Discover C f AM. X-ray microanalysis in the electron microscope Needs to be longer!. X-ray microanalysis in the electron microscope. X-ray microanalysis in EMLab. Oxford Instruments – INCA software. X-ray microanalysis. How are X-rays generated?.
E N D
Discover CfAM X-ray microanalysis in the electron microscope Needs to be longer! X-ray microanalysis in the electron microscope
X-ray microanalysis in EMLab Oxford Instruments – INCA software
X-ray microanalysis • How are X-rays generated? • How are X-rays detected? - Energy dispersive X-ray spectrometry (EDX) • Practical EDX – Point & ID; Mapping • Wavelength dispersive X-ray spectrometry (WDX)
Ejection of inner electron and transition between shells characteristic X-rays Generation of X-rays Deceleration of electrons background X-rays
Energy dispersive X-ray spectrum Atomic levels involved in copper Kα and Kβ emission
Lithium-drifted silicon detector X-ray electron-hole pair voltage pulse proportional to energy of X-ray Pulses are amplified, processed and sent to multichannel analyser
X-ray voltage pulse pulse processor Pulse processing
Process time 6 Process time 1 Processing time (dead time)
Practical EDX on the Quanta ► Aperture 3 ► Turn off chamber camera ► Set Process Time
Point & ID Single point Selected region Line of points Grid of points
Tracey Finn/Amy Smith (Ure Museum) Fragments of Egyptian pot Mapping
Mapping Experimental Archaeology: Flint knapping of antler tools
Crystal Analyzing range, eV Lithium Fluoride Pentaerythritol 15,330 - 4,712 4,994 - 1,535 Wavelength dispersive X-ray spectrometry Bragg diffraction
X-ray Ionises counter gas Electrical pulse Wavelength dispersive X-ray spectrometry Gas proportional counter – contains 90% argon, 10% methane
Wavelength dispersive X-ray spectrometry • Advantages High resolution High sensitivity • Disadvantages Slow - spectrum acquired sequentially Cannot be used at low magnifications
Standards need to be run for all elements to be analysed Wavelength dispersive X-ray spectrometry Not necessary for EDX, because EDX intensities can be simulated
Questions? • Demo