An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. The V.P. C.D.R.(Vibrating Probe) Presented by Zac Eichmeyer And BEMsS (BioElectromagnetic student Society)

  2. What's a Vibrating Probe (V.P.)? • The V.P. is an instrument capable of measuring current densities in a biological medium • Capable of measuring 10 nA/cm2 • It was developed in the 70’s to measure current pulses from an Oocyte • Non-Destructive measurements • Can measure activities in live cells • Offers an order of magnitude increase from current detection methods, and is more versatile than dyes

  3. The Vibrating Probe In Action

  4. How does the V.P. Work? • The V.P. works by vibrating a high impedance probe horizontally ~10mm • The vibration is controlled by a Piezoelectric Actuator • The Probe is connected to a Preamplifier which conditions the signal for the Lock-In Amplifier • The Lock-in amplifier locks the probes measurement to that of its vibration so to only read current density at its max displacement • Reducing Noise, and giving a voltage reading at the max of its amplitude • The data is then collected via a data logger and fed into a computer for analyzing

  5. Project Definition • The V.P. Requirements: • The probe • High impedence, greater than 5 MW • Low length to mass ratio • Piezoelectric Actuator (the bender) • Operate between ± 45V • Bend at least 30mm • The Lock-in Amplifier • Lock on to a 150 Hz – 1kHz signal • The Pre-Amp • Low Noise

  6. Project Definition, Con't • The DAQ • Capable of sampling faster than the vibration frequency • <1 kHz max • The Oscillator • Capable of driving a Piezoelectric Actuator at a frequency less then 1kHz • The Mechanical Manipulator • Connects the probe to the microscope • Rigid • Have 1 degree of motion (up and down) • Thermally and electrically neutral

  7. Project Goals • To be able to measure current densities <100 nA/cm2 • To eliminate/control extraneous noise • To build the appropriate test setup to implement the probe with a microscope • Implement into current BEMSS research • Measure current densities in other biological samples • Track Ion concentrations as they relate to Neutrophils

  8. Project Overview (Top Down) • The V.P. Overview(subsystems) • Probes • Piezoelectric Actuator (Bender) • Pre Amp • Lock In Amplifier • Oscillator • DAQ • Housing • Mechanical Manipulator

  9. Project Overview (Top Down)

  10. V.P. Subsystems and DiversThe Probes • Two Probes • Ref Probe • Used to find a Common • Sensing Probe • Used to find the Potential in a given medium • Construction: • Plate probe in gold • Gold is more bio-friendly than steel • Connect to Housing • Drivers • Length to mass ratio • Impedance of the probe • Material

  11. V.P. Subsystems and DiversThe Probe Selection • Sensing Probe • The Elgiloy Stainless Steel Microelectrodes • Dimensions (D x L) 225mm x 80mm • Impedence is 12 MW • Tip Diamter, 1-2 mm

  12. V.P. Subsystems and DiversThe Piezoelectric Acutator • A Piezoelectric Actuator is a device which can deflect when given a voltage • Will use as the vibration device • Capable of vibrating • Between 100 Hz to 1 kHz • Amplitude of ± 10 mm • Operate with a Vpp of 60 Volts • Must be Mountable onto the microscope test setup

  13. V.P. Subsystems and DiversThe Piezoelectric Acutator • Selection: Physik Instumente-PICMA® Ceramic-Encapsulated Piezo Bender Actuator, PL112.10 • Size (L x W x T): 17.8 x 9.6 x .65 [mm] • Deflection: 160 mm (nominal) • Input Voltage Nominal Voltage ± 45 V • Max Voltage ± 90 V

  14. V.P. Subsystems and DiversThe Pre Amplifiers • Pre-Amplifier Drivers • This conditions the signal to feed it into the Lock-in Amp • Low Noise • Differential input • High Input Impedance • At least a 10x gain • Selection: Model 100 Preamplifier • Available at Applicable electronics

  15. V.P. Subsystem and DriversThe Lock In Amplifier • Lock-in Amplifier • Needs to lock the signal from the probes to the oscillator • 150 Hz to 1 kHz ref input • Needs to be able to output out to a DAQ system • Selection: Stanford Research Systems • Model SR810 DAP Lock-In Amplifier

  16. V.P. Subsystems and DiversThe Oscillator • The Oscillator needs to provide a stable signal in the range of 150 Hz to 1 kHz at 45 Vpp • The probe will need to vibrate at its resonance frequency • This range covers all possible frequencies for the probes • Must be relatively stable • Not crucial because the Lock In amplifier will compensate • Selection: Hewlett Packard Model 204C Oscillator

  17. V.P. Subsystem and DriversThe Housing • Holds the Probe and allows the actuator to vibrate the probe • Requirements • Electrically and thermally neutral • Rigid • Allow light through • Selection: • Custom Made according to specification • Material: Lucite

  18. V.P. Subsystem and Drivers Mechanical Manipulator • Test Setup Drivers • Needs to be mechanically stable • Any vibration will be detected as extraneous noise • Needs to be precise • Accurate to ~5 mm • Needs to mount to a microscope setup

  19. V.P. Subsystems and DiversDAQ • The Data Acquisition System • A/D conversion (Measure voltage) • Data logging capabilities • Needs to sample at least every 10 ms • The HOBO Data Logger • HOBO U12 Temp/RH/2 External Channel Logger • USB connection to Computer • Analog Voltage measurement • 12 bit resolution (.6mV) • 0-2.5 v sensor range • Accuracy 1%

  20. Test Plan • Three Phases : Preliminary, Intermediate and Final • Preliminary • Probes • Measure Impedance • Measure Mass • Piezoelectric Actuator • Verify vibration modes • Oscillator • Verify stability and range • DAQ • Connect to computer and start to log • Mechanical manipulator • Connect to Microscope, show degree of freedom • Lock-in Amplifier • Verify its ability to lock onto a signal

  21. Test Plan con't • Intermediate: • Mechanical • Housing • Make sure everything attaches snuggly • Verify the vibration mode with the Probe • Show light can still get through • Mechanical manipulator • Attach housing, show stability • Electrical • Pre-Amp • Verify the gain • Show stability • Probes • Attach to Pre-amp • Calibrate (as shown in the appendix)

  22. Test Plan Con't • Final phase: • Mechanical • Attach everything to the microscope test setup • Show functionality • 3 degrees of movement (2 by the microscope slide) • Electrical • Connect the Pre-amp and Oscillator to Lock in Amp • Verify its sampling correctly • Connect the Lock in amp to the DAQ • Verify its outputting accurately • Measure well known cell actions • To prove calibration • Start measuring voltages off of live cells

  23. Planned Delverables • Mile Stone 1 Deliverables • Construct housing • Make and Prep the Probes • Complete all of the preliminary testing phase • Milestone 2 Deliverables • Complete all of the intermediate testing phase • Physically attach all pieces together • Begin Live cell tests • Design Expo • A fully working VP system, that can be demonstrated live

  24. Schedule

  25. Budget

  26. Expandability and the Future • Expand it into two or three dimensions • Computer controlled • Mechanically threw the computer • Improve sampling rate • Min>50ms • Integrate into BEMSS reasearch

  27. Conclusion • The VP is an instrument capable of measuring current densities as low as 10 nA/cm2 • Used primarily for measurements in biological mediums • The V.P. Overview(subsystems) • Probes • Piezoelectric Actuator (Bender) • Pre Amp • Lock In Amplifier • Oscillator • DAQ • Housing • Mechanical Manipulator

  28. References • Pictures: • Nuccitelli, Jaffe, “AN ULTRASENSITIVE VIBRATING PROBE FOR MEASURING, STEADY EXTRACELLULAR CURRENTS”, www.jcb.org • UMASS Vibrating Probe Facility- http://marlin.bio.umass.edu/biology/kunkel/conte/polarized-probes.html • Reid, Nuccitell, Zhao, “Non-invasive measurement of bioelectric currents with a vibrating probe” Nature Protocols, Vol 2, #3, 2007, p 661-669 • Information, background and resource • Nuccitelli, Jaffe, “AN ULTRASENSITIVE VIBRATING PROBE FOR MEASURING, STEADY EXTRACELLULAR CURRENTS”, www.jcb.org • “A simple method for preparing a Vibrating Probe” T Nakagawa, http://pcp.oxfordjournals.org • Reid, Nuccitell, Zhao, “Non-invasive measurement of bioelectric currents with a vibrating probe” Nature Protocols, Vol 2, #3, 2007, p 661-669

  29. Any Questions?

  30. AppendixCalibration