480 likes | 731 Views
WHT A MEDICAL PHYSICST CAN DO TO IMPROVE THE CHANCES OF EARLY DETECTION OF BREAST CANCER. Dr. Nabil Maalej Physics Dept., KFUPM 21-10-2004. Outline. Image Formation in Mammography Image Quality Contrast Scatter and Scatter Reduction Quality Assurance Breast Dose
E N D
WHT A MEDICAL PHYSICST CAN DO TO IMPROVE THE CHANCES OF EARLY DETECTION OF BREAST CANCER Dr. Nabil Maalej Physics Dept., KFUPM 21-10-2004
Outline • Image Formation in Mammography • Image Quality • Contrast • Scatter and Scatter Reduction • Quality Assurance • Breast Dose • Entrance Skin Exposure Measurement • Glandular Dose Measurement in Mammography
Breast Cancer (Highest) • For the total Saudi population, the most common ten cancers are (1994-1996 Cancer Registry): • Female breast cancer (8.8%), • Liver cancer (7.5%), • Leukemia (7.5%), • Non-Hodgkin's lymphoma (7.3%) • Colo-rectum cancer (5.5%). • Thyroid cancer (5.2%), • Lung cancer (4.7%) • Nervous system cancer (3.9%) • Stomach cancer (3.8%) and • Urinary bladder cancer (3.5%)
Image Formation
What Can We do to Get Best Image? X-ray Source X-ray Beam Compression Paddle Compressed Breast Table Chest Wall Grid Image Detector
IMAGE CONTRAST
Imaging Quality Parameters CONTRAST RESOLUTION NOISE
Contrast • Image Contrast is a measure of difference between regions in an image • Good image contrast is particularly important when imaging low-contrast objects such as breast cancer
Mammography Contrast • Image contrast is due to the difference in linear attenuation coefficient of different types of tissue in the breast (adipose tissue (fat), fibro-glandular, tumor).
Attenuation Coefficient Contrast decreases toward higher energies the recommended optimum for mammography is in the region 18 - 33 keV depending on tissue thickness and composition.
I I Ib Ib It It S Contrast Degradation with Scatter
Photoelectric Interaction (Primary) T h hν = T + BEelectron
Compton Scattering (Scatter) T h h’ hν = T + BEelectron + hν’
How Can We Reduce Scatter? • Use an Anti-Scatter Grid • Use and Air Gap between the breast and the detector • Line Scanning System • Capillary optics
Anti-scatter grid breast leadsepta detector
Anti-scatter grid Scattered Rays Primary Rays Grid Septa Image Receptor
Grid Parameters Primary Transmission Scatter Transmission
Grid Parameters Total Transmission Bucky Factor
Contrast Improvement Factor This equation states that for a given transmission factor, the contrast can only be improved when the absorbed dose in the breast is also increased by a Bucky factor B.
MEASURING CONTRAST
Contrast Detail Analysis • Contrast detail analysis: an observer reports the size of the smallest object he or she can perceive at a certain contrast level and with a given noise level in the image.
RMI 156 Phantom ACR minimum score is 4 visible fibers, 3 speck groups and 3 masses
RMI 156 Phantom ACR minimum score is 4 visible fibers, 3 speck groups and 3 masses
Radiation Dose to The Breast
The Nationwide Evaluation of X-ray Trends (NEXT) • The Nationwide Evaluation of X-ray Trends (NEXT) is a U.S. national program conducted annually to measure the x-ray exposure that a standard patient receives for selected x-ray examinations
Mean Glandular Dose • The mean, or average, dose to the glandular tissue within the breast • 3 mGy per view is the federal and ACR dose limit for an average-size patient
Entrance Skin Exposure • The entrance skin exposure is the amount of radiation delivered to the patient skin at the entrance point of the X-ray
SETUP For Entrance Skin Exposure Measurement X-Ray Tube Ionization Chamber Breast Equivalent Phantom X-Ray Film Table Film
Entrance Skin Exposure Measurement Dosimeter Mammography Phantom
Conclusion • Using the Contrast Detail phantom or ACR phantom allows medical physicists to evaluate the image quality. This will ensure good image quality and reduce the chances of miss-diagnosis • Using the simple measurement of the entrance skin exposure and breast dose and comparing them with recommended ranges will ensure safer levels of radiation to the patient • Saudi hospitals should implement such quality control and radiation protection procedures to provide quality care for the patients
X-Ray Production Cathode(-) e- Current (mA) Voltage (KV) e- h h Anode (+) • Electrons are accelerated from cathode to anode. • When high energy electrons hit atoms of heavy metals, the atoms produce X-ray photons.
Exposure Factors • Tube Voltage (kVp) • Tube Current (mA) • Time of exposure (s)
Exposure • Exposure due to radiation is the measurement of amount of charge generated by the radiation per unit mass of air [C/kg or R=2.5810-4 C/kg] • Exposure Tube current (mA) • Exposure Time (s) • Exposure Tube Voltage 2 (kVp 2) • Exposure (1/distance 2) Exposure in Air [C/kg or R] (kVp)2(mA.s)/d2
Radiation-Induced Skin Injuries Figure from [17] • Skin “burns” are rare but possible for prolonged fluorocardio & other interventions • FDA has received 60 reports of burns since 1994 ~ 8.6 reported burns per year • How many radiation burns are not reported? 8