Maintenance Training Course. Course Overview. Itemiser 3 Overview Log On and Calibration Menu Options Maintenance Conclusion. Samples are collected with the sample traps and then vaporized by the desorber
E N D
Course Overview • Itemiser3 Overview • Log On and Calibration • Menu Options • Maintenance • Conclusion
Samples are collected with the sample traps and then vaporized by the desorber The vaporized molecules then pass through the semi-permeable membrane and into the detector The molecules are ionized so that they can be mobilized through the detector by an electric field The length of time it takes a substance to pass through the detector is known as its “time of flight” Light ions travel faster than heavier ions Overview
Regenerative Dryers Dopant Sample Pump Detector Pump Sample Filter Electrodes Desorber Signal Processing Pre-amplifier Semi-permeable membrane (low energy beta) 63NI Overview Display Collector Detector Schematic
Overview Why Maintenance is Necessary • Optimize system performance • Keeps the sample and detector tubing clear of contamination • Prevents nuisance alarms • Minimize down time • Reduce maintenance and repair costs
Log On and Calibration Supervisor default pin number is 5 5 5 5. Maintenance pin number is 9 9 9 9.
Log On and Calibration Calibration Methods
Log On and Calibration Select TNT and insert the calibration trap.
Menu Options • Options • Substances • Config/Notes • History • Status • Advanced Menu • Burn in Membrane
Menu Options A timed calibration warning may be set in this box.
Menu Options Substances levels are set at the factory.
Menu Options Advanced Menu Functions
Maintenance Saying yes to “Burn in Membrane” results in a four hour period of raised temperature that burns off any contaminants that may have been introduced with the installation of a new membrane.
Shift Maintenance Calibrate Check printer paper Clean the unit and work area with alcohol
Weekly Maintenance (With unit powered-on) Check Sample and Detector flow Verify nozzle flow Is approximately 0.2 SCFH @ 200 cc/min
Weekly Maintenance (With unit powered-off and wearing gloves) Clean desorber unit Inspect sample filter Clean sample air lines Inspect fan filter Clean the membrane Inspect dopant
Monthly Maintenance (With unit powered-off and wearing gloves) Change the membrane
Conclusion • Itemiser3 Overview • Log On and Calibration • Menu Options • Maintenance • Questions?