1 / 24

Rotronic goes CO2!

Rotronic goes CO2!. CO2. Topics: CO2 in general CO2 transmitter CO2 & Temp. transmitter CO2 special transmitter . CO2 in general. Big topic over the last year, everybody has it.. e.g. Green house effect and environment … . CO2 in general. CO2 in general. Photosynthesis :

meriel
Download Presentation

Rotronic goes CO2!

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Rotronic goes CO2!

  2. CO2 • Topics: • CO2 in general • CO2 transmitter • CO2 & Temp. transmitter • CO2 special transmitter

  3. CO2 in general

  4. Big topic over the last year, everybody has it.. e.g. Green house effect and environment …

  5. CO2 in general

  6. CO2 in general Photosynthesis: Chemical recationCO2 + H2O --> CHO + O2 + H2O Carbondioxide, water usinglightandchlorophyll, oxygenandglucose

  7. CO2 in general

  8. CO2 in general Indoor air quality in offices leads to high or low levels of gases: Since CO2 is exhaled by people at predictable levels the content of Carbon Dioxide - CO2– in air may be a significant indication of air quality. A measure of CO2 indicates the amount of fresh air supply: 15 cfm = 7.0l/s ventilation rate per occupant corresponds to 1000 ppm CO2 20 cfm = 9.2l/s ventilation rate per occupant corresponds to 800 ppm CO2

  9. CO2 in general • Normal CO2 Levels • The effects of increased CO2 summarized: •  normal outdoor level: 350 - 450 ppm • acceptable levels: < 600 ppm • complaints of stiffness and odors: 600 - 1000 ppm • ASHRAE and OSHA standards: 1000 ppm • general drowsiness: 1000 - 2500 ppm • adverse health effects expected: 2500 - 5000 ppm • maximum allowed concentration within a 8 hour working period: 5000 ppm

  10. CO2 in general • Extreme and Dangerous CO2 Levels • slightly intoxicating, breathing and pulse rate increase, nausea: 30,000 ppm • above plus headaches and sight impairment: 50,000 ppm • unconscious, further exposure death: 100.000 ppm

  11. Rotronic CO2 transmitter Portfolio

  12. CO2 transmitter • CO2 transmitter for climate control • EU/US electricaljunction box • Selectable with and without display • for fixed installation “decreasing your energy consumption while creating a healthier indoor air climate”

  13. CO2 transmitter Specification EU: • Measurement range : 0…2000 ppm • Method infrared (NDIR) w. Auto calibration • Accuracy : ± 30 ppm • Supply : 24 VDC/AC • Output 1 : 0…10 VDC (CO2) • Output 2: 4…20 mA (CO2) • Operation temperature : 0…50 °C • Dimensions (H x W x D): 100 x 80 x 25 mm CF3-W-EU CF3-W-EU-Dips

  14. CO2 transmitter Specification US: • Measurement range : 0…2000 ppm • Measurement method: Infrared (NDIR) w. Auto calibration • Accuracy : ± 30 ppm • Supply : 24 VDC/AC • Output 1 : 0…10 VDC (CO2) • Output 2: 4…20 mA (CO2) • Operation temperature : 0…50 °C • Dimensions (H x W x D): 130 x 85 x 30 mm CF3-W-US CF3-W-US-Dips

  15. CO2 transmitter Specification for duct version: • Measurement range : 0…2000 ppm • Method infrared (NDIR) w. Auto calibration • Accuracy : ± 30 ppm • Supply : 24 VDC/AC • Output 1 : 0…10 VDC (CO2) • Output 2: 4…20 mA (CO2) • Operation temperature : 0…50 °C CF3-D CF3-D-Dips

  16. CO2 & Temperature transmitter • CO2 & temperature transmitter for: • BMS • Controlling Ventilation, Air flaps & valves Wall Duct “decreasing your energy consumption while creating a healthier indoor air climate”

  17. CO2 & Temperature transmitter Specification for wall mount: • Measurement range : 0…2000 ppm • Measurement Temperature: 0.. 50 °C • Method infrared (NDIR) w. Auto calibration • Accuracy : ± 30 ppm • Supply : 24 VDC/AC • Output 1 : 0…10 VDC (CO2) • Output 2: 4…20 mA (CO2) • Operation temperature : 0…50 °C • Dimensions (H x W x D): 120 x 82 x 30 mm CF5-W CF5-W-Dips

  18. CO2 & Temperature transmitter Specification for duct mount: • Measurement range : 0…2000 ppm • Measurement Temperature: 0.. 50 °C • Method infrared (NDIR) w. auto calibration • Accuracy : ± 30 ppm • Supply : 24 VDC/AC • Output 1 : 0…10 VDC (CO2) • Output 2: 4…20 mA (CO2) • Operation temperature : 0…50 °C • Dimensions (H x W x D): 142 x 84 x 46 mm • Ductprobe: 245 mm CF5-D CF5-D-Dips

  19. CO2 special transmitter • Special CO2 transmitters: • „Fresh Air Indication“ • Carbon dioxide alarm • Sound & light indication • relays out’s • bigger ranges • for classrooms , offices , controllers and more “decreasing your energy consumption while creating a healthier indoor air climate”

  20. CO2 special transmitter Special CO2 transmitters wall mount: • Measurement range : 0…2000 ppm • Method Infrared (NDIR) w. Auto calibration • Accuracy : ± 30 ppm • Supply : 24 VDC/AC • Output 1 : 0…10 VDC (CO2) • Sound & light indication • Operation temperature : 0…50 °C • Dimensions (H x W x D): EU: 100x 80 x 28 mm // US: 130x 85 x 30 mm CF3-W-EU-Displ-FLI CF3-W-US-Displ-FLI

  21. CO2 special transmitter Special CO2 transmitters alarm and relays: • Measurement range : 0..4% CO2 • Accuracy : ± 200 ppm • Supply : 24 VDC/AC • Output 1 : 0 … 20mA or 0/2 ... 10 VDC (CO2) • Output 2: 0 … 20mA or 0/2 ... 10 VDC (CO2) • Output 3 (relay 1): Open > 1.4% closed >1.5% (CO2) • Output 4 (relay 2): Open > 2.9% closed >3.0% (CO2) • Operation temperature : 0…50 °C • Dimensions (H x W x D): 142 x 84 x 46mm CF8-W-Disp-AL

  22. CO2 special transmitter Special CO2 transmitters for Green House: • Measurement range : 0…4000 ppm • Accuracy : ± 200 ppm • Supply : 24 VDC/AC • Output 1 : 0 … 20mA or 0/2 ... 10 VDC (CO2) • Output 2: 0 … 20mA or 0/2 ... 10 VDC (°C) • Output 3 (relay): Open > 2.0% closed >1.9% (CO2) • Operation temperature : 0…50 °C • Dimensions (H x W x D): 142 x 84 x 46mm CF8-W-Disp-GH

  23. CO2 special transmitter Special CO2 transmitters for incubators: • Measurement range : 0…3000 ppm • Accuracy : ± 200 ppm • Supply : 24 VDC/AC • Output 1 : 0... 5 VDC, 0…2% (CO2) • Output 2: 0… 5 VDC, 0…3 % (CO2) • Operation temperature : 0…50 °C • Dimensions (H x W x D): 142 x 84 x 46mm CF8-D_W-Disp-IN

  24. Thanks for your attention!