1 / 27

Optical Express iDesign System Treatments to Date

Optical Express iDesign System Treatments to Date. Surgery date range: 5/30/2012 – 8/24/2013. Low to Moderate Myopia Methods. iDesign System. WaveScan System. 8,905 consecutive primary LASIK Myopia MSE <= 6D Emmetropia surgical goal 20/20 preop BCVA 1 month followup iFS flap.

Download Presentation

Optical Express iDesign System Treatments to Date

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Optical Express iDesign System Treatments to Date Surgery date range: 5/30/2012 – 8/24/2013

  2. Low to Moderate Myopia Methods iDesign System WaveScan System • 8,905 consecutive primary LASIK • Myopia MSE <= 6D • Emmetropia surgical goal • 20/20 preop BCVA • 1 month followup • iFS flap • 27,290 consecutive primary LASIK • Myopia MSE <= 6D • Emmetropia surgical goal • 20/20 preop BCVA • 1 month followup • iFS flap

  3. iDesign System vs WaveScan LASIK System (UK)Low – Moderate Myopia <= 6D 2013.10.17-RF7705

  4. 1M Predictability MSE

  5. 1M Attempted vs Achieved MSE

  6. 1M Cylinder IRC vs SIRC

  7. 1M Correction Ratio by Intended Cylinder

  8. Axis Shift

  9. Monocular Postop UCVA

  10. “Thinking about your vision during the last week, how satisfiedare you with your vision? (without the use of glasses or contact lenses)”

  11. “Has your overall vision turned out to be: “

  12. “Thinking about your vision during the last week, please rate the degree of difficulty you experienced with Glare WITHOUT WEARING glasses or contact lenses”

  13. “During the last week, how much difficulty do you have driving at night? “

  14. iDesign System for High Myopia

  15. High Myopia Methods iDesign System WaveScan System • 1,227 consecutive primary LASIK • Myopia MSE > 6D • Emmetropia surgical goal • 20/20 preop BCVA • 1 month followup • iFS flap • 3,211 consecutive primary LASIK • Myopia MSE > 6D • Emmetropia surgical goal • 20/20 preop BCVA • 1 month followup • iFS flap

  16. iDesign System vs WaveScan LASIK System (UK)High Myopia MSE > 6D

  17. 1M Predictability MSE

  18. 1M Attempted vs Achieved MSE

  19. 1M Cylinder IRC vs SIRC WaveScan System iDesign System

  20. 1M Correction Ratio by Intended Cylinder iDesign System WaveScan System

  21. Monocular Postop UCVA

  22. “Thinking about your vision during the last week, how satisfiedare you with your vision? (without the use of glasses or contact lenses)”

  23. “Has your overall vision turned out to be“

  24. “Thinking about your vision during the last week, please rate the degree of difficulty you experienced with Glare WITHOUT WEARING glasses or contact lenses”

  25. “During the last week, how much difficulty do you have driving at night? “

  26. Conclusions • Comparison of iDesign System to WaveScan CustomVue LASIK System: • Similar refractive predictability • Better cylinder correction • Better UCVA • Better patient reported quality of vision • Better patient reported night driving • Higher level of patient satisfaction