERGY (MeV) PEAKPLATEAU RATIO F.W.H.M. (mm) Distal-dose falloff d90%-10% (mm) CATANA 30.60 60.2 4.68 3.29 0.81 CCO^ 31.0 60.4 4.85 3.65 0.80 PSI^ 30.0 60.0 4.47 1.10 INFN-LNS ^ taken from BJR Supplement 25 (1996)
THE NEW SCANDITRONIX Si-DIODE IN PROTON BEAMS (Proton Field waterproof Detector-narrow beamno.DZA192 1001) INFN-LNS
DETECTOR PEAK DEPTH (mm) PEAK-PLATEAU RATIO F.W.H.M (mm) Distal-dose falloff (1) d90%-10% (mm) Distal-dose falloff (2) d80%-20% (mm PRACTICAL RANGE (d10%, ICRU 59) THE NEW SCANDITRONIX Si-DIODE IN PROTON BEAMS (Proton Field waterproof Detector-narrow beamno.DZA192 1001) MARKUS PTW 30.14 4.68 3.19 0.71 0.50 31.15 DIODE SCANDITRONIX 30.07 4.89 3.07 0.80 0.60 31.06 CYLINDRICAL MINIDIODE SPECIFICATIONS Detector Material: Hi-pSi, high doped p-type silicon (preirradiated for use in proton beams). Detector diameter: 0.6 mm (t = 60 m) INFN-LNS
MODULATED PROTON DEPTH DOSES FOR EYE THERAPY MODULATION (SOBP) (mm eye tissue) DISTAL-DOSE FALLOFF (1) d90%-10% (mm eye-tissue) DISTAL-DOSE FALLOFF (2) d80%-20% (mm eye tissue) MAXIMUM DOSE IN SOBP % BEAM RANGE (90% DISTAL) (mm eye-tissue) MODULATOR RANGE SHIFTER MOD: 000/00 RS: 14 mm 14.5 0.91 0.65 101.6 14.38 MOD: 010/02 RS: 10 mm 10.5 0.90 0.60 104 18.27 MOD: 009/02 RS: 1.8 mm 20.30 0.80 0.55 103.9 27.04 CCO 17 0.90 0.75 < 102 INFN-LNS
MODULATED PROTON DEPTH DOSES FOR EYE THERAPY 15 MM SOBP vs RANGE SHIFTER INFN-LNS
BEFORE EACH TREATMENT PERIOD TREATMENT DEPTH DOSE PROFILES HAVE TO BE VERIFIED. TO VERIFY RANGE MODULATION AND MAXIMUM RANGE OF THE BEAM DEPTH-DOSE MEASUREMENT WHEEL (CCO DIODE SCANNER) WHEEL IS COMPUTER CONTROLLED, AND RELATIVE DOSE MEASURED AS RATIO BETWEEN WHEEL’S DETECTOR AND REFERENCE DETECTOR TOLERANCES 1) [(MEASURED RANGE) / (REQUIRED RANGE)]: 0.2 mm 2) [(MEASURED MODULATION) / (REQUIRED MODULATION)]: [ -0.1 mm // +0.9 mm ] INFN-LNS
LATERAL OFF-AXIS BEAM PROFILES 1) LATERAL PENUMBRA: d80%-20% 2) Field ratio: 3) L95% 4)SIMMETRY (AREA RATIO): 5) FLATNESS: MD-55-2 RADIOCHROMIC FILMS WERE USED IN COMMISSIONING tissue equivalence – dose rate indipendence Linear dose response – high spatial resolution ONLY ONE CALIBRATION FILE IS NEEDED TO EVALUATE FILMS EXPOSED AT DIFFERENT DEPTHS Energy indipendence 1.3 mm INFN-LNS
He-Ne Scanning LaserDensitomiter PMMA Phantom
INFN-LNS KODAK XV FILMS AND SCANDITRONIX DIODE WERE USED IN COMMISSIONING IF CALIBRATION FILES PRODUCED FOR KODAK XV FILMS AT DEPTH OF SOBP ARE USED FOR TRANSVERSE BEAM MEASUREMENTS GAF MD55-2 = KODAK XV FILM
BEFORE EACH TREATMENT PROTON BEAM PROFILES AT ISOCENTRE ARE TO BE CHECKED IN AIR X-Y DIODE SCANNING COMPUTER CONTROLLED DEVICE PLANNED TOLERANCES 1) LATERAL PENUMBRA (d80%-20%)1.50 mm 2) BEAM SIMMETRY (Sr )3% 3)BEAM FLATNESS: Rt%3%4) FIELD RATIO: H0.90 INFN-LNS
DOSE MONITORING SYSTEM (IN-BEAM DOSE MONITORS) THE PROTON DOSE IS MONITORED BY TWO INDEPENDENT TRANSMISSION UNSEALED ION CHAMBERS, PLACED IMMEDIATELY UPSTREAM OF THE PROTON NOZZLE. TRANSMISSION CHAMBERS HAVE SEPARATE CABLING, BIAS SUPPLIES (800 V) AND CURRENT INTEGRATORS AND ARE ARRANGED AS A REDUNDANT COMBINATIONTO TERMINATE THE BEAM. TRANSMISSION ION CHAMBERS ARE CALIBRATED DAILY AGAINST A REFERENCE PARALLEL-PLATE MARKUS ION CHAMBER, LOCATED AT THE ISOCENTRE. COMMISSIONING TESTS 1)PRECISION 1a)SHORT TERM PRECISION: (CV) OF THE RATIO ROF DOSE MONITOR UNITS TO DOSIMETER SCALE READING FOR n=5 CONSECUTIVE IRRADIATIONS OF 15 Gy. MEASURED CV = 0.1% 1b) LONG TERM PRECISION (WEEKLY STABILITY): MEASURED [(cGy/U.M.)WEEK]:1.5%. 2)LINEARITY THE RATIO V BETWEEN THE MEASURED PROTON DOSE AND DOSE MONITOR UNITS SHALL BE WITHIN 1% AT ALL MONITOR SCALE READINGS, i.e. IN THE CLINICAL RANGE UP TO 15–20 Gy. MEASURED RATIO INFN-LNS
DOSE MONITORING SISTEM TRANSMISSION ION CHAMBER INFN - LNS
DOSE RATE MONITORING (BEAM INTENSITY MONITOR) THE PROTON BEAM RATE IS MONITORED BY THE VOLTAGE SIGNAL (VSF) PROVIDED BY THE FIRST SCATTERING FOIL, LOCATED IN VACUUM IMMEDIATELY UPSTREAM OF THE EXIT WINDOW. IBEAM(nA) = 0.0412 + 6.0835 VSF (R2=0.9999) (VSF) IS USED BY THE C.S.TO STOP THE BEAM IF THE RATE XCEEDS A PRESET LIMIT. NORMAL TREATMENT BEAM CURRENT 4 nA CLINICAL DOSE RATE =1220 Gy/min. U.M. =K Alarm current Zmarkus = 1 mm Full energy beam Normal current INFN-LNS
ABSOLUTE DOSIMETRY (BEAM CALIBRATION, cGy/U.M.) FOR REFERENCE DOSIMETRY, AT LNS, A PLANE-PARALLEL MARKUS ION CHAMBER IS USED IN A WATER PHANTOM, BY EXTENDING TO PROTON BEAMS THE FORMALISM OF THE IAEA 381 CODE OF PRACTICE ON THE USE OF PLANE-PARALLEL ION CHAMBERS IN HIGH ENERGY ELECTRON DOSIMETRY. cGy ( TAB. 7.1 ICRU 59) (Wair/e)cp = 1.031 (ICRU 59) cGy/nC (nC) (sw,air)q pQ = 1.00 (ICRU 59, IAEA TRS-398, MED.PHYS. 1995 :22) FOR DOSE CALIBRATIONS OF INDIVIDUAL PATIENTS, THE CALIBRATION IS MADE IN A UNIFORM DOSE REGION AT THE MIDDLE OF SOBP, AT ISOCENTRE, WITH THE 25 mm DIAMETER REFERENCE COLLIMATOR ON BEAM DELIVERY NOZZLE. FOR DAILY DOSE CHECKS, A PMMA PHANTOM BLOCK IS USED: A RADIOACTIVE CHECK DEVICE (90Sr) IS USED FOR OPERATIONAL AND CONSTANCYCHECKS OF DOSIMETRY SYSTEM (1%, NCRP). INFN-LNS
AN EXRADIN T1 THIMBLE ION CHAMBER [A-150 TE, 0.05 cm3, NK(60Co)] CAN BE USED AT LNS IN THE WATER PHANTOM, ACCORDING TO ICRU 59 BECAUSE OF THE Cavity length AND Outer diameter of the chamber, USE IS LIMITED TO ( ) 15 mm AND (SOBP WIDTH) 12 mm. WORK IN PROGRESS NOW AVAILABLE BY PTW ND,w,60CoFOR T1 EXRADIN AND MARKUS LNS CHAMBERS ADOPTION OF IAEA TRS-398 CODE OF PRACTICE Dw,Q = MQ ND,w,Q0 kQ,Q0 ESTIMATED UNCERTAINTY OF Dw,QUNDER REFERENCE CONDITIONS (1 S.D.) ICRU 59 (Nk=1%) IAEA TRS-398 (ND,w,Q0 =1%) EXRADIN T1 2.6% 2.1% PTW MARKUS 3.1% 2.5% INFN-LNS
INFN-LNS PROTON DOSIMETRY INTERCOMPARISON (ECHED, ICRU 59) BEFORE STARTING LNS PROTON THERAPY FACILITY ) INTERCOMPARISON (1) AT CCO (PHYSICA MEDICA, VOL.XV, N.3) LNS (PTW MARKUS, T1-EXRADIN) CCO (A-150 FW-IC18,REFERENCE DOSE) (Dw) [(PTW MARKUS) / (FW-IC18)]: 1% (Dw) [(T1-exradin) / (FW-IC18)]: 1.4% ) INTERCOMPARISON (2) AT PSI (PHYSICA MEDICA VOL.XVII, S.3, PTCOG XXX 1999) 1) CCO (FW-IC18, MARKUS) 2) LNS (MARKUS, EXRADIN-T1, PTW PMMA FARMER) 3) PSI (PTW GRAPHITE FARMER, MARKUS) 4) TERA ( MARKUS, EXRADIN-T1) % = 1.2% max (%) = 3.5%
INFN-LNS DOSIMETRY INTERCOMPARISON AT PSI (FULLY MODULATED BEAM)
OUTPUT FACTORS (FSDF) ) AT THE INITIAL CALIBRATION OF CYCLOTRON THE OUTPUT DOSE RATE HAS BEEN MEASURED FOR ALL COLLIMATORS AREA ENCOUNTERD IN THE CLINICAL PRACTICE. ) TO EVALUATE IF THERE ARE SIGNIFICANT DROPS OF DOSE PER MONITORUNIT WHEN COLLIMATOR DIAMETER DECREASES TO A FEW MILLIMETERS. ) THE MOST RELIABLE RELATIVE OUTPUT VALUES ARE PROVIDED FROM RADIOCHROMIC DETECTORS, ESPECIALLY FOR NARROWEST BEAMS (VATNITSKY). NO SIGNIFICANT DECREASE OF BEAM OUTPUT (cGy/U.M.) FOR COLLIMATOR DIAMETER UP TO 5 mm INFN-LNS
OUTPUT FACTORS (FSDF) ) FROM EXPERIMENTAL RESULTS WE CAN STATE THE LOWER BOUND OF COLLIMATOR DIAMETER FOR WHICH FRELATIVE OUTPUT IS MEASURED ACCURATELY WITH EACH DETECTOR 1)MARKUS CHAMBER CAN BE USED IN PROTON BEAMS WITH 12 mm 2)TLDs AND SCANDITRONIX DIODE CAN BE USED UP TO =8 mm IN CLINICAL PRACTICE AT LNS GAF-DETECTOR IS USED FOR EVALUATION OF OUTPUT FACTOR FOR SHAPED NARROWEST BEAMS INFN-LNS
PATIENT DOSES (STRAY RADIATION) (PERSONAL MONITORING, TWO PATIENTS TESTED) INFN-LNS A)TWO PERSONAL BADGE-DOSIMETERS WERE PLACED ON THE CHEST OF PATIENTS DURING THE WHOLE PROTON TREATMENT. 1) ENEA FILM BADGE PERSONAL DOSIMETER (, , x)0.05 mSv 2) NRPB PADC NEUTRON PERSONAL DOSEMETER 0.20 mSv PATIENTS COMPLETING TREATMENT INDUCED DOSE RATE 15 Sv/h (10’) 1.5 Sv/h