1 / 9

TPFL Calibration Facilities

TPFL Calibration Facilities. TPFL Facilities Include: Facility for calibrating five hole probes, single and cross hot wire probe Facility for calibrating pressure and temperature sensitive paints Facility for calibrating surface mounted hot films More on next slide.

Download Presentation

TPFL Calibration Facilities

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. TPFL Calibration Facilities • TPFL Facilities Include: • Facility for calibrating five hole probes, single and cross hot wire probe • Facility for calibrating pressure and temperature sensitive paints • Facility for calibrating surface mounted hot films • More on next slide TPFL: The Turbomachinery Performance and Flow Research Laboratory Texas A&M University M. T. Schobeiri

  2. TPFL Calibration of Five Hole-, Single- and and X-wire Probes Turn Table with Two Stepper Motors and Controllers for Pitch and Yaw Traversing Turn Table for Probe Alignment TPFL Five- Hole Probe Calibration Facility Probe Tip and Flow Alignment TPFL: The Turbomachinery Performance and Flow Research Laboratory Texas A&M University M. T. Schobeiri

  3. Calibration Precision Examples: Results, Station 4 TPFL: The Turbomachinery Performance and Flow Research Laboratory Texas A&M University M. T. Schobeiri

  4. Calibration Results, Station 5 TPFL: The Turbomachinery Performance and Flow Research Laboratory Texas A&M University M. T. Schobeiri

  5. TPFL Calibration of PSP, TSP Emitted light from TSP Illuminating laser light Thermo-couple Pressure Gage PC Long pass filter(570 nm) Vacuum Pump Camera Direction of air flow Laser(488nm) Vacuum Chamber PSP coated test sample Heater Setup for Pressure and Temperature Sensitive Paints PSP& TSP Calibration TPFL: The Turbomachinery Performance and Flow Research Laboratory Texas A&M University M. T. Schobeiri

  6. TPFL Calibration Facilities Calibration of Pressure, Temperature Sensitive Paints, Results Calibration of TSP TPFL: The Turbomachinery Performance and Flow Research Laboratory Texas A&M University M. T. Schobeiri

  7. TSP-Application to Turbine Blade, Results Comparison of (a) temperatures from TSP results with thermocouple results for, (b) heat transfer coefficient profiles from TSP with LCT TPFL: The Turbomachinery Performance and Flow Research Laboratory Texas A&M University M. T. Schobeiri

  8. Calibration of Surface Mounted Hot Films • Surface mounted hot-film sensors attached to a low pressure Turbine blade. • Multi-Channel anemometer used • Periodic unsteady inlet flow condition. • Anemometer: Samples were taken at a rate of 10 kHz for each of the 50 revolutions of the wake generator. • The data were ensemble-averaged with respect to the rotational period of the wake generator. Connection pins to anemometer-switch board Blade fixture to test section Hot film probe embedded in a foil, see next slide TPFL: The Turbomachinery Performance and Flow Research Laboratory Texas A&M University M. T. Schobeiri

  9. Calibration of Surface Mounted Hot Films Blade with Surface mounted hot film foil installed in test section Calibration curve for hot film sensors Hot Film Sensor Length: 2.00mm TPFL: The Turbomachinery Performance and Flow Research Laboratory Texas A&M University M. T. Schobeiri

More Related