T82 Single Gas Monitor. 1. On. T82 Start-Up. To turn T82 ON / OFF. Press Function Switch. 8. STEL TWA LO HI ALARM PEAK BEEP BAT CALGAS. 188.8. T82 Start-Up. Display Test. Check the display for proper operation. 9. CO. T82 Start-Up. Sensor ID.
E N D
On T82 Start-Up To turn T82 ON / OFF Press Function Switch 8
STEL TWA LO HI ALARM PEAK BEEP BAT CALGAS 188.8 T82 Start-Up Display Test Check the display for proper operation 9
CO T82 Start-Up Sensor ID Displays the currently installed sensor 10
Bat T82 Start-Up Battery Status & Alarm Check (optional Vibrating alarm check) 11
5 T82 Start-Up 5 Sec. Countdown To access instrument settings press the Function Switch 12
000 T82 Start-Up Normal Operating Mode (15 Second Start-Up Time) 13
000 T82 Viewing Modes Backlight Screen (3 Seconds) Press & release function switch to view modes 14
co T82 Viewing Modes Sensor ID Press & release function switch to view modes 15
-0- T82 Viewing Modes Double Click Function Switch to Zero “BEEP” Press & release function switch to view modes 16
PEAK 23 T82 Viewing Modes Double Click Function Switch to Clear Peak Hold 17
STEL 000 T82 Viewing Modes Note: STEL & TWA averages are cleared each time T82 is turned off. STEL 18
TWA 000 T82 Viewing Modes Note: STEL & TWA averages are cleared each time T82 is turned off. TWA 19
Bat T82 Viewing Modes Battery Status 3 rows = up to 4400 hrs (lithium) 9-volt 2600 hrs (Alkaline) 9-volt 2 rows = less than 75 hrs 20
T82 Viewing Modes Datalog Usage L12 Show user percentage of Datalog memory used. (L01 – L99 or FUL for 100%) 20
T82 Viewing Modes Datalog Start New Session H03 Displays current session from H01 – H99 Double clicking function switch will start a new session. 20
LO ALARM 010 HI ALARM 015 T82 Settings Low Alarm To access T82 settings, press FS during 5 second countdown. To view T82 settings, press & release FS High Alarm To changeT82 settings, use internal + or - keys 21
STEL ALARM 015 TWA ALARM 010 T82 Settings STEL Alarm To access T82 settings, press FS during 5 second countdown. To view T82 settings, press & release FS TWA Alarm To changeT82 settings, use internal + or - keys 22
CALGAS BEEP OFF T82 Settings Calibration Gas To access T82 settings, press FS during 5 second countdown. 025 To view T82 settings, press & release FS Confidence Beep To changeT82 settings, use internal + or - keys 23
LCL T82 Settings Clear Datalog To access T82 settings, press FS during 5 second countdown. To view T82 settings, press & release FS To Clear T82 Datalog Double click the function switch 23
000 CAL P T82 Calibration From the Normal Operating Mode... Attach calibration cup and press & hold FS until “CAL” appears on display Apply gas, wait for 2-5 minutes for toxic gases, 30 sec. for O2. A “P” for pass or “F” for fail will appear along with a “Beep”. Turn off gas and remove cup. 24